Speedy Hardware Specification<br />

Supply Voltage<br />

8 to 24 volt DC<br />

Power consumption<br />

2.1 watt<br />

IP rating<br />

IP68<br />

Operating temperature range -20˚C to +50˚C (+40˚C in Ex zone 1)<br />

Ex approval (optional)<br />

II 2G EEx ib IIB T4<br />

RS 485 interface<br />

38400 baud, 8 data bit, 1 stop bit, no<br />

parity<br />

Measurement principle<br />

Ultrasonic continuous wave DOPPLER<br />

Measurement frequency<br />

1 MHz<br />

Flow velocity measurement range +/- 6m/s<br />

Temperature measurement range -20˚C to +50˚C<br />

Medium contacting material Polyurethane, stainless steel 1.4571<br />

(V4A), PVDF<br />

Cable type<br />

LiYC11Y 2x1.5 + 1x2x0.34<br />

Speedy Wiring Details<br />

The Pulsar <strong>FlowCERT</strong> is interfaced to Speedy via a digital communications<br />

interface through the RS485 terminals. The speedy also takes its power from<br />

the <strong>FlowCERT</strong> unit.<br />

The terminal strip for the wall mount <strong>FlowCERT</strong> is as detailed below.<br />

There is also a wiring diagram inside the terminal access cover.<br />

Details on connecting the Speedy sensor to the <strong>FlowCERT</strong> are as follows:<br />

Speedy cable <strong>FlowCERT</strong> terminals Description<br />

Black 34 Shield<br />

Green 35 RS485 -<br />

White 36 RS485 +<br />

Red 41 24 volt DC supply<br />

Blue 44 0 volt supply GND<br />