Deltabar S PMD70/75 FMD76/77/78
Deltabar S, PMD70, PMD75, FMD76, FMD77, FMD78 - from AoteWell
Deltabar S, PMD70, PMD75, FMD76, FMD77, FMD78 - from AoteWell
- No tags were found...
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
www.aotewell.com www.fa-market.com info@aotewell.com<br />
<strong>Deltabar</strong> S<br />
Certificates and approvals<br />
CE mark<br />
Ex approvals<br />
The device meets the legal requirements of the relevant EC directives.<br />
Endress+Hauser confirms that the device has been successfully tested by applying the CE mark.<br />
•ATEX<br />
•FM<br />
•CSA<br />
•NEPSI<br />
•IECEx<br />
• GOST on request<br />
• Also combinations of different approvals<br />
All explosion protection data are given in separate documentation which is available upon request. The<br />
Ex documentation is supplied as standard with all devices approved for use in explosion hazardous areas.<br />
ä 85, "Safety Instructions" and "Installation/Control Drawings" sections.<br />
Suitability for hygienic<br />
processes<br />
The <strong>Deltabar</strong> S is suitable for use in hygienic processes.<br />
Overview of suitable process connections from ä 81 ff.<br />
Many versions meet the requirements of 3A-Sanitary Standard No. 74 and<br />
are certified by the EHEDG.<br />
Suitable fittings and seals must be used to ensure hygiene-compliant<br />
design according to the specifications of 3A and EHEDG.<br />
Note!<br />
The gap-free connections can be cleaned without residue using<br />
the usual cleaning methods.<br />
TYPE EL<br />
October 2007<br />
74 -<br />
Marine certificate<br />
• GL: <strong>FMD76</strong>, FMD<strong>78</strong>, <strong>PMD70</strong>, PMD<strong>75</strong><br />
• ABS: <strong>FMD76</strong>, FMD<strong>78</strong>, <strong>PMD70</strong>, PMD<strong>75</strong><br />
Functional Safety SIL /<br />
IEC 61508 Declaration of<br />
Conformity (optional)<br />
Overfill protection<br />
CRN approvals<br />
Pressure Equipment Directive<br />
(PED)<br />
The <strong>Deltabar</strong> S with 4 to 20 mA output signal has been developed to IEC 61508 standard. The device can be<br />
used for flow, level and differential pressure monitoring up to SIL 3.<br />
È For a detailed description of the safety functions with <strong>Deltabar</strong> S, settings and characteristic quantities for<br />
functional safety, please refer to the "Functional Safety Manual - <strong>Deltabar</strong> S" SD189.<br />
È For devices with SIL / IEC 61508 Declaration of Conformity see ä 69 ff, feature 100 "Additional<br />
option 1" and feature 110 "Additional option 2" version E "SIL / IEC 61508, Declaration of Conformity".<br />
WHG. See "Ordering information" ä 69 (see also ZE259P/00/DE).<br />
Some device versions have CRN approval. For a CRN-approved device, a CRN-approved process connection<br />
( ä 69 ff, feature 70 "Process connection") has to be ordered with a CSA approval ( ä 69 ff, feature 10<br />
"Approval"). These devices are fitted with a separate plate bearing the registration number 0F10524.5C.<br />
The devices <strong>PMD70</strong>, PMD<strong>75</strong>, <strong>FMD76</strong>, FMD<strong>77</strong> and FMD<strong>78</strong> correspond to Article 3 (3) of the EC directive<br />
97/23/EC (Pressure Equipment Directive) and have been designed and manufactured according to good<br />
engineering practice.<br />
The following also applies:<br />
– FMD<strong>78</strong> with pipe diaphragm seal 1.5"/PN40:<br />
Suitable for stable gases in group 1, category II<br />
– PMD<strong>75</strong>, PN 420<br />
Suitable for stable gases in group 1, category I<br />
Standards and guidelines DIN EN 60<strong>77</strong>0 (IEC 60<strong>77</strong>0):<br />
Transmitters for use in industrial-process control systems<br />
Part 1: Methods for performance evaluation<br />
DIN 16086:<br />
Electrical pressure measuring instruments, pressure sensors, pressure transmitters, pressure measuring<br />
instruments, concepts, specifications on data sheets<br />
EN 61326-X:<br />
EMC product family standard for electrical equipment for measurement, control and laboratory use.<br />
68 Endress+Hauser<br />
AoteWell International.inc sales@aotewell.com Tel: +852-6563-2160