140 GeneChip ® <strong>Mapping</strong> <strong>500K</strong> Assay Manual<br />

• Fields that are highlighted in bold require an entry.<br />

• Drop-down menus are available for Sample/Project information<br />

(default information can be used or new information can be<br />

entered).<br />

• The Experiment Name must be unique.<br />

• Appropriate library files must be installed for a probe array to<br />

appear in the drop-down menu.<br />

3. From the File menu click Save As, or click the Save icon on the tool<br />

bar to register the experiment into the database.<br />

The Sample Information fields can be customized. See the<br />

GeneChip ® Operating Software User’s Guide for further<br />

information.<br />

STEP 2: PREPARING THE FLUIDICS STATION<br />

The Fluidics Station 450 is used to wash and stain the probe arrays; it<br />

is operated using GeneChip Operating Software.<br />

Setting Up the Fluidics Station<br />

1. Turn on the Fluidics Station using the toggle switch on the lower<br />

left side of the machine.<br />

2. Select Run → Fluidics from the menu bar in GCOS.<br />

The Fluidics Station dialog box appears with a drop-down list for<br />

selecting the experiment name for each of the fluidics station<br />

modules. A second drop-down list is accessed for choosing the<br />

Protocol for each of the fluidics station modules. Use the radio<br />

buttons to access each module.<br />

Refer to the GeneChip ® Fluidics Station User’s Guide for<br />

instructions on connecting and addressing multiple fluidics<br />

stations.