Fig . 14) Dispersion surfaces for Barium in<br />

the (110)-plane . The solidlines correspond<br />

to the main symmetry directions,<br />

the points to the various observed pealc<br />

positions of intensity . The transition from<br />

lighter to darker points corresponds to that<br />

from longitudinal to transverse polarisation<br />

of the phonons .<br />

A fmal aspect for optimising the spectrometer shall now be mentioned . Considering the dispersion<br />

for Barium it becomes obvious that the dependence on wave vector for the so called<br />

acoustic phonons is not exceedingly large towards the zone boundary . Optical phonons (sec<br />

the lecture on phonons and magnons) exhibit this property in most cases even more clearly .<br />

This means that one could achieve more intensity or shorter measuring times by reducing the<br />

-resolution and keeping the energy resolution constant, i.e . keeping the distinctness of the<br />

various phonon branches . This can be achieved by means of a horizontally culved monochromator<br />

and/or analyzer as drafted in Fig . 15) . About a factor of five in irrtensity may bc<br />

gained by this horizontal arrangement. Experiments using doubly focusing deflection crystals<br />

exist since a couple of years requiring, however, a high degree of skill and experience .<br />

k<br />

1<br />

1<br />

0#0#0#0#0 0=0=0=0=0<br />

1 2 3 4 5 1 2 3 4 5<br />

k#k#k#k#k<br />

k=k=k=k=k<br />

1 2 3 4 5 1 2 3 4 5<br />

Fig. 15)LA hand side : energy and resolution are determined by the deflection angle O as well as by the collimators<br />

absorbing all neutrons travelling outside the accepted divergence . Right hand side : for the so called<br />

monochromatic focusing, all deflection angles and thus the energy of the neutrons are equal due to the curved<br />

crystal such that the divergent (i .e . relaxed Q-resolution) but monochromatic neutrons merge together at the focusing<br />

point . Caution! the average energy of the set-up bas now become dependent on the sample position due<br />

to the omitted collimators . The resolution will depend on the size of source and sample relative to the radius of<br />

curvature .<br />

9- 13