<strong>CAPILLARYS</strong> <strong>HEMOGLOBIN</strong>(E) - 2010/10<br />
• When an abnormal hemoglobin is detected, use other means of identification (e.g., globin chain electrophoresis), or consult or send sample to a<br />
specialized laboratory.<br />
IMPORTANT: It is also necessary to analyze the hematologic state, as complementary results.<br />
• The migration of a hemoglobin variant close to Hb A involves an underestimation of Hb A fraction and that of the variant and consequently, an<br />
overestimation of Hb A2 fraction. In order to quantify Hb A2 with precision, it is necessary to delete the separate integration of both variants and<br />
Hb A, and to quantify these fractions together.<br />
• Some homozygous "S" subjects receive a "Hydrea"® (hydroxyurea) treatment that can induce synthesis of foetal hemoglobin. With <strong>CAPILLARYS</strong><br />
<strong>HEMOGLOBIN</strong>(E) procedure, the mobility of the induced hemoglobin F is not different from the physiological hemoglobin F.<br />
• Due to the resolution and sensitivity limits of zone electrophoresis, it is possible that some hemoglobin variants may not be detected with this<br />
method.<br />
Troubleshooting<br />
Call SEBIA Technical Service of the supplier when the test fails to perform while the instruction for the preparation and storage of materials, and for<br />
the procedure were carefully followed.<br />
Kit reagent Safety Data Sheets and information on waste products elimination are available from the Technical Service of the supplier.<br />
PERFORMANCE DATA<br />
Results obtained using the <strong>CAPILLARYS</strong> <strong>HEMOGLOBIN</strong>(E) procedure with the <strong>CAPILLARYS</strong> 2 FLEX PIERCING system indicate a very good<br />
reproducibility for quantitative analysis with a mean CV % of about 0.9 % for each hemoglobin component. All electrophoregrams were also interpreted<br />
visually.<br />
Results presented below have been obtained using the standard parameters of the <strong>CAPILLARYS</strong> software (smoothing 0 and hemoglobin fractions<br />
automatic quantification with <strong>HEMOGLOBIN</strong>(E) analysis program).<br />
Reproducibility within run<br />
Six (6) different blood samples (normal bloods A and E, blood F with increased Hb A2, blood B with Hb S, blood C with increased Hb F and blood G<br />
with Hb F, S and C components) were run using the <strong>CAPILLARYS</strong> <strong>HEMOGLOBIN</strong>(E) procedure in 8 capillaries of 3 <strong>CAPILLARYS</strong> 2 FLEX PIERCING<br />
systems and with 3 lots of <strong>CAPILLARYS</strong> <strong>HEMOGLOBIN</strong>(E) kits.<br />
The mean, SD and CV (n = 8) were calculated for each hemoglobin component, each sample, each lot of kit and each system.<br />
The table shows the values for the 6 tested samples for each hemoglobin component calculated with the 3 lots of kit and the 3 <strong>CAPILLARYS</strong> 2 FLEX<br />
PIERCING systems.<br />
Hemoglobin component Hb A Hb A2 Hb F Hb S Hb C<br />
Blood sample A : lot No. 1 & system No. 1 - lot No. 2 & system No. 2 - lot No. 3 & system No. 3<br />
MEAN (%) 97.4 - 97.35 - 97.4 2.6 - 2.65 - 2.6 / / /<br />
SD 0.03 - 0.05 - 0.05 0.03 - 0.05 - 0.05 / / /<br />
CV (%) 0.03 - 0.1 - 0.05 1.3 - 1.9 - 1.8 / / /<br />
Blood sample B : lot No. 1 & system No. 1 - lot No. 2 & system No. 2 - lot No. 3 & system No. 3<br />
MEAN (%) 58.1 - 58.1 - 58.0 3.0 - 3.0 - 3.0 / 38.9 - 38.9 - 39.0 /<br />
SD 0.10 - 0.16 - 0.06 0.05 - 0.03 - 0.03 / 0.12 - 0.17 - 0.05 /<br />
CV (%) 0.2 - 0.3 - 0.1 1.7 - 1.1 - 1.1 / 0.3 - 0.4 - 0.1 /<br />
Blood sample C : lot No. 1 & system No. 1 - lot No. 2 & system No. 2 - lot No. 3 & system No. 3<br />
MEAN (%) 83.0 - 83.0 - 83.14 2.3 - 2.3 - 2.32 14.7 - 14.7 - 14.54 / /<br />
SD 0.08 - 0.13 - 0.10 0.03 - 0.03 - 0.04 0.05 - 0.10 - 0.10 / /<br />
CV (%) 0.1 - 0.2 - 0.1 1.4 - 1.4 - 1.9 0.3 - 0.7 - 0.7 / /<br />
Blood sample E : lot No. 1 & system No. 1 - lot No. 2 & system No. 2 - lot No. 3 & system No. 3<br />
MEAN (%) 97.5 - 97.4 - 97.3 2.5 - 2.6 - 2.7 / / /<br />
SD 0.04 - 0.04 - 0.07 0.04 - 0.04 - 0.07 / / /<br />
CV (%) 0.04 - 0.04 - 0.1 1.7 - 1.7 - 2.5 / / /<br />
Blood sample F : lot No. 1 & system No. 1 - lot No. 2 & system No. 2 - lot No. 3 & system No. 3<br />
MEAN (%) 93.6 - 93.6 - 93.25 6.4 - 6.4 - 6.75 / / /<br />
SD 0.07 - 0.16 - 0.11 0.07 - 0.16 - 0.11 / / /<br />
CV (%) 0.1 - 0.2 - 0.1 1.1 - 2.5 - 1.7 / / /<br />
Blood sample G : lot No. 1 & system No. 1 - lot No. 2 & system No. 2 - lot No. 3 & system No. 3<br />
MEAN (%) 50.1 - 50.05 - 48.85 2.2 - 2.25 - 2.45 24.5 - 24.57 - 24.26 18.7 - 18.59 - 19.66 4.5 - 4.54 - 4.78<br />
SD 0.25 - 0.19 - 0.21 0.04 - 0.05 - 0.09 0.08 - 0.11 - 0.10 0.19 - 0.11 - 0.11 0.08 - 0.05 - 0.04<br />
CV (%) 0.5 - 0.4 - 0.4 1.9 - 2.2 - 3.5 0.3 - 0.4 - 0.4 1.0 - 0.6 - 0.6 1.8 - 1.1 - 0.9<br />
In addition, none of the repeats showed false positive or false negative values.<br />
- 52 -