OCM Parameters<br />

PMD Setup<br />

P700 Primary Measuring Device Type<br />

This parameter is used to select the type of Primary Measuring Device and<br />

enable additional parameters required to calculate the flow of the particular<br />

Primary Measuring Device chosen (P701). Options are as follows:<br />

0 = Off (Default)<br />

1 = Exponent<br />

2 = BS3680 Flume<br />

3 = BS3680 Weir<br />

4 = Area/Velocity<br />

5 = Special<br />

6 = Universal<br />

P701 Primary Measuring Device<br />

Enter the Primary Measuring Device used.<br />

If P700 = 1 (Exponent)<br />

Select from the following options:<br />

1 = Suppressed Rectangular Weir<br />

2 = Cipolletti (Trapezoidal) Weir<br />

3 = Venturi Flume<br />

4 = Parshall Flume<br />

5 = Leopold Lagco Flume<br />

6 = V- notch Weir,<br />

7 = Others<br />

8 = Rectangular Weir with End Contractions<br />

If P700 = 2 (BS 3680 Flume)<br />

Select from the following options:<br />

1 = Rectangular<br />

2 = Rectangular with hump<br />

3 = U-throated<br />

4 = U-Throated with hump<br />