0.080 MWCFF24<br />

X-ray tube assembly "Vari-Aqua"<br />

with water-cooling and Be-window tube<br />

Technical data<br />

<strong>Petrick</strong><br />

1. Geometry: according assembly drawing<br />

2. Type of operating: DC<br />

3. Max. operating voltage: max. 50 kV (but just max. approvable voltage of tube)<br />

4. Long-term capacity rating: depending on type of tube<br />

5. Max. housing temperature: 55°C<br />

6. Leakage radiation: < 2,0 µSv/h at 50mm (50kV, 1000µA)<br />

7. Weight: 3.000 g<br />

Emitter housing:<br />

<strong>GmbH</strong><br />

The X-ray tube assembly made of brass is used as protection housing for small fixed-anode X-ray<br />

tubes from <strong>Petrick</strong> <strong>GmbH</strong> and is completely free of lead-parts.<br />

To avoid overheating of the emitter, a temperature limiter is integrated, which is daisy-chained to the<br />

cathode heating, and it is equipped with a water cooling system.<br />

When achieving max. housing temperature, the temperature limiter turns off the heating of the tube.<br />

Applied insulating materials allow an anode voltage of up to 50 kV DC.<br />

A diaphragm system for compensation of oil extension protects the emitter against flashovers in any<br />

position, because no air bubble is located inside.<br />

Applied insulating oil is specially prepared und filled in under vacuum.<br />

Applicable X-ray tubes:<br />

Tubes with various focal spots, different anode materials and different radiation emission-windows<br />

(100µ und 200µ).<br />

Tube heating:<br />

Tube heating is connected by a 4-pole female connector.<br />

If a grid-controlled tube is applied and shall work with grid control, the joint between black and brown<br />

wire of the provided heating-cable has to be ripped and the necessary grid voltage to be supplied<br />

between black (-) and brown (+) wire. The tube never has to be operated with open grid connection.<br />

Heating data depend on the applied type of tube.<br />

High voltage connection:<br />

High voltage connection results from a high voltage-cable with UHF-plug 5/8-24 UNEF.<br />

Necessarily the connection has to be protected against impurity and humidity by the protection cap,<br />

which is included in delivery. High voltage-cable has to be cleaned if necessary and the dry surface to<br />

be fine coated with Silicone Paste P4 (available from us) before installation.<br />

HV-cables and heating-cables in customized length as well as hose-connection elements for watercooling<br />

are optional available.<br />

Page 2 from 6


<strong>Petrick</strong><br />

0.080 MWCFF24<br />

X-ray tube assembly "Vari-Aqua"<br />

with water-cooling and Be-window tube<br />

Circuit diagram<br />

BNC bulkhead recept. jack<br />

Grid voltage (black)<br />

Heating voltage (brown)<br />

Heating voltage (grey)<br />

When working with grid control,<br />

rip here and supply grid<br />

voltage separately.<br />

-<br />

+<br />

Emitter<br />

- + + -<br />

<strong>GmbH</strong><br />

Heating connection DC<br />

(brown and grey) HV-connection DC<br />

Generator<br />

Page 3 from 6


<strong>Petrick</strong><br />

0.080 MWCFF24<br />

X-ray tube assembly "Vari-Aqua"<br />

with water-cooling and Be-window tube<br />

Detail A, M1:1<br />

Fitting dimensions<br />

200µ-window 100µ-window<br />

6,7<br />

109,0<br />

Ø 15,0<br />

4,1<br />

M 4 thread depht 6 mm<br />

0-ring flute<br />

53,5 x 60,6 x 2,0<br />

ø 47,37<br />

18,9<br />

Ø 69,5<br />

Ø 69,0<br />

4,1<br />

Ø 12,6 G7 7,6<br />

187,0<br />

Grid control - (black)<br />

Ground<br />

HV-connection DC<br />

5/8 - 24 UNEF<br />

Connection water-cooling:<br />

RECTUS Art.Nr.21KBAW10MPN<br />

Hose connection:<br />

RECTUS Art.Nr.21SBKO06MPN<br />

76,0<br />

41,3<br />

<strong>GmbH</strong><br />

Filament - (brown) Filament + (grey)<br />

Do NEVER release<br />

the lock screw<br />

33,1<br />

Ø 63,5<br />

24,0°<br />

Detail A<br />

M 1 : 2<br />

Page 4 from 6


<strong>Petrick</strong><br />

0.080 MWCFF24<br />

X-ray tube assembly "Vari-Aqua"<br />

with water-cooling and Be-window tube<br />

Radiation protection<br />

Brass<br />

Plastic<br />

Brass<br />

Brass<br />

-X-ray tube assembly can emit harmful X-rays<br />

-There is a danger of implosion<br />

-User is account for personal protection by a housing<br />

10 10<br />

8<br />

10 10<br />

12<br />

4<br />

4<br />

<strong>GmbH</strong><br />

Stainless steel<br />

Rubber<br />

Plastic<br />

Oil<br />

Copper<br />

Nickel<br />

FeNiCo<br />

Brass<br />

Monel<br />

Beryllium<br />

( * )<br />

Nickel<br />

Tungsten<br />

Glass<br />

Molybdenum<br />

( * ):<br />

Tungsten<br />

Molybdenum<br />

Tantalum<br />

Copper<br />

Chrome<br />

Rhodium<br />

Cobalt<br />

Copper<br />

Brass<br />

Stainless steel<br />

Page 5 from 6


55<br />

50<br />

°C 45<br />

40<br />

35<br />

0.080 MWCFF24<br />

X-ray tube assembly "Vari-Aqua"<br />

with water-cooling and Be-window tube<br />

Temperature-curve<br />

°C<br />

60<br />

55<br />

50<br />

45<br />

40<br />

35<br />

30<br />

25<br />

20<br />

15<br />

10<br />

Characteristic at 23°C ambient temperature<br />

0,50 0,10 0,05<br />

l/min<br />

<strong>Petrick</strong><br />

Housing temperature at 180 W<br />

0 30 60 90 120 150 180 210 240<br />

min<br />

50 W<br />

1,8<br />

1,6<br />

1,4<br />

1,2<br />

l/min 1<br />

0,8<br />

0,6<br />

0,4<br />

0,2<br />

Pressure-flow-diagram<br />

<strong>GmbH</strong><br />

0 0,5 1,0 1,5 2,0<br />

bar<br />

2,5 3,0 3,5 4,0<br />

Page 6 from 6