80K-40 80K-15<br />

80K-15, 85RF<br />

Specialty probes<br />

80K-6 85RF<br />

42<br />

Caution!<br />

High voltage<br />

80K-40 High Voltage Probe<br />

Extends the voltage measuring<br />

capability of an ac/dc multimeter<br />

to 40,000 V peak ac<br />

or dc. This probe is designed<br />

for taking measurements on<br />

low-energy circuits within<br />

equipment only, such as televisions<br />

or photo copy machines.<br />

• Voltage range: 1 kV to<br />

40,000 kV, dc or peak ac,<br />

28 kV rms ac<br />

• Accuracy: 20 kV to 35 kV;<br />

± 1 % at 20 °C to 30 °C<br />

• Division ratio: 1000:1<br />

• Meets IEC 61010-2-031:1993,<br />

Type B, 40 kV dc or peak ac,<br />

28 kV rms ac, overvoltage<br />

CAT I (voltage derived from<br />

limited energy transformer)<br />

DO NOT use these probes to measure high<br />

voltages on power distribution systems. The<br />

probes provide high accuracy when used<br />

with a multimeter having 10 megohm input<br />

impedance.<br />

80K-6 High Voltage Probe<br />

Extends the voltage measuring<br />

capability of an ac/dc digital<br />

multimeter to 6,000 V<br />

peak ac or dc. This probe is<br />

designed for taking measurements<br />

on low-energy circuits<br />

within equipment only.<br />

• Voltage range: 0 kV to<br />

6 kV, dc or peak ac<br />

• Accuracy: dc to 500 Hz ± 1 %;<br />

500 Hz to 1 kHz ± 2 %; above<br />

1 kHz output falls, -30 % at<br />

10 kHz<br />

• Division ratio: 1000:1<br />

• Meets IEC 61010-2-031:1993,<br />

Type B, 40 kV dc or peak ac,<br />

28 kV rms ac, overvoltage<br />

CAT I (voltage derived from<br />

limited energy transformer)<br />

80K-15 High Voltage Probe<br />

Extends the voltage measuring<br />

capability of an ac/dc digital<br />

multimeter to 15,000 V peak ac<br />

or dc. This probe is designed<br />

for taking measurements on<br />

low-energy circuits within<br />

equipment only.<br />

• Voltage range: 1 kV to 15 kV<br />

dc or peak ac, 10 kV ac rms<br />

• Accuracy: dc ± 2 % 60 Hz;<br />

± 5 %<br />

• Division ratio: 1000:1<br />

• Meets IEC 61010-2-031:1993,<br />

Type B, 40 kV dc or peak ac,<br />

28 kV rms ac, overvoltage<br />

CAT I (voltage derived from<br />

limited energy transformer)<br />

High frequency<br />

85RF High<br />

Frequency Probe<br />

Converts a dc voltmeter with an<br />

input impedance of 10 megohms<br />

into a high frequency (100 kHz<br />

to 500 MHz) RF voltmeter. The<br />

probe’s dc output is calibrated to<br />

be equivalent to the rms value<br />

of a sine wave input.<br />

• Voltage range: 0.25 V to<br />

3 V ac rms<br />

• Maximum input: 30 V rms,<br />

200 V dc<br />

• AC to dc ratio: 1:1<br />

• Ratio accuracy at 25 MHz and<br />

loaded with 10 megohms<br />

–Above 0.5 V: ± 0.5 dB<br />

–Below 0.5 V: ± 1.0 dB<br />

–Add ± 0.2 dB in RF fields of<br />

1 V/m to 3 V/m<br />

• Frequency response relative to<br />

25 MHz:<br />

–100 KHz to 100 MHz<br />

± 0.5 dB<br />

–100 MHz to 200 MHz<br />

± 2.0 dB<br />

–200 MHz to 500 MHz<br />

± 3.0 dB