4. Self test during treatment4.1 The equipment should perform a self testduring treatment automatically in fixedtime slot not less than 10 min to ensure theintegrity of close system.5. Heparin pump5.1 Infusion rate : 0-10 ml/hr in 1ml/hrincrement.5.2 Accuracy : ±5%5.3 Counter pressure:5.4 Stop time: Heparinization stop time(before end of treatment) 0 – 9 hr 59 minuser adjustable in 1 min increment5.5 Accumulated heparin volume: 0 to 999.9ml6. Pressure monitoring & alarm6.1 Venous pressure monitoring range: -700 to+700mmhg Accuracy : ±5 mmhg6.2 Venous pressure alarm adjustable high &low alarm limit. Alarm limit should spreadand be reset automatically on adjustmentof blood flow.6.3 Arterial pressure monitoring Range : -700to +700 mmhg Accuracy ±5mmhg6.4 Arterial pressure alarm adjustable high &low alarm limit. Alarm limit should spreadand be reset automatically on adjustmentof blood flow.7. Air detection7.1 Method: Ultrasonic detection7.2 Drip chamber size: 22 mm diameter7.3 Sensitivity – Bubbles >µl will be trappedby the drip chamber.8. Dialysis fluid preparation and monitoring8.1 It should have separate conductivitymonitoring for Acid Concentration,Bicarbonate concentration, Total dialysate