www.aotewell.com www.fa-market.com info@aotewell.com<br />

<strong>Deltabar</strong> S<br />

Performance characteristics – metal process isolating<br />

diaphragms<br />

Reference accuracy –<br />

PMD<strong>75</strong>, FMD<strong>77</strong>, FMD<strong>78</strong><br />

The reference accuracy comprises the non-linearity (terminal based), hysteresis and non-reproducibility as per<br />

IEC 60<strong>77</strong>0. The data refer to the calibrated span.<br />

The following applies for the root-extracting characteristic curve:<br />

The accuracy data of the <strong>Deltabar</strong> S are taken into the accuracy calculation of the flow rate with a factor of 0.5.<br />

PMD<strong>75</strong><br />

Measuring cell<br />

% of the set span<br />

10 mbar (0.15 psi)<br />

30 mbar (0.45 psi)<br />

• TD 1:1<br />

• TD > 1:1<br />

=<br />

=<br />

±0.15<br />

±0.15 x TD<br />

100 mbar (1.5 psi)<br />

• TD 1:1 to TD 4:1<br />

•TD > 4:1<br />

=<br />

=<br />

±0.0<strong>75</strong><br />

(0.012 x TD + 0.027)<br />

500 mbar (7.5 psi)<br />

• TD 1:1 to TD 15:1<br />

• TD > 15:1<br />

=<br />

=<br />

±0.0<strong>75</strong><br />

(0.0015 x TD + 0.053)<br />

Platinum version:<br />

00 mbar (1.5 psi)<br />

•TD 1:1 = ±0.05<br />

FMD<strong>77</strong>, FMD<strong>78</strong><br />

Measuring cell FMD<strong>77</strong> FMD<strong>78</strong><br />

% of the set span (influence of the diaphragm seal included)<br />

100 mbar (1.5 psi)<br />

• TD 1:1 to TD 4:1<br />

• TD > 4:1<br />

= ±0.15<br />

= (0.03 x TD + 0.03)<br />

• TD 1:1 to TD 4:1<br />

• TD > 4:1<br />

= ±0.15<br />

= (0.03 x TD + 0.03)<br />

500 mbar<br />

(7.5 psi)<br />

3 bar (45 psi)<br />

16 bar (240 psi)<br />

• TD 1:1 to TD 15:1<br />

•TD > 15:1<br />

= ±0.0<strong>75</strong><br />

= (0.0015 x TD + 0.053)<br />

• TD 1:1 to TD 4:1<br />

•TD > 4:1<br />

= ±0.15<br />

= (0.02 x TD + 0.07)<br />

40 bar (600 psi) <br />

• TD 1:1 to TD 4:1<br />

•TD > 4:1<br />

= ±0.15<br />

= (0.02 x TD + 0.07)<br />

Total performance – PMD<strong>75</strong><br />

The "Total performance" specification comprises the non-linearity including hysteresis, non-reproducibility, the<br />

thermal change of the zero point as well as the influence of the static pressure (p st = 70 bar (1050 psi)).<br />

All specifications apply to the temperature range –10 to +60 °C (+14 to +140 °F).<br />

Measuring cell<br />

AISI 316L/1.4435, Alloy, gold-rhodium or<br />

Monel process isolating diaphragm<br />

Tantalum process isolating diaphragm<br />

% of the set span<br />

500 mbar<br />

(7.5 psi) to TD 2:1<br />

0.15 0.30<br />

Total error<br />

The total error comprises the long-term stability and the total performance.<br />

All specifications apply to the temperature range –10 to +60 °C (+14 to +140 °F) and a turn down of 1:1.<br />

Measuring cell<br />

AISI 316L/1.4435, Alloy, gold-rhodium or<br />

Monel process isolating diaphragm<br />

Tantalum process isolating diaphragm<br />

% of URL/year<br />

10 mbar (0.15 psi)<br />

30 mbar (0.45 psi)<br />

0.33 <br />

100 mbar (1.5 psi) 0.33 0.48<br />

500 mbar (7.5 psi) 0.20 0.35<br />