Sensitometric checks may be carried out to evaluate if the performance <strong>of</strong><br />

films in the automatic processors is being maximized. These checks involve<br />

measurement <strong>of</strong> basic fog level, speed and average gradient made at 1° C<br />

intervals <strong>of</strong> temperature. The range <strong>of</strong> temperature measurement depends on<br />

the type <strong>of</strong> chemistry in use, whether cold or hot developer. These three<br />

measurements: fog level, speed, and average gradient, should then be<br />

plotted against temperature and compared with the manufacturer's supplied<br />

figures.<br />

Charlie Chong/ Fion Zhang