30 Fig. 3.2a: Apparatus for the calibration of long-stem platinum resistance thermometers at the argon triple point using a sealed cell: 1, long-stem platinum resistance thermometer; 2, stainless-steel body of cell; 3, thermometer tube; 4, polyurethane foam; 5, solid-liquid argon; 6, bath of liquid nitrogen; 7, helium gas inlet; 8, manometer; 9, valve; 10, filling tube for liquid nitrogen; 11, cryostat [Bonnier and Moser (1983)];

31 Fig. 3.2b: Apparatus for the calibration and comparison of long-stem and capsule platinum resistance thermometers from the triple point of argon to the triple point of indium [Ancsin and Phillips (1984)].

30<br />

Fig. 3.2a: Apparatus <strong>for</strong> <strong>the</strong> calibration of long-stem platinum resistance <strong>the</strong>rmometers at<br />

<strong>the</strong> argon triple point using a sealed cell: 1, long-stem platinum resistance<br />

<strong>the</strong>rmometer; 2, stainless-steel body of cell; 3, <strong>the</strong>rmometer tube; 4,<br />

polyurethane foam; 5, solid-liquid argon; 6, bath of liquid nitrogen; 7, helium gas<br />

inlet; 8, manometer; 9, valve; 10, filling tube <strong>for</strong> liquid nitrogen; 11, cryostat<br />

[Bonnier and Moser (1983)];