allows a system to remotely monitor the <strong>CMS</strong> operation, if desired,allowing more specific diagnostics. 14Value Function Comment0 NOT READY Unit is powering-up, or there issome problem1 READY Ready to start a test 152 TESTING Test in progress3 WAITING Waiting between tests 16128 FAULT OPTICAL LED failure / sensor blocked /filled with air129 FAULT FLOWLOWFlow too low for reliable test 17130 FAULT FLOWHIGH131 FAULT LOGGING Fault with data logging132 FAULT WATERSENSORWater sensor failureTable 16.1The TEST STATUS Register14 However the fault conditions are also indicated on the front panel LED, while ``NoResult’’ in the case of a fault is indicated using special result values as previouslydescribed.15 <strong>User</strong> has not set tests to occur automatically.16 <strong>User</strong> has set a non-zero test interval.17 Or fluid is totally clean (no particle counts). Flow alarm can be turned off by userif this is a problem, for example cleaning rigs.56 Fault Finding