Micropilot M<br />
Performance characteristics<br />
Reference operating<br />
conditions<br />
Maximum measured error<br />
• temperature = +20 °C (68 °F) ±5 °C (9 °F)<br />
• pressure = 1013 mbar abs. (14.7 psia) ±20 mbar (0.3 psi)<br />
• relative humidity (air) = 65 % ±20%<br />
• ideal reflector<br />
• no major interference reflections inside the signal beam<br />
Typical statements for reference conditions, include linearity, repeatability, and hysteresis:<br />
.<br />
Type of device to 10 m ex 10 m<br />
<strong>FMR</strong> <strong>230</strong> ±10 mm ±0.1% of measuring range<br />
<strong>FMR</strong> <strong>231</strong> ±10 mm ±0.1% of measuring range<br />
<strong>FMR</strong> <strong>240</strong> ±3 mm ±0.03% of measuring range<br />
Resolution<br />
Reaction time<br />
Influence of ambiente<br />
temperature<br />
Digital / analog in % 4…20 mA<br />
• <strong>FMR</strong> <strong>230</strong>: 1mm / 0.1 % of measuring range<br />
• <strong>FMR</strong> <strong>231</strong>: 1mm / 0.1 % of measuring range<br />
• <strong>FMR</strong> <strong>240</strong>: 1mm / 0.1 % of measuring range<br />
The reaction time depends on the parameter settings (min. 1 s). In case of fast level changes, the<br />
instrument needs the reaction time to indicate the new value.<br />
0.006% / 10 K referring to max. measuring range<br />