SYSTem System<strong>SMIQ</strong>:SYSTem:SERRor?This command returns a list of all errors existing at the point of time of the query. This listcorresponds to the indication on the ERROR page with manual control (cf. Section ErrorMessages).Example: :SYST:SERR?Response: -221, "Settings conflict", 153, "Input voltage out of range":SYSTem:VERSion?The command returns the SCPI version number the instrument acts in accordance with. Thiscommand is a query and thus has no *RST value.Example: :SYST:VERS? Response: 1994.0:SYSTem:SSAVe 1...1000This command saves the current device setting at the memory location indicated.:SYSTem:SREStore 1...1000This command loads a device status that was stored using the :SYSTem:SSAVe command(RESTORE). One of 1000 available memory locations is selected by entering a numeral.! This command has the same effect as the:SYSTem:SREStore command. The setting timehowever is 300 µs less. It is optimized for highest speed and does not comply with the SCPIsyntax regulations. Exactly 3 bytes are transmitted including the '!' (which is the identifier ofthis command). With the last byte, EOI has to be activated as delimiter.The memory location is binary-coded in the 2 bytes indicated.1125.5555.03 3.206 E-9