Product Specification<br />

Physical<br />

Wall Mount<br />

Outside dimensions<br />

Weight<br />

Enclosure material/description<br />

235 x 184 x 120 mm<br />

(9.25 x 7.24 x 4.72 inch)<br />

Nominal 1 kg (2.2 lbs)<br />

Polycarbonate, flame resistant to<br />

UL91<br />

Cable entry detail 10 cable entry knock outs, 1 x M16,<br />

5 x M20 underside,<br />

4 x 18mm dia (PG11) at rear<br />

Transducer cable extensions<br />

Maximum separation<br />

2-core screened<br />

1000 m (3,280 ft)<br />

Environmental<br />

IP Rating (Wall) IP65 (NEMA 4X)<br />

Max. & min. temperature (electronics)<br />

-20 ºC to +50 ºC (-4˚F to +120˚F)<br />

Flammable atmosphere approval<br />

Safe area: compatible with approved<br />

dB transducers (see transducer spec'<br />

sheet)<br />

CE approval EMC approval to BS EN 50081-1:1992<br />

for emissions and BS EN50082-<br />

2:1995 for immunity,<br />

and to BS EN61010-1:1993 for low<br />

voltage directive<br />

Performance<br />

Accuracy<br />

Resolution<br />

Max. range<br />

Min. range<br />

Rate response<br />

Dependant on application and sensor<br />

used, typically with standard ultrasonic<br />

transducer, 0.25% of the measured<br />

range or 6 mm (0.24") (whichever is<br />

greater). DUET 0.025% of Max. range.<br />

Dependant on application and sensor<br />

used, typically with standard ultrasonic<br />

transducer, 0.1% of the measured<br />

range or 2 mm (0.08") (whichever is<br />

greater). DUET 0.1 mm.<br />

Dependant on application and<br />

transducer (maximum 15m (49ft)<br />

dB15)<br />

Dependent upon application and<br />

transducer (minimum zero dB Mach3)<br />

fully adjustable<br />

Echo Processing<br />

Description<br />

Outputs<br />

Analogue output (x2)<br />

DATEM (Digital Adaptive Tracking of<br />

Echo Movement)<br />

Isolated (floating) output (to 150V) of<br />

4-20 mA or 0-20 mA into 500Ω (user<br />

programmable and adjustable) 1µA<br />

resolution<br />