Edwards Signaling Catalog

08.08.2016 Views

the peak intensity of the beam. This measurement is important in the world of signaling devices where beam intensity is an important application factor. Suppose that two LED’s each emit 0.1 lm total in a narrow beam: One has a 10° solid angle and the other a 5° angle. The 10° LED has an intensity of 4.2 cd, and the 5° LED an intensity of 16.7 cd. They both output the same total amount of light, however - 0.1 lm. A flashlight with a million candela beam may be very bright, but only within its extremely narrowly focused beam. Illuminance Illuminance is the total amount of visible light illuminating, (or incident upon), a point on a surface from all directions above the surface. Illuminance is equivalent to irradiance weighted with the response curve of the human eye. Standard unit for illuminance is Lux (lx), or lumens per square meter (lm/m 2 ). A surface will receive one Luxx of illuminance from a point light source that emits one candela of luminous intensity in its direction from a distance of 1 m. When using the non-standard US units, this translates into one footcandle received from a one candela source one foot away. Illuminance is measured whenever the light level of a particular surface has to be specified. For example, these measurements are required to characterize the light falling on a projector screen or to design light fixtures in a building. Two laws of physics that affect illuminance measurements are the inverse square law and the cosine law. The inverse square law states that the intensity per unit area of a surface varies inversely with the square of the distance between the light source and the detector. Therefore, if illuminance is measured at a particular distance from a source, it is possible to calculate the illuminance at other distances. The cosine law states that the illumination of a surface decreases as a function of the cosine of the incident angle of illumination. This happens because, as the angle of illumination is moved away from the perpendicular to the surface, the area of illumination increases and the flux density per area decreases. Shining a flashlight on a piece of paper at different angles will clearly illustrate this. Illuminance meters use a cosine diffuser that lights and weighs each ambient source’s flux density by the cosine of the angle at which it illuminates the surface, therefore providing cosine corrected results. Usually, a luminance meter has a lens to restrict the field of view of the detector. The human eye is the best-known example of a luminance meter. The unit of luminance is the candela per square meter (cd/m 2 ) in metric units or the footlambert (fl) in English units. The conversion factor is 1 cd/m 2 = 0.2919 fl. A perfectly diffuse source has what is known as a “lambertian” surface and reflects light in all directions following the cosine law. Because of the ways in which light propagates through threedimensional space– spreading out, becoming concentrated, reflecting off shiny or matte surfaces– and because light consists of many different wavelengths, the number of fundamentally different kinds of light measurement is large, and so are the units that represent them. For example, offices are typically "brightly" illuminated by an array of many recessed fluorescent lights for a combined high luminous flux. A laser pointer has very low luminous flux (it could not illuminate a room) but is blindingly bright in one direction (high luminous intensity in that direction). Luminance Luminance, the most commonly measured photometric quantity, is required whenever it is necessary to know the apparent brightness of an object or source. Luminance is the luminous flux emitted from a source per unit solid angle per unit area in a given direction and is, therefore, the luminous intensity per unit area. Luminance measurements are constant, regardless of the distance between the source and the detector because, as the intensity measured by a detector decreases with distance, the area of the measuring field increases proportionately. TM www.edwardssignaling.com 23

TM Science of Light Photometers Photometric measurements are made with instruments called photometers. The devices function by collecting light through some kind of input optics, passing it through a spectral modifying filter and then measuring the light with a photosensitive detector. The filter is carefully trimmed to modify the detector response so that it matches the CIE photopic (or scotopic) function. The detector converts the incoming light energy into an electrical signal, which is then amplified and displayed. Because the filter/detector combination approximates the eye response, the measured electrical signal is a true measure of the light as perceived by a human observer. Note that traditional photometers, because of their inability to simulate the response of the human eye at the ends of the visible spectrum generate significantly flawed data when testing red, blue, and some styles of white LEDs. Spectroradiometers Spectroradiometers are another class of instruments that can be used to perform photometric measurements. A spectroradiometer measures the radiant flux of light at different wavelengths and then mathematically multiplies these spectral values with the CIE-defined photopic values at those wavelengths. By summing these multiplications at small wavelength intervals throughout the visible spectrum, a spectroradiometer accurately calculates photometric quantities. A good spectroradiometer can offer high accuracy for measuring any kind of light source, as there are no filters causing spectral mismatches. These instruments do not require special calibration factors for measuring narrowband LEDs, high-intensity discharge lamps, CRT phosphors, laser projectors, etc. Proper Selection of Visual Signaling Devices General Selecting the “best ” visual signaling device for any particular application can be a daunting challenge for many users. However, the reason behind this complexity is really quite simple: There is much more to consider than the signaling device itself. Choosing properly begins with understanding the reality that we are dealing with a complex system involving not just the signaling device itself, but light and the way our eyes “see” it as well as the way our brains process and perceive it. When comparing two different warning lights, the first question typically asked is which one is “brighter”? This can be a complicated question when one is comparing very different light sources such as rotating incandescent lights, xenon strobes and variable flash rate LEDs of different colors. Let’s briefly discuss three different commonly specified “intensity” ratings: Peak Candela or Peak Candlepower Peak Candela can be defined as the maximum light intensity generated by a flashing light during its light pulse. It indicates NOTHING ABOUT HOW BRIGHT THE LIGHT APPEARS TO THE HUMAN EYE. Peak candela alone cannot be used to directly compare two warning lights. In addition there is no set multiplication factor for converting peak candela, a unit of luminous intensity, to either candela seconds or effective candela, both units of luminous energy. Edwards discourages the use of peak candela ratings when comparing warning lights. Perceptual Phenomena 1. Broca-Sulzer Effect A brief, relatively bright flash of light (optimal flash duration of 0.05 to 0.1 s) is subjectively perceived to be brighter than a longer flash of greater luminance intensity. 2. Brücke-Bartley Effect Below the critical flicker frequency (i.e., the frequency where a flashing light appears constant), the apparent brightness of a flashing light will gradually increase as the frequency is reduced and reach a point (approximately 8 to 10 Hz) where it appears brighter than an uninterrupted light source of equal luminance. 3. Bloch’s Law For sufficiently short stimulus durations, detection threshold decreases inversely with the duration of the stimulus. (Under about 100 milliseconds stimulus duration it is possible to exchange the amount of light for the duration and maintain a constant effect.) 4. Blondel Rey The product of flash intensity times its duration is equal to the asymptotic threshold value times the sum of the duration plus a “visual response time constant”. 24 www.edwardssignaling.com

TM<br />

Science of Light<br />

Photometers<br />

Photometric measurements are made with instruments called<br />

photometers. The devices function by collecting light through<br />

some kind of input optics, passing it through a spectral modifying<br />

filter and then measuring the light with a photosensitive<br />

detector. The filter is carefully trimmed to modify the detector<br />

response so that it matches the CIE photopic (or scotopic)<br />

function. The detector converts the incoming light energy into<br />

an electrical signal, which is then amplified and displayed.<br />

Because the filter/detector combination approximates the eye<br />

response, the measured electrical signal is a true measure of<br />

the light as perceived by a human observer.<br />

Note that traditional photometers, because of their inability to<br />

simulate the response of the human eye at the ends of the<br />

visible spectrum generate significantly flawed data when<br />

testing red, blue, and some styles of white LEDs.<br />

Spectroradiometers<br />

Spectroradiometers are another class of instruments that<br />

can be used to perform photometric measurements. A<br />

spectroradiometer measures the radiant flux of light at different<br />

wavelengths and then mathematically multiplies these spectral<br />

values with the CIE-defined photopic values at those<br />

wavelengths. By summing these multiplications at small<br />

wavelength intervals throughout the visible spectrum, a<br />

spectroradiometer accurately calculates photometric quantities.<br />

A good spectroradiometer can offer high accuracy for<br />

measuring any kind of light source, as there are no filters<br />

causing spectral mismatches. These instruments do not<br />

require special calibration factors for measuring narrowband<br />

LEDs, high-intensity discharge lamps, CRT phosphors, laser<br />

projectors, etc.<br />

Proper Selection of<br />

Visual <strong>Signaling</strong> Devices<br />

General<br />

Selecting the “best ” visual signaling device for any particular<br />

application can be a daunting challenge for many users.<br />

However, the reason behind this complexity is really quite<br />

simple: There is much more to consider than the signaling<br />

device itself. Choosing properly begins with understanding the<br />

reality that we are dealing with a complex system involving not<br />

just the signaling device itself, but light and the way our eyes<br />

“see” it as well as the way our brains process and perceive it.<br />

When comparing two different warning lights, the first question<br />

typically asked is which one is “brighter”? This can be a complicated<br />

question when one is comparing very different light<br />

sources such as rotating incandescent lights, xenon strobes<br />

and variable flash rate LEDs of different colors.<br />

Let’s briefly discuss three different commonly specified<br />

“intensity” ratings:<br />

Peak Candela or Peak Candlepower<br />

Peak Candela can be defined as the maximum light intensity<br />

generated by a flashing light during its light pulse. It indicates<br />

NOTHING ABOUT HOW BRIGHT THE LIGHT APPEARS<br />

TO THE HUMAN EYE. Peak candela alone cannot be used<br />

to directly compare two warning lights. In addition there is no<br />

set multiplication factor for converting peak candela, a unit of<br />

luminous intensity, to either candela seconds or effective candela,<br />

both units of luminous energy. <strong>Edwards</strong> discourages the<br />

use of peak candela ratings when comparing warning lights.<br />

Perceptual Phenomena<br />

1. Broca-Sulzer Effect<br />

A brief, relatively bright flash of light (optimal flash duration of<br />

0.05 to 0.1 s) is subjectively perceived to be brighter than a<br />

longer flash of greater luminance intensity.<br />

2. Brücke-Bartley Effect<br />

Below the critical flicker frequency (i.e., the frequency where<br />

a flashing light appears constant), the apparent brightness<br />

of a flashing light will gradually increase as the frequency is<br />

reduced and reach a point (approximately 8 to 10 Hz) where it<br />

appears brighter than an uninterrupted light source of<br />

equal luminance.<br />

3. Bloch’s Law<br />

For sufficiently short stimulus durations, detection threshold<br />

decreases inversely with the duration of the stimulus. (Under<br />

about 100 milliseconds stimulus duration it is possible to<br />

exchange the amount of light for the duration and maintain a<br />

constant effect.)<br />

4. Blondel Rey<br />

The product of flash intensity times its duration is equal to the<br />

asymptotic threshold value times the sum of the duration plus<br />

a “visual response time constant”.<br />

24<br />

www.edwardssignaling.com

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!