<strong>Magellan</strong> ® <strong>1000i</strong>Omni-Directional Imaging ScannerProduct Reference Guide


Datalogic Scanning, Inc.959 Terry StreetEugene, Oregon 97402Telephone: (541) 683-5700Fax: (541) 345-7140An Unpublished Work - All rights reserved. No part of the contents of this documentation or the proceduresdescribed therein may be reproduced or transmitted in any form or by any means without prior written permission ofDatalogic Scanning, Inc. or its subsidiaries or affiliates ("Datalogic" or “Datalogic Scanning”). Owners of Datalogicproducts are hereby granted a non-exclusive, revocable license to reproduce and transmit this documentation for thepurchaser's own internal business purposes. Purchaser shall not remove or alter any proprietary notices, includingcopyright notices, contained in this documentation and shall ensure that all notices appear on any reproductions ofthe documentation.Should future revisions of this manual be published, you can acquire printed versions by contacting your Datalogicrepresentative. Electronic versions may either be downloadable from the Datalogic website (www.scanning.datalogic.com)or provided on appropriate media. If you visit our website and would like to make comments or suggestionsabout this or other Datalogic publications, please let us know via the "Contact Datalogic" page.DisclaimerDatalogic has taken reasonable measures to provide information in this manual that is complete and accurate, however,Datalogic reserves the right to change any specification at any time without prior notice.Datalogic is a registered trademark of Datalogic S.p.A. and the Datalogic logo is a trademark of Datalogic S.p.A. alllicensed to Datalogic Scanning, Inc. All other trademarks and trade names referred to herein are property of theirrespective owners.This product may be covered by one or more of the following patents: 4603262 • 4639606 • 4652750 • 4672215 • 4699447 •4709369 • 4749879 • 4786798 • 4792666 • 4794240 • 4798943 • 4799164 • 4820911 • 4845349 • 4861972 • 4861973 • 4866257• 4868836 • 4879456 • 4939355 • 4939356 • 4943127 • 4963719 • 4971176 • 4971177 • 4991692 • 5001406 • 5015831 • 5019697• 5019698 • 5086879 • 5115120 • 5144118 • 5146463 • 5179270 • 5198649 • 5200597 • 5202784 • 5208449 • 5210397 • 5212371• 5212372 • 5214270 • 5229590 • 5231293 • 5232185 • 5233169 • 5235168 • 5237161 • 5237162 • 5239165 • 5247161 • 5256864• 5258604 • 5258699 • 5260554 • 5274219 • 5296689 • 5298728 • 5311000 • 5327451 • 5329103 • 5330370 • 5347113 • 5347121• 5371361 • 5382783 • 5386105 • 5389917 • 5410108 • 5420410 • 5422472 • 5426507 • 5438187 • 5440110 • 5440111 • 5446271• 5446749 • 5448050 • 5463211 • 5475206 • 5475207 • 5479011 • 5481098 • 5491328 • 5493108 • 5504350 • 5508505 • 5512740• 5541397 • 5552593 • 5557095 • 5563402 • 5565668 • 5576531 • 5581707 • 5594231 • 5594441 • 5598070 • 5602376 • 5608201• 5608399 • 5612529 • 5629510 • 5635699 • 5641958 • 5646391 • 5661435 • 5664231 • 5666045 • 5671374 • 5675138 • 5682028• 5686716 • 5696370 • 5703347 • 5705802 • 5714750 • 5717194 • 5723852 • 5750976 • 5767502 • 5770847 • 5786581 • 5786585• 5787103 • 5789732 • 5796222 • 5804809 • 5814803 • 5814804 • 5821721 • 5822343 • 5825009 • 5834708 • 5834750 • 5837983• 5837988 • 5852286 • 5864129 • 5869827 • 5874722 • 5883370 • 5905249 • 5907147 • 5923023 • 5925868 • 5929421 • 5945670• 5959284 • 5962838 • 5979769 • 6000619 • 6006991 • 6012639 • 6016135 • 6024284 • 6041374 • 6042012 • 6045044 • 6047889• 6047894 • 6056198 • 6065676 • 6069696 • 6073849 • 6073851 • 6094288 • 6112993 • 6129279 • 6129282 • 6134039 • 6142376• 6152368 • 6152372 • 6155488 • 6166375 • 6169614 • 6173894 • 6176429 • 6188500 • 6189784 • 6213397 • 6223986 • 6230975• 6230976 • 6237852 • 6244510 • 6259545 • 6260763 • 6266175 • 6273336 • 6276605 • 6279829 • 6290134 • 6290135 • 6293467• 6303927 • 6311895 • 6318634 • 6328216 • 6332576 • 6332577 • 6343741 • 6454168 • 6478224 • 6568598 • 6578765 • 6705527• 6974084 • 6991169 • 7051940 • 7170414 • 7172123 • 7201322 • 7204422 • 7215493 • 7224540 • 601 26 118.6 • AU703547 •D312631 • D313590 • D320011 • D320012 • D323492 • D330707 • D330708 • D349109 • D350127 • D350735 • D351149 •D351150 • D352936 • D352937 • D352938 • D352939 • D358588 • D361565 • D372234 • D374630 • D374869 • D375493 •D376357 • D377345 • D377346 • D377347 • D377348 • D388075 • D446524 • EP0256296 • EP0260155 • EP0260156 •EP0295936 • EP0325469 • EP0349770 • EP0368254 • EP0442215 • EP0498366 • EP0531645 • EP0663643 • EP0698251 •EP01330772 • GB2252333 • GB2284086 • GB2301691 • GB2304954 • GB2307093 • GB2308267 • GB2308678 • GB2319103 •GB2333163 • GB2343079 • GB2344486 • GB2345568 • GB2354340 • ISR107546 • ISR118507 • ISR118508 • JP1962823 •JP1971216 • JP2513442 • JP2732459 • JP2829331 • JP2953593 • JP2964278 • MEX185552 • MEX187245 • RE37166 • OtherPatents Pending


Minimum Reads ........................................................................................................... 5-63Optical and Read Performance Parameters ..............................................................................A-1Scanner Dimensions ............................................................................................................A-1Physical Properties ...............................................................................................................A-2Electrical Parameters ...........................................................................................................A-2Environmental Parameters ....................................................................................................A-2Other Parameters ................................................................................................................A-2Standard Cable Pinouts (Primary Interface Cables) ...................................................................B-1RS-232 .........................................................................................................................B-1IBM Port 5B/9B/17 .........................................................................................................B-1IBM USB .......................................................................................................................B-2USB & USB Keyboard .....................................................................................................B-2Wand Emulation ............................................................................................................B-2Keyboard Wedge ............................................................................................................B-2Defaults by Symbology ........................................................................................................ D-1Interface Default Exceptions ................................................................................................ D-1IBM Interfaces .............................................................................................................. D-2RS-232 Wincor/Nixdorf .................................................................................................. D-2Wand Emulation ........................................................................................................... D-3Keyboards .................................................................................................................... D-3Keyboard Model Cross Reference ...........................................................................................E-1Accepting RS-232 Commands ............................................................................................... F-1i-iv<strong>Magellan</strong> ® <strong>1000i</strong>


Chapter 1Getting StartedAbout This ManualManual ConventionsThe <strong>Magellan</strong> ® <strong>1000i</strong> Omni-Directional Imaging Scanner offers hands-free scanning for small,easily handled items and handheld scanning for bulkier items. Its aggressive imaging performanceand intuitive operation reduces user training and speeds checkout for better customerservice.This manual presents advanced user information which includes connection, programming,product and cable specifications, and other useful references. For additional information, suchas installation, maintenance, troubleshooting and warranty information, see the Quick ReferenceGuide (QRG). Copies of other publications for this product are downloadable free ofcharge from the website listed on the back cover of this manual.On leaving the factory, units are programmed for the most common terminal and communicationssettings. If you need to change these settings, custom programming can be accomplishedby scanning the bar codes in this guide.Bold text and a yellow-highlighted background indicates the most common default setting for afeature/option.The symbols listed below are used in this manual to notify the reader of key issues or proceduresthat must be observed when using the scanner:NOTENotes contain information necessary for properlydiagnosing, repairing and operating thescanner.The CAUTION symbol advises you of actionsthat could damage equipment or property.CAUTIONProduct Reference Guide 1-1


Getting StartedConnecting the ScannerThe scanner kit you ordered to match your interface should provide a compatible cable for yourinstallation. Use the appropriate instructions below to connect the scanner to the terminal, PCor other host device.Upon completing the connection via the appropriate interface instructions below, proceed tothe Interface Related Features section of this manual and scan the bar code to select the correctinterface type.RS-232 Serial Connection — Turn off power to the terminal/PC and connect the scannerto the terminal/PC serial port via the RS-232 cable as shown in Figure 1. If the terminal will notsupport POT (Power Off the Terminal) to supply scanner power, use the approved power supply(AC Adapter). Plug the AC Adapter barrel connector into the socket on the RS-232 cable connectorand the AC Adapter plug into a standard power outlet.Figure 1. RS-232 Serial or USB Connection using A/C AdapterConnectHereI/F CableTo Host/TerminalA/C Adapter(if needed)For 220-230 VACadapters, thecord must befacing down asshown in theillustration. Ifinstalled upwards,it will pose anundue strain on thesocket outlet.USB Connection — Connect the scanner to a USB port on the terminal/PC using the correctUSB cable for the interface type you ordered. Reference Figure 1 and Figure 2.NOTEUSB installations may require a power connection viaan approved A/C Adapter as shown in Figure 1. Forexample, this would be the case if the scanner is connectedalong with a number of other devices to a nonpoweredUSB hub.Figure 2. Other Connection Typesor...or...WandIB MUS BK eyboardWedgeor...1-2 <strong>Magellan</strong> ® <strong>1000i</strong>


ProgrammingProgrammingWand Emulation Connection — Connect the scanner to the appropriated port on the terminal/PCusing the correct cable for the interface type you ordered. Reference Figure 2.IBM Connection — Connect the scanner to the IBM port on the terminal/PC using the correctIBM cable. Reference Figure 2.Keyboard Wedge Connection — Before connection, turn off power to the terminal/PC.The Keyboard Wedge cable has a ‘Y’ connection from the scanner. Connect the female to themale end from the keyboard and the remaining end at the keyboard port at the terminal/PC.Reference Figure 2.The scanner is typically factory-configured with a set of default features standard to the interfacetype you ordered. After scanning the interface bar code from the Interface Related Features section,you can select other options and customize your scanner through use of the instructionsand programming bar codes available in that section and also the Data Editing and Symbologieschapters of this manual.Using the Programming Bar CodesThis manual contains feature descriptions and bar codes which allow you to reconfigure yourscanner. Some programming bar code labels, like the label below for resetting defaults, requireonly the scan of that single label to enact the change. Most of the programming labels in thismanual, however, require the scanner to be placed in Programming Mode prior to scanningthem. Scan a START/END bar code once to enterProgramming Mode. Once the scanner is inProgramming Mode, you can scan a number of parameter settings before scanning the START/END bar code a second time, which will then accept your changes, exit Programming Modeand return the scanner to normal operation.Resetting the Standard Product DefaultsIf you are unsure of what programming options are in your scanner, or you’ve changed someoptions and want the factory settings restored, scan the Standard Product Default Settings barcode below. This will copy the factory configuration for the currently active interface to the currentconfiguration.Standard Product Default SettingsThe programming section lists the factory default settings for each of the menu commands forthe standard RS-232 interface in BOLD text on the following pages. Exceptions to default settingsfor the other interfaces can be found in Appendix D, Default Settings.Product Reference Guide 1-3


Getting StartedLED and Beeper IndicatorsGreen LED IndicationsThe scanner’s beeper sounds and its green LED illuminates to indicate various functionsor errors on the scanner. The tables below list these indications. One exception to thebehaviors listed in the tables is that the scanner’s functions are programmable, and mayor may not be turned on. For example, certain indications, such as the power-up beepcan be disabled using programming bar code labels.LEDINDICATIONINDICATIONCOMMENTPower-onindicationGood ReadIndicationBright green flashBright green flashIndicates the scanner has finished all its power up tests and isnow ready for ooperation.Indicates a bar code has been read and decoded.Scanner Ready Constant dim green The scanner is ready for operation.Sleep ModeHost DisableDiagnosticsBEEPER FUNCTIONSConstant green flash(100mS on, 1900mS off)Constant green flash at 1 Hz(100mS on, 900mS off)Varies (see "Error Codes" onpage 1-5 for more information)The scanner is in Sleep Mode. To wake the scanner up, movean object in front of its window or press the button atop theunit.The scanner is disabled due to receiving a disble commandfrom the POS terminal.The LED can provide diagnostic feedback if the scanner discoversa problem during SelfTest.BEEPERINDICATIONPower On BeepGood ReadIndicationDiagnosticsProgrammingModeIndicationsSingle beepSingle beepINDICATIONVaries (see "Error Codes" onpage 1-5 for more information)Varies (see the Product ReferenceGuide for more information aboutscanner programming)COMMENTThe Power-On LED indication is a configurable feature whichcan be enabled or disabled. When enabled, this beep Indicatesthe scanner has finished all its power up tests and isnow ready for operation.The good read beep indication is configurable. Optionsinclude: Enable/disable, frequency, duration and volume. Seethe Product Reference Guide (PRG) for more information.The Beeper can provide diagnostic feedback if the scannerdiscovers a problem during SelfTest.The Beeper will sound as programming bar code labels arescanned, indicating progress during scanner configuration.1-4 <strong>Magellan</strong> ® <strong>1000i</strong>


Error CodesError CodesUpon startup, if the scanner flashes its indicator LED or sounds an unexpected series of beeptones (other than normal power-up indications), this means the scanner has not passed its automaticSelftest and has entered FRU 1 isolation mode. If the scanner is reset or the trigger ispulled, the sequence will be repeated. The following table describes the LED flashes/beep codesassociated with an error found.NUMBER OFLED FLASHES/BEEPSERRORCORRECTIVE ACTION1 Configuration2 Interface PCB6 Main PCB10 Button Error12 Imager Module1314Software IDFailureCPLD/CodeMismatchContact Helpdesk for assistance1. Field Replaceable Unit (FRU)Product Reference Guide 1-5


Getting StartedNOTES1-6 <strong>Magellan</strong> ® <strong>1000i</strong>


Chapter 2General FeaturesDouble Read TimeoutThe Double Read Timeout feature sets a time limit that determines how much time must passbefore reading the same label again (e.g. two identical items in succession).START / ENDPROGRAMMING BARCODES0.1 Second0.2 Second0.3 Second0.4 SecondDEFAULT0.5 Second0.6 SecondProduct Reference Guide 2-1


General FeaturesDouble Read Timeout — cont.START / ENDPROGRAMMING BARCODES0.7 Second0.8 Second0.9 Second1 Second2-2 <strong>Magellan</strong> ® <strong>1000i</strong>


Label Gone TimeoutLabel Gone TimeoutThis feature sets the time after the last label segment is seen before the scanner prepares for anew label.START / ENDPROGRAMMING BARCODESSets the label gone timeout duration using hex values from 000 to 255 in increments of ten milliseconds (10ms or 0.01 seconds). Toconfigure this feature, scan the “START/END” bar code above to place the unit in Programming Mode, then the “Set Label Gone Timeout,”followed by the three digits (zero padded) from the Alphanumeric table in Appendix C, Alpha-Numeric Pad representing thedesired time value. Exit programming mode by scanning the “START/END” bar code again.DEFAULT SETTING FOR THIS FEATURE: 320 milliseconds (032)Set Label Gone TimeoutProductivity Index Reporting (PIR)When PIR is enabled, label quality data is appended to decoded data before being presented tothe POS. The PIR feature allows the scanner to provide information to an external computer,indicating how easy the label was to read.NOTEThis value-added feature is a factory-programmedoption. Contact your dealer for information aboutupgrading your system to include this advancedcapability.START / ENDPROGRAMMING BARCODESDisableDEFAULTEnableProduct Reference Guide 2-3


General FeaturesSleep ModeThis feature specifies the amount of time with no bar code reads before the scanner enters sleep mode.START / ENDPROGRAMMING BARCODES15 Seconds30 Seconds1 Minute2 Minutes3 Minutes4 Minutes5 Minutes2-4 <strong>Magellan</strong> ® <strong>1000i</strong>


Sleep ModeSleep Mode — cont.START / ENDPROGRAMMING BARCODES6 Minutes7 Minutes8 Minutes9 Minutes10 MinutesDEFAULT12 Minutes15 Minutes30 Minutes1 HourProduct Reference Guide 2-5


General FeaturesLED and Beeper IndicatorsPower On AlertDisables or enables the indication (a single beep) that the scanner has finished all its power uptests and is now ready for operation.START / ENDPROGRAMMING BARCODESDisableEnableDEFAULT2-6 <strong>Magellan</strong> ® <strong>1000i</strong>


LED and Beeper IndicatorsGood Read: When to IndicateThis feature specifies when the scanner will provide indication (beep and/or flash its green LED)upon successfully reading a bar code. Choices are:• Good Read = Indicate after decode• Good Read = Indicate after transmit• Good Read = Indicate after CTS goes inactive, then activeNOTEThis option (Indicate after CTS goes inactive, thenactive), which uses CTS, is only valid for RS-232 interfaces.START / ENDPROGRAMMING BARCODESAfter DecodeDEFAULTAfter TransmitAfter CTS goes inactive, then activeProduct Reference Guide 2-7


General FeaturesGood Read Beep ControlThis feature enables/disables the scanner’s ability to beep upon a successful decode of a bar code.START / ENDPROGRAMMING BARCODESDisableEnableDEFAULTGood Read Beep FrequencyAdjusts the good read beep to sound at a selectable low, medium or high frequency, selectablefrom the list below. (Controls the beeper’s pitch/tone.)START / ENDPROGRAMMING BARCODESLowMediumDEFAULTHigh³ 0 0 1 4 0 22-8 <strong>Magellan</strong> ® <strong>1000i</strong>


LED and Beeper IndicatorsGood Read Beep LengthSpecifies the duration of a good read beep.START / ENDPROGRAMMING BARCODES60msec80msecDEFAULT100msec120msec140msec160msec180msec200msecProduct Reference Guide 2-9


General FeaturesGood Read Beep VolumeSelects the beeper volume (loudness) upon a good read beep. There are three selectable volumelevels.START / ENDPROGRAMMING BARCODESLowMediumHighDEFAULT2-10 <strong>Magellan</strong> ® <strong>1000i</strong>


General FeaturesTarget Mode Linger TimeSpecifies the time duration the scanner remains in the targeted mode of operation after readinga bar code before reverting to Omni-directional Mode.START / ENDPROGRAMMING BARCODESShort DurationMedium DurationDEFAULTLong Duration2-12 <strong>Magellan</strong> ® <strong>1000i</strong>


Scanning FeaturesWake Up IntensityThis feature indicates the percentage of ambient light change which will trigger the scanner towake up from Sleep Mode. Lower settings provide greater sensitivity. The seelectable range forthis setting is 5% to 15%.START / ENDPROGRAMMING BARCODES5%6%7%8%9%10%DEFAULT11%12%Product Reference Guide 2-13


General FeaturesWake Up Intensity — cont.START / ENDPROGRAMMING BARCODES13%14%15%2-14 <strong>Magellan</strong> ® <strong>1000i</strong>


Chapter 3Interface Related FeaturesAt the time of this writing, the Scanner supports the interfaces listed in Table 3-1. Select thedesired interface type from the table, then reference the page number given for the customizablefeatures section associated with each interface. See Table 3-2 for a description of each KeyboardWedge interface type (A through Y as listed).Table 3-1. Interfaces SupportedRS-232 Page Keyboard Wedge PageRS-232 Standard 3-9 Keyboard Wedge H aRS-232 Wincor-Nixdorf 3-9 Keyboard Wedge I a 3-30IBM Keyboard Wedge J a 3-30IBM 4683 Port 5B 3-24 Keyboard Wedge K a 3-30IBM 4683 Port 9B 3-24 Keyboard Wedge L a 3-30IBM 4683 Port 17 3-24 Keyboard Wedge M a 3-30USB Keyboard Wedge N a 3-30IBM USB 3-24 Keyboard Wedge N a 3-30USB Keyboard 3-24 Keyboard Wedge O a 3-30Wand Emulation Keyboard Wedge P a 3-30Wand Emulation 3-26 Keyboard Wedge Q a 3-30Keyboard Wedge 3-30 Keyboard Wedge R a 3-30Keyboard Wedge A a3-30 Keyboard Wedge S a 3-30Keyboard Wedge B a 3-30 Keyboard Wedge T a 3-30Keyboard Wedge C a 3-30 Keyboard Wedge U a 3-30Keyboard Wedge D a 3-30 Keyboard Wedge V a 3-30Keyboard Wedge E a 3-30 Keyboard Wedge W a 3-30Keyboard Wedge F a 3-30 Keyboard Wedge X a 3-30Keyboard Wedge G a 3-30 Keyboard Wedge Y a 3-30a. Consult Table for more information regarding keyboard interface types.NOTEThe correct interface cable is included for the scannerinterface type you ordered.Product Reference Guide 3-1


Interface Related FeaturesTable 3-2. Keyboard Wedge Interface ReferenceI/F TypePCs SupportedA PC/XT w/Alternate Key EncodingAT, PS/2 25-286, 30-286, 50, 50Z, 60, 70, 80, 90 & 95 w/Alternate KeyBEncodingC PS/2 25 and 30 w/Alternate Key EncodingD PC/XT w/Standard Key EncodingAT, PS/2 25-286, 30-286, 50, 50Z, 60, 70, 80, 90 & 95 w/Standard KeyEEncodingF PS/2 25 and 30 w/Standard Key EncodingG IBM 3xxx w/122 keyboardH IBM 3xxx w/102 keyboardI PS/55 5530T w/104 keyboardJ NEC 9801K WYSE 30/30+ WY-30 Keyboard 83 KeysWYSE 60/85/99 GT/150/160/285 Style IBM Enhanced PC, 520/520ESL Style IBM Enhanced PC FRWYSE 55/65/65 ES/120/185/325 Style IBM Enhanced PCWYSE 60/85/99 GT/150/160/285 ANSI Keyboard 105 Keys, 520/520 ESM ANSI Keyboard 105 KeysWYSE 55/65/65 ES/120/185/325 ANSI Keyboard 105 KeysWYSE 60/85/99 GT/150/160/285 ASCII Kbd, 520/520 ES ASCII KbdNWYSE 55/65/65 ES/120/185/325 ASCII KeyboardWYSE 60/85/99 GT/150/160/285 ANSI W285 Keyboard 105 Keys, 520/O 520 ES ANSI W285 Keyboard 105 KeysWYSE 55/65/65 ES/120/185/325 ANSI W285 Keyboard 105 KeysP WYSE WINTERM 3320 SEIBM 3153QIBM 316X, 3179/3180/319X/3270R IBM 3151/3152-010, 347X/348XS DIGITAL VT 220/320/330/340/350/382T DIGITAL VT420U DIGITAL VT 510/520 IBM ANSI Style KeyboardV DIGITAL VT 510/520 IBM PC Style KeyboardW SUN SPARC 5/10X SUN 420/440, ITXY WYSE 370/355 Style Enhanced IBM PCNOTEReference Appendix E, Keyboard Function Key Mappingsfor more information about keyboards.3-2 <strong>Magellan</strong> ® <strong>1000i</strong>


Interface SelectionInterface SelectionSTART / ENDPROGRAMMING BARCODESRS-232 StandardRS-232 Wincor-NixdorfIBM 4683 Port 5BIBM 4683 Port 9BIBM 4683 Port 17IBM USBUSB KeyboardWand EmulationKeyboard Wedge AProduct Reference Guide 3-3


Interface Related FeaturesInterface Selection — cont.START / ENDPROGRAMMING BARCODESKeyboard Wedge BKeyboard Wedge CKeyboard Wedge DKeyboard Wedge EKeyboard Wedge FKeyboard Wedge GKeyboard Wedge H³ 0 0 0 1 3 2Keyboard Wedge IKeyboard Wedge J3-4 <strong>Magellan</strong> ® <strong>1000i</strong>


Interface SelectionInterface Selection — cont.START / ENDPROGRAMMING BARCODESKeyboard Wedge KKeyboard Wedge LKeyboard Wedge MKeyboard Wedge NKeyboard Wedge OKeyboard Wedge P³ 0 0 0 1 3 BKeyboard Wedge QKeyboard Wedge RKeyboard Wedge SProduct Reference Guide 3-5


Interface Related FeaturesInterface Selection — cont.START / ENDPROGRAMMING BARCODESKeyboard Wedge TKeyboard Wedge UKeyboard Wedge VKeyboard Wedge WKeyboard Wedge XKeyboard Wedge Y3-6 <strong>Magellan</strong> ® <strong>1000i</strong>


Interface FeaturesInterface FeaturesObey/Ignore Host CommandsWhen set to ignore host commands, the scanner will ignore all host commands except for theminimum set necessary to keep the interface active and transmit labels For normal operation ofthe interface, select Obey Host Commands.START / ENDPROGRAMMING BARCODESObey Host CommandsDEFAULTIgnore Host CommandsProduct Reference Guide 3-7


Interface Related FeaturesInterface Features — cont.Host Transmission BuffersSpecifies the number of host transmission(s) that may be buffered. By buffering data from a barcode, the scanner can continue to read a new bar code while the old one is being transmitted tothe host. Selecting BUFFERS = 1 means that the first bar code must be transmitted before anew one can be read. A selection of BUFFERS = 2 means that a new bar code can be read whiledata from the first bar code is transmitted.When a DISABLE SCANNER command is received from the host, the scanner will continue totransmit all data that is buffered.START / ENDPROGRAMMING BARCODESHost Transmission Buffers = 1Host Transmission Buffers = 2DEFAULT3-8 <strong>Magellan</strong> ® <strong>1000i</strong>


Interface FeaturesRS-232 Interface FeaturesSTART / ENDPROGRAMMING BARCODES1200 Baud2400 Baud4800 Baud9600 BaudDEFAULT19200 Baud38400 Baud57600 Baud115200 BaudProduct Reference Guide 3-9


Interface Related FeaturesRS-232 Interface Features — cont.START / ENDPROGRAMMING BARCODES7 Data Bits8 Data BitsDEFAULT1 Stop BitDEFAULT2 Stop BitsParity = NoneDEFAULTParity = EvenParity = Odd3-10 <strong>Magellan</strong> ® <strong>1000i</strong>


Interface FeaturesRS-232 Interface Features — cont.Hardware Flow ControlDisable Hardware Control— The scanner transmits to the host regardless of any activity onthe CTS line.Enable CTS Flow Control— The CTS signal controls transmission of data to the host.Enable CTS Scan Control— The CTS line must be active for the scanner to read and transmitdata. While the CTS line is inactive, the scanner remains in a host-disabled state; followinga successful label transmission, the CTS signal must transition to inactive and then to active toenable scanning for the next label.START / ENDPROGRAMMING BARCODESDisable Hardware ControlDEFAULTEnable CTS Flow ControlEnable CTS Scan ControlProduct Reference Guide 3-11


Interface Related FeaturesRS-232 Interface Features — cont.Intercharacter DelayThis delay is inserted after each data character transmitted. If the transmission speed is too high,the system may not be able to receive all characters. You may need to adjust the delay to makethe system work properly.START / ENDPROGRAMMING BARCODESInter-Char Delay = No DelayDEFAULTInterchar Delay = 10 msecInterchar Delay = 20 msecInterchar Delay = 30 msecInterchar Delay = 40 msecInterchar Delay = 50 msecInterchar Delay = 60 msecInterchar Delay = 70 msec3-12 <strong>Magellan</strong> ® <strong>1000i</strong>


Interface FeaturesIntercharacter Delay — cont.START / ENDPROGRAMMING BARCODESInterchar Delay = 80 msecInterchar Delay = 90 msecSoftware Flow ControlDisables/Enables software control using XON/XOFF characters.START / ENDPROGRAMMING BARCODESDisable Software Flow ControlDEFAULTEnable Software Flow ControlProduct Reference Guide 3-13


Interface Related FeaturesRS-232 Interface Features — cont.Host EchoWhen enabled, this feature passes all data through the scanner to the host as it comes in. Thisfeature is used for applications where “daisy chaining” of RS-232 devices onto the same cable isnecessary. If, for example, one of the devices in the chain is a terminal where someone is enteringdata while another person is simultaneously scanning a bar code requiring transmission to thehost, the scanner will wait for the RS-232 channel to be quiet for a specified period of time (setvia RS-232 Host Echo Quiet Interval). The scanner can be set to observe this delay before sending itsdata in order to avoid RS-232 transmission conflicts.START / ENDPROGRAMMING BARCODESDisable Host EchoDEFAULTEnable Host Echo3-14 <strong>Magellan</strong> ® <strong>1000i</strong>


Interface FeaturesRS-232 Interface Features — cont.Host Echo Quiet IntervalThis setting specifies the time interval of RS-232 channel inactivity which must transpire beforethe scanner will break the host echo loop to transmit the bar code data that has just beenscanned to the host.START / ENDPROGRAMMING BARCODESHost Echo Quiet Interval = 0msecHost Echo Quiet Interval = 10msecDEFAULTHost Echo Quiet Interval = 20msecHost Echo Quiet Interval = 30msecHost Echo Quiet Interval = 40msecHost Echo Quiet Interval = 50msecHost Echo Quiet Interval = 60msecHost Echo Quiet Interval = 70msecProduct Reference Guide 3-15


Interface Related FeaturesHost Echo Quiet Interval — cont.START / ENDPROGRAMMING BARCODESHost Echo Quiet Interval = 80msecHost Echo Quiet Interval = 90msecHost Echo Quiet Interval = 100msecSignal Voltage: Normal/TTLSpecifies whether the RS-232 interface provides TTL levels on the output pins TxD and RTS.START / ENDPROGRAMMING BARCODESSignal Voltage: Normal RS-232DEFAULTSignal Voltage: TTL³3-16 <strong>Magellan</strong> ® <strong>1000i</strong>


Interface FeaturesRS-232 InvertEnables/disables inversion of RS-232 TXD and RXD signals.START / ENDPROGRAMMING BARCODESDisable RS-232 InvertDEFAULTEnable RS-232 InvertBeep on ASCII BELEnables/disables ability of scanner to beep (sound a good read tone) on receiving an ASCII BEL(07 hex).START / ENDPROGRAMMING BARCODESEnable Beep on ASCII BELDEFAULTDisable Beep on ASCII BELProduct Reference Guide 3-17


Interface Related FeaturesBeep on Not on FileSelect for the host to beep (or not) when a not-on-file (host command) condition is detected bythe host.START / ENDPROGRAMMING BARCODESDisable Beep on Not On FileEnable Beep on Not On FileDEFAULT3-18 <strong>Magellan</strong> ® <strong>1000i</strong>


Interface FeaturesACK NAK OptionsThis enables/disables the ability of the scanner to support the RS-232 ACK/NAK protocol.When configured, the scanner and/or host sends an “ACK” when it receives data properly, andsends “NAK” when the data is in error. Selections for this option are:• Disable• Enable for label transmission — the scanner expects an ACK/NAK response from the hostwhen a label is sent• Enable for host-command acknowledge — the scanner will respond with ACK/NAKwhen the host sends a command• Enable for label transmission and host-command acknowledgeSTART / ENDPROGRAMMING BARCODESDisable ACK NAKDEFAULTEnable ACK NAK for TransmissionEnable ACK NAK for host command acknowledgeEnable ACK NAK for transmission and host commandProduct Reference Guide 3-19


Interface Related FeaturesRS-232 Interface Features — cont. — cont.ACK CharacterSTART / ENDPROGRAMMING BARCODESSets the ACK character from the set of ASCII characters or any decimal value from 000 to 255. Pad entries of less than three digitswith zeros, as in “005”. To configure this feature, scan the “START/END” bar code above to place the unit in Programming Mode,then the “Set ACK Character,” followed by the digits from the Alphanumeric table in Appendix C, Alpha-Numeric Pad representingyour desired character. Exit programming mode by again scanning the “START/END” barcode above.DEFAULT SETTING FOR THIS FEATURE: 006Set ACK CharacterNAK CharacterSTART / ENDPROGRAMMING BARCODESSets the NAK character from the set of ASCII characters or any decimal value from 000 to 255. Pad entries of less than three digitswith zeros, as in “005”. To configure this feature, scan the “START/END” bar code above to place the unit in Programming Mode,then the “Set NAK Character,” followed by the digits from the Alphanumeric table in Appendix C, Alpha-Numeric Pad representingyour desired character. Exit programming mode by again scanning the “START/END” barcode above.DEFAULT SETTING FOR THIS FEATURE: 021Set NAK Character3-20 <strong>Magellan</strong> ® <strong>1000i</strong>


Interface FeaturesRS-232 Interface Features — cont.Retry on ACK NAK TimeoutEnables/disables retry after the configurable ACK NAK Timeout Value (set in the following feature)has expired.START / ENDPROGRAMMING BARCODESDisable Retry on ACK NAK TimeoutEnable Retry on ACK NAK TimeoutDEFAULTACK NAK Timeout ValueSTART / ENDPROGRAMMING BARCODESThis item specifies the time the scanner will wait for an ACK character from the host following a label transmission.000 = Infinite timeout001 - 075 = Timeout in 200-millisecond incrementsTo configure this feature, scan the “START/END” bar code above to place the unit in Programming Mode, then the “Set ACK NAKTimeout Value,” followed by the three digits (zero padded) from the Alphanumeric table in Appendix C, Alpha-Numeric Pad representingyour desired value. Exit programming mode by again scanning the “START/END” barcode above.DEFAULT SETTING FOR THIS FEATURE: 001 (200 msec)Set ACK NAK Timeout ValueProduct Reference Guide 3-21


Interface Related FeaturesRS-232 Interface Features — cont.ACK NAK Retry CountSTART / ENDPROGRAMMING BARCODESThis feature sets the number of times for the scanner to retry a label transmission under a retry condition.000 = No retry001 - 254 = Retry for the specified number of times255 = Retry foreverTo configure this feature, scan the “START/END” bar code above to place the unit in Programming Mode, then the “Set ACK NAKRetry Count,” followed by the three digits (zero padded) from the Alphanumeric table in Appendix C, Alpha-Numeric Pad representingyour desired retry count. Exit programming mode by again scanning the “START/END” barcode aboveDEFAULT SETTING FOR THIS FEATURE: 003Set ACK NAK Timeout Value3-22 <strong>Magellan</strong> ® <strong>1000i</strong>


Interface FeaturesACK NAK Error HandlingThis item specifies the method the scanner will use to handle errors detected while waiting toreceive the ACK character from the host. Errors include unrecognized host commands and communicationerrors such as parity or framing errors. Choices are:00 = Ignore errors detected (recommended setting)01 = Process error as valid ACK character (risk of lost label data)02 = Process error as valid NAK character (risk of duplicate label data)START / ENDPROGRAMMING BARCODESIgnore Errors DetectedDEFAULTProcess error as valid ACK characterProcess error as valid NAK characterProduct Reference Guide 3-23


Interface Related FeaturesRS-232 Interface Features — cont.Transmission Failure IndicationEnables/disables bad-label indication upon transmission failure.START / ENDPROGRAMMING BARCODESDisable Transmission Error IndicationEnable Transmission Error IndicationDEFAULTIBM-USB Interface FeaturesIBM-USB Device usageThe IBM-USB protocol allows for the scanner to be identified as one of two different types ofbar code scanners. Depending on what other scanners you may already have connected to aIBM-USB POS, you may need to change this setting to enable all devices to communicate.Options are:• Table Top Scanner• Handheld ScannerSTART / ENDPROGRAMMING BARCODESConfigure as Table Top ScannerConfigure as Handheld ScannerDEFAULT3-24 <strong>Magellan</strong> ® <strong>1000i</strong>


Interface FeaturesIBMIBM Transmit Labels in Code 39 FormatThis feature enables/disables scanner's ability to set a symbology identifier for a specified label toCode 39 before transmitting that label data to an IBM host. This applies to: Code 128, Codabarand Code 93 for IBM USB; Code 128, Codabar and Code 93 for IBM Port 5B; and Codabarand Code 93 for IBM Port 9B.START / ENDPROGRAMMING BARCODESDisable Convert to Code 39DEFAULTEnable Convert to Code 39Product Reference Guide 3-25


Interface Related FeaturesWand EmulationSupported SymbologiesThe Wand Emulation interface will transmit bar code data as a wand device would. This interfacewill transmit the following bar code symbologies:• UPC/EAN• UPC/EAN with addons• Code 39• Full ASCII Code 39• Interleaved 2 of 5• Codabar• Code 128Pharmacode 39 is transmitted as Code 39. All other bar code symbology types read by the scannerwill be transmitted as Code 128.Wand Emulation Bar Code FormatThe following format settings are required for the wand emulation interface. These settings havebeen pre-configured at the factory for Wand Emulation scanners.• UPC-A bar codes must include all 12 digits.• UPC-E bar codes must contain 8 digits, including a system digit, 6 data digits, and thecheck digit.• EAN-13 bar codes must have all 13 digits.• EAN-8 bar codes must include all 8 digits.• Code 39, Code 39 Full ASCII, and Pharmacode 39 bar codes must NOT contain start /stop characters.• Codabar bar codes must include the start / stop characters, presented in the ABCD format.• Interleaved 2 of 5 bar codes must have an even number of digits.3-26 <strong>Magellan</strong> ® <strong>1000i</strong>


Interface FeaturesWand Emulation — cont.Bar/Space PolarityLow/High — Black will be transmitted as a low voltage level (0 to +0.7V) and space as highlevel (+2.4 to +5.25V).High/Low — Black will be transmitted as a high voltage level (+2.4 to +5.25V) and space aslow level (0 to +0.7V).START / ENDPROGRAMMING BARCODESBar/Space = Low/HighBar/Space = High/LowDEFAULTWand Idle StateThis feature specifies the level of the wand output signal when idle. TTL logic levels:High voltage level (+2.4 to +5.25V)Low voltage level (0 to +0.7V).START / ENDPROGRAMMING BARCODESWand Idle State = LowDEFAULTWand Idle State = HighProduct Reference Guide 3-27


Interface Related FeaturesWand Emulation — cont.Signal SpeedThe speed of the transmission can be set. This selects the width of the minimum narrow bar.330 microseconds660 microsecondsSTART / ENDPROGRAMMING BARCODESSignal Speed = 330mSSignal Speed = 660mSDEFAULTTransmit Trailing NoiseThe transmission of noise pulses after the label may be enabled or disabled.START / ENDPROGRAMMING BARCODESDisable Trailing NoiseEnable Trailing NoiseDEFAULT3-28 <strong>Magellan</strong> ® <strong>1000i</strong>


Interface FeaturesWand Emulation — cont.Transmit Leading NoiseThe transmission of noise pulses before the bar code may be enabled or disabled.START / ENDPROGRAMMING BARCODESDisable Leading NoiseEnable Leading NoiseDEFAULTSymbology ConversionWand Emulation can convert all bar codes to a single symbology. Choices are:No ConversionConvert to Code 39Convert to Code 128START / ENDPROGRAMMING BARCODESNo Symbology ConversionDEFAULTConvert to C39Convert to C128Product Reference Guide 3-29


Interface Related FeaturesKeyboard WedgeandUSB KeyboardAs a keyboard interface, the scanner supports most popular PCs and IBM terminals. The installationof the wedge is a fairly simple process that doesn’t require any changes of software or hardware.NOTEAll of the options in this section apply to the KeyboardWedge, however, only some apply to USB Keyboard.Keyboard LayoutThe Keyboard Layout option supports many countries. For details about Keyboard Layout,please refer to your operating system manual.START / ENDPROGRAMMING BARCODESUSADEFAULTBelgiumBritainDenmarkFranceGermany3-30 <strong>Magellan</strong> ® <strong>1000i</strong>


Interface FeaturesKeyboard Wedge — cont.START / ENDPROGRAMMING BARCODESItalyNorwayPortugalSpainSwedenSwitzerlandJapan 106 KeyHungaryCzechProduct Reference Guide 3-31


Interface Related FeaturesKeyboard Wedge — cont.START / ENDPROGRAMMING BARCODESSlovakiaRomaniaCroatiaPolandCaps Lock StateSpecifies the format in which the scanner sends character data.START / ENDPROGRAMMING BARCODESDisable Caps LockDEFAULTCaps Lock “ON”Shift Lock “ON”3-32 <strong>Magellan</strong> ® <strong>1000i</strong>


Interface FeaturesKeyboard Wedge — cont.Power-On SimulationNOTEThis feature does not apply to the USB Keyboardinterface.All PCs check the keyboard status during the power-on Selftest. It is recommended that youenable this function if you are working without a keyboard installation. It simulates keyboardtiming and passes the keyboard status to the PC during power-on.START / ENDPROGRAMMING BARCODESDisable Power-on SimulationDEFAULT³Enable Power-on SimulationProduct Reference Guide 3-33


Interface Related FeaturesControl CharactersSpecifies how the scanner transmits ASCII control characters to the host. Choices are:• Disable Control Characters• Enable transmission of control characters to host• Send characters between 00H and 1FH according to a special function-key mappingtable. (This is used to send keys that are not in the normal ASCII set; a unique set is providedfor each available scancode set. Reference Appendix E, Keyboard Function Key Mappings.)START / ENDPROGRAMMING BARCODESDisable Control CharactersDEFAULTEnable Transmission of Control CharactersEnable Function Key Mapping3-34 <strong>Magellan</strong> ® <strong>1000i</strong>


Interface FeaturesKeyboard Wedge — cont.Wedge Quiet IntervalNOTEThis feature does not apply to the USB Keyboardinterface.Quiet Interval is the amount of time to look for keyboard activity before the scanner breaks thekeyboard connection in order to transmit data to the host.START / ENDPROGRAMMING BARCODESSelectable from 000 (no interval) to 255 in 10 msec increments. To configure this feature, scan the “START/END” bar code above toplace the unit in Programming Mode, then the Set Wedge Quiet Interval bar code followed by the three digits (zero padded) fromthe Alphanumeric table in Appendix C, Alpha-Numeric Pad representing your desired length. Exit programming mode by againscanning the “START/END” barcode above.DEFAULT SETTING FOR THIS FEATURE:010 (100 msec)Set Wedge Quiet IntervalProduct Reference Guide 3-35


Interface Related FeaturesKeyboard Wedge — cont.Intercharacter DelayNOTEThis feature does not apply to the USB Keyboard interface.START / ENDPROGRAMMING BARCODESOne-half of the delay specified below is inserted between scancodes within each character. If the transmission speed is too high, thesystem may not be able to receive all characters. You may need to adjust the delay to make the system work properly. Selectablefrom 000 to 255 in 10msec increments.To configure this feature, scan the “START/END” bar code above to place the unit in Programming Mode, then the “Set IntercharacterDelay,” followed by the three digits (zero padded) from the Alphanumeric table in Appendix C, Alpha-Numeric Pad representingyour desired length. Exit programming mode by again scanning the “START/END” barcode above/DEFAULT SETTING FOR THIS FEATURE:000 (No Delay)Set Intercharacter Delay3-36 <strong>Magellan</strong> ® <strong>1000i</strong>


Chapter 4Data EditingData Editing OverviewCAUTIONIt is not recommended to use these features with IBMor Wand Emulation interfaces.When a bar code is scanned, additional information can be sent to the host computer alongwith the bar code data. This combination of bar code data and supplementary user-defineddata is called a “message string.” The features in this chapter can be used to build specificuser-defined data into a message string.There are several types of selectable data characters that can be sent before and after scanneddata. You can specify if they should be sent with all symbologies, or only with specific symbologies.Figure 4-1 shows the available elements you can add to a message string:Figure 4-1. Breakdown of a Message StringLabel ID Transmission:Enable this option totransmit the Label ID youconfigure for the scannedsymbology.OR...PrefixLabel IDAIM IDBar Code DataLabel IDSuffixAIM ID: This function is used to identify and display thecommon label identifier for its symbology. When enabled,this ID code will be transmitted before the scanned barcode data.00 - 20 Characters (ASCII)Please Keep In Mind...• Modifying a message string is not a mandatory requirement. Data editing is sophisticatedfeature allowing highly customizable output for advanced users. Factory default settingsfor data editing is typically set to NONE.• A prefix or suffix may be applied (reference the Symbologies chapter for these settings)across all symbologies (set via the Global features in this chapter).• You can add any character from the ASCII Chart (from 00-FF) on the inside back cover ofthis manual as a prefix, suffix or Label ID.Product Reference Guide 4-1


Data Editing• Enter prefixes and suffixes in the order in which you want them to appear on the output.Global Prefix/SuffixUp to 20 ASCII characters may be added as a prefix (in a position before the bar code data) and/or as a suffix (in a position following the bar code data) as indicated in Figure 4-2.Figure 4-2. Prefix and Suffix PositionsOR...PrefixLabel IDAIM IDBar Code DataLabel IDSuffix00 - 20 Characters (ASCII)ANDORExample: Setting a PrefixIn this example, we’ll set a prefix for all symbologies.1. Determine which ASCII character(s) are to be added to scanned bar code data. In thisexample, we’ll add a dollar sign (‘$’) as a prefix.2. Scan the START bar code.3. Scan the SET PREFIX bar code.4. Reference the ASCII Chart on the inside back cover of this manual, to find the hex valueassigned to the desired character. The corresponding hex number for the ‘$’ character is24. To enter this selection code, scan the ‘2’ and ‘4’ bar codes from Appendix C, Alpha-Numeric Pad.5. Scan the END bar code to exit Programming Mode.NOTEIf less than the expected string of 20 characters areselected, scan the END bar code twice to accept theselections and exit Programming Mode.6. The resulting message string would appear as follows:Scanned bar code data:12345Resulting message string output: $123454-2 <strong>Magellan</strong> ® <strong>1000i</strong>


Global Prefix/SuffixGlobal Prefix/Suffix — cont.START / ENDPROGRAMMING BARCODESSets up to 20 characters each from the set of ASCII characters or any hex value from 0 to FF. To configure this feature, scan the“START/END” bar code above to place the unit in Programming Mode, then the “Set Prefix” or “Set Suffix,” followed by the digits fromthe Alphanumeric table in Appendix C, Alpha-Numeric Pad representing your desired character(s). Reference the section,"Example: Setting a Prefix", for more information. Exit programming mode by scanning the “START/END” barcode again (scan“START/END” twice if less than 20 characters have been selected).DEFAULT SETTING PREFIX: 00 (None)DEFAULT SETTING SUFFIX: 0D (CR)Set PrefixSet SuffixProduct Reference Guide 4-3


Data EditingAIM IDAIM (Automatic Identification Manufacturers) label identifiers are assigned from a globallystandardized list — as opposed to custom label ID characters you select yourself — and can beincluded with scanned bar code data. AIM label identifiers consist of three characters as follows:• A close brace character (ASCII ‘]’), followed by...• A code character (see the table below), followed by...• A modifier character (the modifier character is symbol dependent)SYMBOLOGY CHAR SYMBOLOGY CHARUPC/EAN E Code 128/EAN 128 CCode 39 A MSI/Plessey MCodabarFRSS (RSS-14, RSSExpanded)eInterleaved 2 of 5 I Standard 2 of 5 SCode 93 G ISBN X aa. ISBN (X with a 0 modifier character)Figure 4-3. AIM IDPrefixLabel IDAIM IDBar Code DataLabel IDSuffixSTART / ENDPROGRAMMING BARCODESDisable AIM IDDEFAULTEnable AIM ID4-4 <strong>Magellan</strong> ® <strong>1000i</strong>


Label IDLabel IDA Label ID is a customizable code of up to two ASCII characters (00-FF), used to identify a barcode (symbology) type. See Appendix D, Default Settings, for a listing for common symbologies. Itcan be appended previous to or following the transmitted bar code data depending upon howthis option is enabled. This feature provides options for configuring custom Label IDs individuallyper symbology. If you wish to program the scanner to always include an industry standardlabel identifier for ALL symbology types, see the previous feature, AIM ID.To configure a Label ID:1. Scan the START bar code.2. Select Label ID position as either BEFORE or AFTER by scanning the appropriate barcode.3. Scan a bar code to select the symbology for which you wish to configure a custom LabelID.4. Determine the desired character(s) (you may choose either one or two) which will representthe Label ID for the selected symbology. Next, turn to the ASCII Chart on the insideback cover of this manual and find the equivalent hex digits associated with your choice ofLabel ID. For example, if you wish to select an equal sign (=) as a Label ID, the chart indicatesits associated hex characters as 3D.5. Turn to Appendix C, Alpha-Numeric Pad and scan the bar codes representing the hex charactersdetermined in the previous step. For the example given, the characters ‘3’ and ‘D’would be scanned.6. Scan the END bar code to exit programming mode.Figure 4-4. Label ID Position OptionsPrefixLabel IDAim IDBar Code DataLabel IDSuffixORProduct Reference Guide 4-5


Data EditingLabel ID — cont.START / ENDPROGRAMMING BARCODESLabel ID Transmission: DisableLabel ID Position: Before Bar Code DataDEFAULTLabel ID Position: After Bar Code DataSet UPC-A Label ID Character(s)DEFAULT SETTING FOR THIS FEATURE: A (41 hex)Set UPC-A w/P2 Addon Label ID Character(s)DEFAULT SETTING FOR THIS FEATURE: A (41 hex)Set UPC-A w/P5 Addon Label ID Character(s)DEFAULT SETTING FOR THIS FEATURE: A (41 hex)Set UPC-A w/C128 Addon Label ID Character(s)DEFAULT SETTING FOR THIS FEATURE: A (41 hex)Set UPC-E Label ID Character(s)DEFAULT SETTING FOR THIS FEATURE: E (45 hex)4-6 <strong>Magellan</strong> ® <strong>1000i</strong>


Label IDLabel ID — cont.START / ENDPROGRAMMING BARCODESDEFAULT SETTING FOR THIS FEATURE: E (45 hex)Set UPC-E w/P2 Addon Label ID Character(s)Set UPC-E w/P5 Addon Label ID Character(s)DEFAULT SETTING FOR THIS FEATURE: E (45 hex)DEFAULT SETTING FOR THIS FEATURE: E (45 hex)Set UPC-E w/C128 Addon Label ID Character(s)Set EAN-8 Label ID Character(s)DEFAULT SETTING FOR THIS FEATURE: FF (4646 hex)DEFAULT SETTING FOR THIS FEATURE: FF (4646 hex)Set EAN-8 w/P2 Addon Label ID Character(s)Set EAN-8 w/P5 Addon Label ID Character(s)DEFAULT SETTING FOR THIS FEATURE: FF (4646 hex)DEFAULT SETTING FOR THIS FEATURE: FF (4646 hex)Set EAN-8 w/C128 Addon Label ID Character(s)Set EAN-13 Label ID Character(s)DEFAULT SETTING ³ 0FOR 0THIS 2 FEATURE: 7 F (46 hex)Product Reference Guide 4-7


Data EditingLabel ID — cont.START / ENDPROGRAMMING BARCODESDEFAULT SETTING FOR THIS FEATURE: F (46 hex)Set EAN-13 w/P2 Addon Label ID Character(s)Set EAN-13 w/P5 Addon Label ID Character(s)DEFAULT SETTING FOR THIS FEATURE: F (46 hex)DEFAULT SETTING FOR THIS FEATURE: F (46 hex)Set EAN-13 w/C128 Addon Label ID Character(s)Set ISBN Label ID Character(s)DEFAULT SETTING FOR THIS FEATURE: I (49 hex)DEFAULT SETTING ³ FOR 0 3THIS 4FEATURE: C IA (4941 hex)Set IATA Label ID Character(s)Set GTIN Label ID Character(s)DEFAULT SETTING FOR THIS FEATURE: G (47 hex)DEFAULT SETTING ³ FOR THIS FEATURE: G2 (4732 hex)Set GTIN w/P2 addon Label ID Character(s)Set GTIN w/P5 addon Label ID Character(s)DEFAULT SETTING ³ FOR 0 5THIS 6FEATURE: 2 G5 (4735 hex)4-8 <strong>Magellan</strong> ® <strong>1000i</strong>


Label IDLabel ID — cont.START / ENDPROGRAMMING BARCODESDEFAULT SETTING FOR THIS FEATURE: G8 (4738 hex)Set GTIN w/C128 addon Label ID Character(s)Set RSS-14 Label ID Character(s)DEFAULT SETTING FOR THIS FEATURE: R4 (5234 hex)DEFAULT SETTING ³ FOR 0 THIS 4 DFEATURE: B RX (5258 hex)Set RSS Expanded Label ID Character(s)Set Code 39 Label ID Character(s)DEFAULT SETTING FOR THIS FEATURE: * (2A hex)DEFAULT SETTING FOR THIS FEATURE: A (41 hex)Set Pharmacode 39 Label ID Character(s)Set Code 128 Label ID Character(s)DEFAULT SETTING FOR THIS FEATURE: # (23 hex)DEFAULT SETTING FOR THIS FEATURE: i (69 hex)Set I 2 of 5 Label ID Character(s)Product Reference Guide 4-9


Data EditingLabel ID — cont.START / ENDPROGRAMMING BARCODESSet Codabar Label ID Character(s)DEFAULT SETTING ³ 0FOR 0THIS 3 FEATURE: 7 % (25 hex)Set Code 93 Label ID Character(s)DEFAULT SETTING FOR THIS FEATURE: & (26 hex)Set Code 11 Label ID Character(s)DEFAULT SETTING FOR THIS FEATURE: @ (40 hex)Set MSI/Plessey Label ID Character(s)DEFAULT SETTING FOR THIS FEATURE: @ (40 hex)Set Std 2 of 5 Label ID Character(s)DEFAULT SETTING FOR THIS FEATURE: s (73 hex)Set EAN UCC Composite Label ID Character(s)DEFAULT SETTING FOR THIS FEATURE: 04-10 <strong>Magellan</strong> ® <strong>1000i</strong>


Case ConversionCase ConversionThis feature can convert scanned bar code data to either all lower case (a through z) or all uppercase (A through Z) characters.NOTECase conversion affects ONLY scanned bar code data,and does not affect Label ID, Prefix, Suffix, or otherappended data.START / ENDPROGRAMMING BARCODESDisableDEFAULTConvert to Upper CaseConvert to Lower CaseProduct Reference Guide 4-11


Data EditingCharacter ConversionCharacter conversion is an eight byte configuration item. The eight bytes are 4 character pairsrepresented in hexadecimal ASCII values. The first character in the pair is the character thatwill be converted. The second character in the pair is the character to convert to. If the characterto convert in a pair is FF, then no conversion is done.For example, if you have the character conversion configuration item set to the following:41423132FFFFFFFFThe first pair is 4142 or AB (41 hex is an ASCII capital A, 42 hex is an ASCII capital B) andthe second pair is 3132 or 12 (31 hex is an ASCII 1, 32 is an ASCII 2). The other two pairsare FFFF and FFFF.With the label, AG15TA81, it would look as follows after the character conversion:BG25TB82.The A characters were converted to the B character and the 1 characters were converted to thenumeral 2 character. Nothing is done with the last two character pairs, since they are all FF.To set Character Conversion:1. Scan the START/END bar code.2. Scan the Character Conversion bar code.3. Determine the desired string. Up to sixteen positions can be determined as in the aboveexample. Next, turn to the ASCII Chart on the inside back cover of this manual and findthe equivalent hex digits needed to fulfill the string.4. Turn to Appendix C, Alpha-Numeric Pad and scan the bar codes representing the hex charactersdetermined in the previous step.5. Scan the START/END bar code to exit Programming Mode.NOTEIf less than the expected string of 16 characters areselected, scan the START/END bar code twice to acceptthe selections and exit Programming Mode.START / ENDPROGRAMMING BARCODESDEFAULT SETTING FOR THIS FEATURE:FFFFFFFFFFFFFFFF hex (no conversion)Character Conversion4-12 <strong>Magellan</strong> ® <strong>1000i</strong>


Chapter 5SymbologiesThe scanner supports the following symbologies (bar code types). Options for each symbologyare included in this chapter.·UPC-A ·Pharmacode 39·UPC-E ·Code 128·EAN-13 ·Interleaved 2 of 5·EAN-8·Codabar·RSS-14 ·Code 93·RSS Expanded·MSI/Plessey·Code 39 ·Standard 2 of 5Factory Defaults — for the standard RS-232 interface are indicated in bold text throughoutthis section. Reference Appendix D, Default Settings for default exceptions for your interface.UPC-ADisable/Enable UPC-AWhen disabled, the scanner will not read UPC-A bar codes.START / ENDPROGRAMMING BARCODESDisable UPC-AEnable UPC-ADEFAULTProduct Reference Guide 5-1


SymbologiesUPC-A — continuedCheck Digit TransmissionEnable this option to transmit the check digit along with UPC-A bar code data.START / ENDPROGRAMMING BARCODESDon’t Send Check DigitSend Check DigitDEFAULTExpand UPC-A to EAN-13Expands UPC-A data to the EAN-13 data format. Selecting this feature also changes the symbologyID to match those required for EAN-13.START / ENDPROGRAMMING BARCODESDon’t Expand to EAN-13DEFAULTExpand to EAN-135-2 <strong>Magellan</strong> ® <strong>1000i</strong>


UPC-AUPC-A — continuedNumber System TransmissionThis feature enables/disables transmission of UPC-A System Number.START / ENDPROGRAMMING BARCODESDisable Number System TransmissionEnable Number System TransmissionDEFAULTUPC-A Minimum ReadsThis feature specifies the minimum number of consecutive times a UPC-A label must bedecoded before it is accepted as good read.START / ENDPROGRAMMING BARCODESMinimum = 1 ReadDEFAULTMinimum = 2 ReadsMinimum = 3 ReadsMinimum = 4 ReadsProduct Reference Guide 5-3


SymbologiesUPC-A — continuedUPC-A In-store Minimum ReadsThis feature specifies the minimum number of consecutive times an in-store printed label mustbe decoded before it is accepted as good read.START / ENDPROGRAMMING BARCODESMinimum = 1 ReadDEFAULTMinimum = 2 ReadsMinimum = 3 ReadsMinimum = 4 Reads5-4 <strong>Magellan</strong> ® <strong>1000i</strong>


UPC-EUPC-EThe following options apply to the UPC-E symbology.Disable/Enable UPC-EWhen disabled, the scanner will not read UPC-E bar codes.START / ENDPROGRAMMING BARCODESDisable UPC-EEnable UPC-EDEFAULTCheck Digit TransmissionEnable this option to transmit the check digit along with UPC-E bar code data.START / ENDPROGRAMMING BARCODESDon’t Send Check DigitSend Check DigitDEFAULTProduct Reference Guide 5-5


SymbologiesUPC-E — continuedNumber System DigitThe Number System Digit (NSD) which is always a zero (0) in the leading position can beoptionally included (or not) with scanned bar code data.START / ENDPROGRAMMING BARCODESExclude Number System DigitDEFAULTInclude Number System DigitExpand to UPC-E to UPC-AEnables/disables expansion of UPC-E labels to UPC-A. Selecting this feature also changes thesymbology ID to match those required for UPC-A.START / ENDPROGRAMMING BARCODESDon’t Expand UPC-E to UPC-ADEFAULTExpand UPC-E to UPC-A5-6 <strong>Magellan</strong> ® <strong>1000i</strong>


UPC-EUPC-E — continuedExpand UPC-E to EAN13Enables/disables expansion of UPC-E labels to EAN-13. Selecting this feature also changes thesymbology ID to match those required for EAN-13.START / ENDPROGRAMMING BARCODESDon’t Expand UPC-E to EAN-13DEFAULTExpand UPC-E to EAN-13Minimum ReadsThis feature specifies the minimum number of consecutive times a UPC-E label must bedecoded before it is accepted as good read.START / ENDPROGRAMMING BARCODESMinimum = 1 ReadMinimum = 2 ReadsDEFAULTMinimum = 3 ReadsMinimum = 4 ReadsProduct Reference Guide 5-7


SymbologiesGTINThe following options apply to the GTIN label data format.Expand UPC/EAN to GTINWhen this feature is enabled, the scanner will translate UPC/EAN labels to the 14 digit GTINformat.START / ENDPROGRAMMING BARCODESDon’t Expand to GTINDEFAULTExpand to GTIN5-8 <strong>Magellan</strong> ® <strong>1000i</strong>


EAN-13EAN-13The following options apply to the EAN-13 symbology.Disable/Enable EAN-13When disabled, the scanner will not read EAN-13 bar codes.START / ENDPROGRAMMING BARCODESDisable EAN-13Enable EAN-13DEFAULTCheck Digit TransmissionEnable this option to transmit the check digit along with EAN-13 bar code data.START / ENDPROGRAMMING BARCODESDon’t Send Check DigitSend Check DigitDEFAULTProduct Reference Guide 5-9


SymbologiesEAN-13 — continuedEAN-13 Flag 1 CharacterEnables/disables transmission of an EAN/JAN13 Flag1 character.START / ENDPROGRAMMING BARCODESDon’t Transmit EAN-13 Flag 1 CharTransmit EAN-13 Flag 1 CharDEFAULTISBNWhen enabled, this feature truncates the leading three digits from labels that contain ISBN(International Standard Book Number) and appends an ISBN check character to the end of thelabel. These codes are used for books and magazines. Labels with ISBN codes start with "978".Example:Bar code data: “9789572222720”Output: “9572222724”START / ENDPROGRAMMING BARCODESDisable ISBNDEFAULTEnable ISBN5-10 <strong>Magellan</strong> ® <strong>1000i</strong>


EAN-13EAN-13 — continuedMinimum ReadsThis feature specifies the minimum number of consecutive times an EAN-13 label must bedecoded before it is accepted as good read.START / ENDPROGRAMMING BARCODESMinimum = 1 ReadDEFAULTMinimum = 2 ReadsMinimum = 3 ReadsMinimum = 4 ReadsProduct Reference Guide 5-11


SymbologiesEAN-8The following options apply to the EAN-8 symbology.Disable/Enable EAN-8When disabled, the scanner will not read EAN-8 bar codes.START / ENDPROGRAMMING BARCODESDisable EAN-8Enable EAN-8DEFAULTCheck Digit TransmissionEnable this option to transmit the check Digit along with EAN-8 bar code data.START / ENDPROGRAMMING BARCODESDon’t Send Check DigitSend Check DigitDEFAULT5-12 <strong>Magellan</strong> ® <strong>1000i</strong>


EAN-8EAN-8 — continuedExpand EAN-8 to EAN-13 — Expands EAN-8 data to the EAN-13 data format. Selectingthis feature also changes the symbology ID to match those required for EAN-13.START / ENDPROGRAMMING BARCODESDon’t Expand to EAN-13DEFAULTExpand to EAN-13Minimum ReadsThis feature specifies the minimum number of consecutive times an EAN-8 label must bedecoded before it is accepted as good read.START / ENDPROGRAMMING BARCODESMinimum = 1 ReadDEFAULTMinimum = 2 ReadsMinimum = 3 ReadsMinimum = 4 ReadsProduct Reference Guide 5-13


SymbologiesEnable EAN Two-LabelEnables/disables the ability of the scanner to decode EAN two-label pairs.START / ENDPROGRAMMING BARCODESDisable EAN Two-LabelEnable EAN Two-Label5-14 <strong>Magellan</strong> ® <strong>1000i</strong>


Add-onsAdd-onsAdd-ons (or supplemental characters) are commonly added to the end of UPC/EAN bar codes.The scanner will read the add-ons if they are enabled and in the field of view. Three add-ontypes are supported: 2-digit, 5-digit and Code 128 add-ons. Supported options are:None — This option directs the scanner to ignore add-on portion of a UPC/EAN bar code butstill read the main portion of the bar code.2 Digits — The scanner will optionally read 2-digit add-ons with the UPC/EAN label.5 Digits — The scanner will optionally read 5-digit add-ons with the UPC/EAN label.Code 128 Add-on — The scanner will optionally read Code 128 add-ons with the UPC/EAN label.NOTEContact Customer Support foradvanced programming of optionaland conditional add-ons.Product Reference Guide 5-15


SymbologiesAdd-ons — continuedSTART / ENDPROGRAMMING BARCODESDisable Optional 2-Digit Add-onsDEFAULTEnable Optional 2-Digit Add-onsDisable Optional 5-Digit Add-onsDEFAULTEnable Optional 5-Digit Add-onsDisable Optional Code 128 Add-onsDEFAULTEnable Optional Code 128 Add-ons5-16 <strong>Magellan</strong> ® <strong>1000i</strong>


RSS-14RSS-14The following options apply to the RSS-14 symbology.Disable/Enable RSS-14When this feature is disabled, the scanner will not read RSS-14 bar codes.START / ENDPROGRAMMING BARCODESDisable RSS-14DEFAULTEnable RSS-14UCC/EAN 128 EmulationWhen enabled, RSS-14 bar codes will be translated to the UCC/EAN 128 label data format.START / ENDPROGRAMMING BARCODESDisable UCC/EAN 128 EmulationDEFAULTEnable UCC/EAN 128 EmulationProduct Reference Guide 5-17


SymbologiesRSS-14 — continuedMinimum ReadsThis feature specifies the minimum number of consecutive times an RSS-14 label must bedecoded before it is accepted as good read.START / ENDPROGRAMMING BARCODESMinimum = 1 ReadDEFAULTMinimum = 2 ReadsMinimum = 3 ReadsMinimum = 4 Reads5-18 <strong>Magellan</strong> ® <strong>1000i</strong>


RSS ExpandedRSS ExpandedThe following options apply to the RSS Expanded symbology.Disable/Enable RSS ExpandedWhen this feature is disabled, the scanner will not read RSS Expanded bar codes.START / ENDPROGRAMMING BARCODESDisable RSS ExpandedDEFAULTEnable RSS ExpandedUCC/EAN 128 EmulationWhen enabled, RSS Expanded bar codes will be translated to the UCC/EAN 128 label data format.START / ENDPROGRAMMING BARCODESDisable UCC/EAN 128 EmulationDEFAULTEnable UCC/EAN 128 EmulationProduct Reference Guide 5-19


SymbologiesRSS Expanded — continuedLength ControlFixed Length Decoding — When fixed length decoding is enabled, the scanner will decode abar code if the label length matches one of the configurable fixed lengths.Variable Length Decoding — When variable length decoding is enabled, the scanner willdecode a bar code if the label length falls in the range of the configurable minimum and maximumlength.Configuring Fixed Length Decoding:1. Scan the START/END bar code.2. Scan the Fixed Length Decoding bar code.3. Scan the START/END bar code.4. Set Length 1 to the first fixed length by following the RSS Expanded Length 1, Length 2Programming Instructions below.5. Set Length 2 to the second fixed length (or to ‘000’ if there is only one fixed length) byfollowing the RSS Expanded Length 1, Length 2 Programming Instructions below.Configuring Variable Length Decoding:1. Scan the START/END bar code.2. Scan the Variable Length Decoding bar code.3. Scan the START/END bar code.4. Set Length 1 to the first variable length by following the RSS Expanded Length 1, Length 2Programming Instructions below.5. Set Length 2 to the second variable length by following the RSS Expanded Length 1,Length 2 Programming Instructions below.START / ENDPROGRAMMING BARCODESVariable Length DecodingDEFAULTFixed Length Decoding5-20 <strong>Magellan</strong> ® <strong>1000i</strong>


RSS ExpandedRSS Expanded — continuedRSS Expanded Length 1, Length 2 Programming Instructions1. Scan the START/END bar code.2. Scan either the Set Length 1 or Set Length 2 bar code.3. Turn to Appendix C, Alpha-Numeric Pad and scan the three digits (zero padded) representingthe length in decimal notation.NOTEFor RSS Expanded bar codes, only the datacharacters are included in the length calculations.4. Scan the START/END bar code.START / ENDPROGRAMMING BARCODESSet Length 1DEFAULT SETTING FOR THIS FEATURE: 008Set Length 2DEFAULT SETTING FOR THIS FEATURE: 014Product Reference Guide 5-21


SymbologiesRSS Expanded — continuedMinimum ReadsThis feature specifies the minimum number of consecutive times an RSS Expanded label mustbe decoded before it is accepted as good read.START / ENDPROGRAMMING BARCODESMinimum = 1 ReadDEFAULTMinimum = 2 ReadsMinimum = 3 ReadsMinimum = 4 Reads5-22 <strong>Magellan</strong> ® <strong>1000i</strong>


Code 39Code 39The following options apply to the Code 39 symbology.Disable/Enable Code 39When this feature is disabled, the scanner will not read Code 39 bar codes.START / ENDPROGRAMMING BARCODESDisable Code 39Enable Code 39DEFAULTCheck Character CalculationWhen enabled, the scanner will calculate the check character of the labels. Turn this option ononly when a checksum is present in the Code 39 labels.START / ENDPROGRAMMING BARCODESDisable Check Char CalculationDEFAULTEnable Check Char CalculationProduct Reference Guide 5-23


SymbologiesCheck Character TransmitEnable this option to transmit the check character with scanned bar code data.START / ENDPROGRAMMING BARCODESDisable Check Char TransmissionEnable Check Char TransmissionDEFAULTStart/Stop CharactersEnables/disables transmission of Code39 start and stop characters.START / ENDPROGRAMMING BARCODESDon’t Transmit Start/Stop CharactersDEFAULTTransmit Start/Stop Characters5-24 <strong>Magellan</strong> ® <strong>1000i</strong>


Code 39Code 39 Full ASCIIEnables/disables the translation of Code 39 characters to Code 39 full-ASCII characters.START / ENDPROGRAMMING BARCODESDisable Code 39 Full ASCIIDEFAULTEnable Code 39 Full ASCIIProduct Reference Guide 5-25


SymbologiesCode 39 — continuedLength ControlFixed Length Decoding — When fixed length decoding is enabled, the scanner will decode abar code if the label length matches one of the configurable fixed lengths.Variable Length Decoding — When variable length decoding is enabled, the scanner willdecode a bar code if the label length falls in the range of the configurable minimum and maximumlength.Configuring Fixed Length Decoding:1. Scan the START/END bar code.2. Scan the Fixed Length Decoding bar code.3. Scan the START/END bar code.4. Set Length 1 to the first fixed length by following the Code 39 Length 1, Length 2 ProgrammingInstructions below.5. Set Length 2 to the second fixed length (or to ‘000’ if there is only one fixed length) byfollowing the Code 39 Length 1, Length 2 Programming Instructions below.Configuring Variable Length Decoding:1. Scan the START/END bar code.2. Scan the Variable Length Decoding bar code.3. Scan the START/END bar code.4. Set Length 1 to the first variable length by following the Code 39 Length 1, Length 2 ProgrammingInstructions below.5. Set Length 2 to the second variable length by following the Code 39 Length 1, Length 2 ProgrammingInstructions below.START / ENDPROGRAMMING BARCODESVariable Length DecodingDEFAULTFixed Length Decoding5-26 <strong>Magellan</strong> ® <strong>1000i</strong>


Code 39Code 39 Length 1, Length 2 Programming Instructions1. Scan the START/END bar code.2. Scan either the Set Length 1 or Set Length 2 bar code.3. Turn to Appendix C, Alpha-Numeric Pad and scan the three digits (zero padded) representingthe length in decimal notation.NOTEFor Code 39 bar codes, all check, data and full ASCIIshift characters are included in the length calculations.Start/Stop characters are not included.4. Scan the START/END bar code.START / ENDPROGRAMMING BARCODESSet Length 1DEFAULT SETTING FOR THIS FEATURE: 003Set Length 2Quiet ZonesDEFAULT SETTING FOR THIS FEATURE: 050This feature enables/disables the requirement that quiet zones must be present for Code 39 barcodes.START / ENDPROGRAMMING BARCODESDon’t require Quiet ZonesDEFAULTRequire Quiet ZonesProduct Reference Guide 5-27


SymbologiesCode 39 — continuedCode 39 StitchingEnables/disables stitching for Code 39 labels. When parts of a Code 39 bar code are presentedto the scanner with this feature enabled, the bar code parts will be assembled by the scanner’ssoftware, and the data will be decoded if all bar code proofing requirements are met.START / ENDPROGRAMMING BARCODESDisable Code 39 StitchingDEFAULTEnable Code 39 StitchingMinimum ReadsThis feature specifies the minimum number of consecutive times a Code 39 label must bedecoded before it is accepted as good read.START / ENDPROGRAMMING BARCODESMinimum = 1 ReadDEFAULTMinimum = 2 ReadsMinimum = 3 ReadsMinimum = 4 Reads5-28 <strong>Magellan</strong> ® <strong>1000i</strong>


Pharmacode 39Pharmacode 39The following options apply to the Pharmacode 39 symbology.Disable/Enable Pharmacode 39When this feature is disabled, the scanner will not read Pharmacode 39 bar codes.START / ENDPROGRAMMING BARCODESDisable Pharmacode 39DEFAULTEnable Pharmacode 39Start/Stop CharactersEnables or disables transmission of Pharmacode 39 start/stop characters.START / ENDPROGRAMMING BARCODESDon’t Transmit Start/Stop CharactersDEFAULTTransmit Start/Stop CharactersProduct Reference Guide 5-29


SymbologiesCheck Character TransmitEnable this option to transmit the check character with scanned bar code data.START / ENDPROGRAMMING BARCODESDisable Check Char TransmissionEnable Check Char TransmissionDEFAULT5-30 <strong>Magellan</strong> ® <strong>1000i</strong>


Code 128Code 128The following options apply to the Code 128 symbology.Disable/Enable Code 128When this feature is disabled, the scanner will not read Code 128 bar codes.START / ENDPROGRAMMING BARCODESDisable Code 128Enable Code 128DEFAULTTransmit Function CharactersEnables/disables transmission of Code128 function characters 1, 2, 3, and 4.Function codes are transmitted as follows:• FNC1 = 80 hex• FNC2 = 81 hex• FNC3 = 82 hex• FNC4 = 83 hexSTART / ENDPROGRAMMING BARCODESDon’t Transmit Function CharactersDEFAULTTransmit Function CharactersProduct Reference Guide 5-31


SymbologiesCode 128 — continuedLength ControlFixed Length Decoding — When fixed length decoding is enabled, the scanner will decode abar code if the label length matches one of the configurable fixed lengths.Variable Length Decoding — When variable length decoding is enabled, the scanner willdecode a bar code if the label length falls in the range of the configurable minimum and maximumlength.Configuring Fixed Length Decoding:1. Scan the START bar code.2. Scan the Fixed Length Decoding bar code.3. Scan the END bar code.4. Set Length 1 to the first fixed length by following the Code 128 Length 1, Length 2 ProgrammingInstructions below.5. Set Length 2 to the second fixed length (or to ‘000’ if there is only one fixed length) byfollowing the Code 128 Length 1, Length 2 Programming Instructions below.Configuring Variable Length Decoding:1. Scan the START bar code.2. Scan the Variable Length Decoding bar code.3. Scan the END bar code.4. Set Length 1 to the first variable length by following the Code 128 Length 1, Length 2 ProgrammingInstructions below.5. Set Length 2 to the second variable length by following the Code 128 Length 1, Length 2Programming Instructions below.START / ENDPROGRAMMING BARCODESVariable Length DecodingDEFAULTFixed Length Decoding5-32 <strong>Magellan</strong> ® <strong>1000i</strong>


Code 128Code 128 — continuedCode 128 Length 1, Length 2 Programming Instructions1. Scan the START bar code.2. Scan either the Set Length 1 or Set Length 2 bar code.3. Turn to Appendix C, Alpha-Numeric Pad and scan the three digits (zero padded) representingthe length in decimal notation.NOTE4. Scan the END bar code.For Code 128 bar codes, only the data charactersare included in the length calculations.START / ENDPROGRAMMING BARCODESDEFAULT SETTING FOR THIS FEATURE: 001Set Length 1Set Length 2DEFAULT SETTING FOR THIS FEATURE: 080Code 128 Conversion to Code 39Enables/disables expansion of Code 128 labels to Code 39.START / ENDPROGRAMMING BARCODESDisableDEFAULTEnableProduct Reference Guide 5-33


SymbologiesCode 128 — continuedCode 128 StitchingEnables/disables stitching for Code 128 labels. When parts of a Code 128 bar code are presentedto the scanner with this feature enabled, the bar code parts will be assembled by the scanner’ssoftware, and the data will be decoded if all bar code proofing requirements are met.START / ENDPROGRAMMING BARCODESDisable Code 128 StitchingDEFAULTEnable Code 128 StitchingMinimum ReadsThis feature specifies the minimum number of consecutive times a Code 128 label must bedecoded before it is accepted as good read.START / ENDPROGRAMMING BARCODESMinimum = 1 ReadDEFAULTMinimum = 2 ReadsMinimum = 3 ReadsMinimum = 4 Reads5-34 <strong>Magellan</strong> ® <strong>1000i</strong>


Interleaved 2 of 5Interleaved 2 of 5The following options apply to the Interleaved 2 of 5 (I 2 of 5) symbology.Disable/Enable Interleaved 2 of 5When this feature is disabled, the scanner will not read Interleaved 2 of 5 bar codes.START / ENDPROGRAMMING BARCODESDisable Interleaved 2 of 5DEFAULTEnable Interleaved 2 of 5Check Digit CalculationWhen enabled, the scanner will calculate the check digit of the labels.START / ENDPROGRAMMING BARCODESDisable Check Digit CalculationDEFAULTEnable Check Digit CalculationProduct Reference Guide 5-35


SymbologiesInterleaved 2 of 5 — continuedCheck Digit TransmitEnable this option to transmit the check digit with scanned bar code data.START / ENDPROGRAMMING BARCODESDisable Check Digit CalculationDEFAULTEnable Check Digit Calculation5-36 <strong>Magellan</strong> ® <strong>1000i</strong>


Interleaved 2 of 5Interleaved 2 of 5 — continuedLength ControlFixed Length Decoding — When fixed length decoding is enabled, the scanner will decode abar code if the label length matches one of the configurable fixed lengths.Variable Length Decoding — When variable length decoding is enabled, the scanner willdecode a bar code if the label length falls in the range of the configurable minimum and maximumlength.Configuring Fixed Length Decoding:1. Scan the START bar code.2. Scan the Fixed Length Decoding bar code.3. Scan the END bar code.4. Set Length 1 to the first fixed length by following the Interleaved 2 of 5 Length 1, Length 2Programming Instructions below.5. Set Length 2 to the second fixed length (or to ‘000’ if there is only one fixed length) byfollowing the Interleaved 2 of 5 Length 1, Length 2 Programming Instructions below.Configuring Variable Length Decoding:1. Scan the START bar code.2. Scan the Variable Length Decoding bar code.3. Scan the END bar code.4. Set Length 1 to the first variable length by following the Interleaved 2 of 5 Length 1, Length2 Programming Instructions below.5. Set Length 2 to the second variable length by following the Interleaved 2 of 5 Length 1,Length 2 Programming Instructions below.START / ENDPROGRAMMING BARCODESVariable Length DecodingDEFAULTFixed Length DecodingProduct Reference Guide 5-37


SymbologiesInterleaved 2 of 5 — continuedInterleaved 2 of 5 Length 1, Length 2 Programming Instructions1. Scan the START bar code.2. Scan either the Set Length 1 or Set Length 2 bar code.3. Turn to Appendix C, Alpha-Numeric Pad and scan the three digits (zero padded) representingthe length in decimal notation.NOTEFor Interleaved 2 of 5 bar codes, lengths mustbe an even number. Additionally, all check anddata characters are included in the length calculations.4. Scan the END bar code.START / ENDPROGRAMMING BARCODESSet Length 1DEFAULT SETTING FOR THIS FEATURE: 006Set Length 2DEFAULT SETTING FOR THIS FEATURE: 0505-38 <strong>Magellan</strong> ® <strong>1000i</strong>


Interleaved 2 of 5Interleaved 2 of 5 — continuedInterleaved 2 of 5 StitchingEnables/disables stitching for Interleaved 2 of 5 labels. When parts of an Interleaved 2 of 5 barcode are presented to the scanner with this feature enabled, the bar code parts will be assembledby the scanner’s software, and the data will be decoded if all bar code proofing requirements aremet.NOTEOnly functions when Fixed Length Decoding is enabled.START / ENDPROGRAMMING BARCODESDisable Interleaved 2 of 5 StitchingDEFAULTEnable Interleaved 2 of 5 StitchingProduct Reference Guide 5-39


SymbologiesInterleaved 2 of 5 — continuedMinimum ReadsThis feature specifies the minimum number of consecutive times an Interleaved 2 of 5 labelmust be decoded before it is accepted as good read.START / ENDPROGRAMMING BARCODESMinimum = 1 ReadDEFAULTMinimum = 2 ReadsMinimum = 3 ReadsMinimum = 4 Reads5-40 <strong>Magellan</strong> ® <strong>1000i</strong>


CodabarCodabarThe following options apply to the Codabar symbology.Disable/Enable CodabarWhen this feature is disabled, the scanner will not read Codabar bar codes.START / ENDPROGRAMMING BARCODESDisable CodabarDEFAULTEnable CodabarCheck Character VerificationWhen enabled, the scanner will verify the check character of the labels.START / ENDPROGRAMMING BARCODESDisable Check Char VerificationDEFAULTEnable Check Char VerificationProduct Reference Guide 5-41


SymbologiesCheck Character TransmitEnable this option to transmit the check character with scanned bar code data.START / ENDPROGRAMMING BARCODESDisable Check Char TransmissionEnable Check Char TransmissionDEFAULT5-42 <strong>Magellan</strong> ® <strong>1000i</strong>


CodabarCodabar — continuedLength ControlFixed Length Decoding — When fixed length decoding is enabled, the scanner will decode abar code if the label length matches one of the configurable fixed lengths.Variable Length Decoding — When variable length decoding is enabled, the scanner willdecode a bar code if the label length falls in the range of the configurable minimum and maximumlength.Configuring Fixed Length Decoding:1. Scan the START bar code.2. Scan the Fixed Length Decoding bar code.3. Scan the END bar code.4. Set Length 1 to the first fixed length by following the Codabar Length 1, Length 2 ProgrammingInstructions below.5. Set Length 2 to the second fixed length (or to ‘000’ if there is only one fixed length) byfollowing the Codabar Length 1, Length 2 Programming Instructions below.Configuring Variable Length Decoding:1. Scan the START bar code.2. Scan the Variable Length Decoding bar code.3. Scan the END bar code.4. Set Length 1 to the first variable length by following the Codabar Length 1, Length 2 ProgrammingInstructions below.5. Set Length 2 to the second variable length by following the Codabar Length 1, Length 2Programming Instructions below.START / ENDPROGRAMMING BARCODESVariable Length DecodingDEFAULTFixed Length DecodingProduct Reference Guide 5-43


SymbologiesCodabar — continuedCodabar Length 1, Length 2 Programming Instructions1. Scan the START bar code.2. Scan either the Set Length 1 or Set Length 2 bar code.3. Turn to Appendix C, Alpha-Numeric Pad and scan the three digits (zero padded) representingthe length in decimal notation.NOTEFor Codabar bar codes, all start, stop, checkand data characters are included in the lengthcalculations.4. Scan the END bar code.START / ENDPROGRAMMING BARCODESSet Length 1DEFAULT SETTING FOR THIS FEATURE: 003Set Length 2Quiet ZonesDEFAULT SETTING FOR THIS FEATURE: 050This feature enable/disables the requirement that quiet zones must be present for Codabar barcodes.START / ENDPROGRAMMING BARCODESDon’t require Quiet ZonesDEFAULTRequire Quiet Zones5-44 <strong>Magellan</strong> ® <strong>1000i</strong>


CodabarCodabar — continuedStart/Stop Character TypeCodabar has four pairs of Start/Stop patterns. Select one pair to match your application.START / ENDPROGRAMMING BARCODESStart/Stop Type: ABCD/TN*EStart/Stop Type: ABCD/ABCDStart/Stop Type: abcd/tn*eStart/Stop Type: abcd/abcdDEFAULTStart/Stop Character TransmissionThe transmission of start and end characters of Codabar is selected below.START / ENDPROGRAMMING BARCODESDisable Start/Stop Char TransmissionEnable Start/Stop Char TransmissionDEFAULTProduct Reference Guide 5-45


SymbologiesStart/Stop Character MatchThis feature enables/disables the requirement that start and stop characters match.START / ENDPROGRAMMING BARCODESDisable Start/Stop Char MatchDEFAULTEnable Start/Stop Char MatchCodabar StitchingEnables/disables stitching for Codabar labels. When parts of a Codabar label are presented tothe scanner with this feature enabled, the bar code parts will be assembled by the scanner’s software,and the data will be decoded if all bar code proofing requirements are met.NOTEOnly functions when Fixed Length Decoding is enabled.START / ENDPROGRAMMING BARCODESDisable Codabar StitchingDEFAULTEnable Codabar Stitching5-46 <strong>Magellan</strong> ® <strong>1000i</strong>


CodabarCodabar — continuedMinimum ReadsThis feature specifies the minimum number of consecutive times a Codabar label must bedecoded before it is accepted as a good read.START / ENDPROGRAMMING BARCODESMinimum = 1 ReadDEFAULTMinimum = 2 ReadsMinimum = 3 ReadsMinimum = 4 ReadsProduct Reference Guide 5-47


SymbologiesCode 93The following options apply to the Code 93 symbology.Disable/Enable Code 93When this feature is disabled, the scanner will not read Code 93 bar codes.START / ENDPROGRAMMING BARCODESDisable Code 93DEFAULTEnable Code 935-48 <strong>Magellan</strong> ® <strong>1000i</strong>


Code 93Code 93 — continuedLength ControlFixed Length Decoding — When fixed length decoding is enabled, the scanner will decode abar code if the label length matches one of the configurable fixed lengths.Variable Length Decoding — When variable length decoding is enabled, the scanner willdecode a bar code if the label length falls in the range of the configurable minimum and maximumlength.Configuring Fixed Length Decoding:1. Scan the START/END bar code.2. Scan the Fixed Length Decoding bar code.3. Scan the START/END bar code.4. Set Length 1 to the first fixed length by following the Code 93 Length 1, Length 2 ProgrammingInstructions below.5. Set Length 2 to the second fixed length (or to ‘000’ if there is only one fixed length) byfollowing the Code 93 Length 1, Length 2 Programming Instructions below.Configuring Variable Length Decoding:1. Scan the START/END bar code.2. Scan the Variable Length Decoding bar code.3. Scan the START/END bar code.4. Set Length 1 to the first variable length by following the Code 93 Length 1, Length 2 ProgrammingInstructions below.5. Set Length 2 to the second variable length by following the Code 93 Length 1, Length 2 ProgrammingInstructions below.START / ENDPROGRAMMING BARCODESVariable Length DecodingDEFAULTFixed Length DecodingProduct Reference Guide 5-49


SymbologiesCode 93 — continuedCode 93 Length 1, Length 2 Programming Instructions1. Scan the START bar code.2. Scan either the Set Length 1 or Set Length 2 bar code.3. Turn to Appendix C, Alpha-Numeric Pad and scan the three digits (zero padded) representingthe length in decimal notation.NOTE4. Scan the END bar code.For Code 93 bar codes, only the data charactersare included in the length calculations.START / ENDPROGRAMMING BARCODESDEFAULT SETTING FOR THIS FEATURE: 001Set Length 1Set Length 2DEFAULT SETTING FOR THIS FEATURE: 0505-50 <strong>Magellan</strong> ® <strong>1000i</strong>


Code 93Code 93 — continuedCode 93 StitchingEnables/disables stitching for Code 93 bar codes. When parts of a Code 93 label are presentedto the scanner with this feature enabled, the bar code parts will be assembled by the scanner’ssoftware, and the data will be decoded if all bar code proofing requirements are met.START / ENDPROGRAMMING BARCODESDisable Code 93 StitchingDEFAULTEnable Code 93 StitchingMinimum ReadsThis feature specifies the minimum number of consecutive times a Code 93 label must bedecoded before it is accepted as a good read.START / ENDPROGRAMMING BARCODESMinimum = 1 ReadDEFAULTMinimum = 2 ReadsMinimum = 3 ReadsMinimum = 4 ReadsProduct Reference Guide 5-51


SymbologiesMSI/PlesseyThe following options apply to the MSI/Plessey symbology.Disable/Enable MSI/PlesseyWhen this feature is disabled, the scanner will not read MSI/Plessey bar codes.START / ENDPROGRAMMING BARCODESDisable MSI/PlesseyDEFAULTEnable MSI/PlesseyCheck Digit VerificationThis feature specifies whether one or two check digits are to be calculated and verified.START / ENDPROGRAMMING BARCODESDisable Check Digit VerificationDEFAULTEnable Check Digit Verification5-52 <strong>Magellan</strong> ® <strong>1000i</strong>


MSI/PlesseyMSI/Plessey — continuedCheck Digit TransmitWhen this option is enabled, the scanner will transmit one-digit or two-digit check digits,depending upon the setting for check digit verification.START / ENDPROGRAMMING BARCODESDisable Check Digit TransmissionEnable Check Digit TransmissionDEFAULTNumber of Check CharactersSpecifies number of MSI/Plessey check characters to be calculated and verifiedSTART / ENDPROGRAMMING BARCODES1 Check CharacterDEFAULT2 Check CharactersProduct Reference Guide 5-53


SymbologiesMSI/Plessey — continuedLength ControlFixed Length Decoding — When fixed length decoding is enabled, the scanner will decode abar code if the label length matches one of the configurable fixed lengths.Variable Length Decoding — When variable length decoding is enabled, the scanner willdecode a bar code if the label length falls in the range of the configurable minimum and maximumlength.Configuring Fixed Length Decoding:1. Scan the START/END bar code.2. Scan the Fixed Length Decoding bar code.3. Scan the START/END bar code.4. Set Length 1 to the first fixed length by following the MSI/Plessey Length 1, Length 2 ProgrammingInstructions below.5. Set Length 2 to the second fixed length (or to ‘000’ if there is only one fixed length) byfollowing the MSI/Plessey Length 1, Length 2 Programming Instructions below.Configuring Variable Length Decoding:1. Scan the START/END bar code.2. Scan the Variable Length Decoding bar code.3. Scan the START/END bar code.4. Set Length 1 to the first variable length by following the MSI/Plessey Length 1, Length 2Programming Instructions below.5. Set Length 2 to the second variable length by following the MSI/Plessey Length 1, Length 2Programming Instructions below.START / ENDPROGRAMMING BARCODESVariable Length DecodingDEFAULTFixed Length Decoding5-54 <strong>Magellan</strong> ® <strong>1000i</strong>


MSI/PlesseyMSI/Plessey — continuedMSI/Plessey Length 1, Length 2 Programming Instructions1. Scan the START bar code.2. Scan either the Set Length 1 or Set Length 2 bar code.3. Turn to Appendix C, Alpha-Numeric Pad and scan the three digits (zero padded) representingthe length in decimal notation.NOTEFor MSI/Plessey bar codes, all check and datacharacters are included in the length calculations.4. Scan the END bar code.START / ENDPROGRAMMING BARCODESSet Length 1DEFAULT SETTING FOR THIS FEATURE: 004Set Length 2DEFAULT SETTING FOR THIS FEATURE: 016Product Reference Guide 5-55


SymbologiesMSI/Plessey — continuedMSI/Plessey StitchingEnables/disables stitching for MSI/Plessey bar codes. When parts of an MSI/Plessey label arepresented to the scanner with this feature enabled, the bar code parts will be assembled by thescanner’s software, and the data will be decoded if all bar code proofing requirements are met.NOTEOnly functions when Fixed Length Decoding is enabled.START / ENDPROGRAMMING BARCODESDisable MSI/Plessey StitchingDEFAULTEnable MSI/Plessey Stitching5-56 <strong>Magellan</strong> ® <strong>1000i</strong>


MSI/PlesseyMSI/Plessey — continuedMinimum ReadsThis feature specifies the minimum number of consecutive times an MSI/Plessey label must bedecoded before it is accepted as good read.START / ENDPROGRAMMING BARCODESMinimum = 1 ReadDEFAULTMinimum = 2 ReadsMinimum = 3 ReadsMinimum = 4 ReadsProduct Reference Guide 5-57


SymbologiesStandard 2 of 5The following options apply to the Standard 2 of 5 symbology.Disable/Enable Standard 2 of 5When this feature is disabled, the scanner will not read Standard 2 of 5 bar codes.START / ENDPROGRAMMING BARCODESDisable Std 2 of 5DEFAULTEnable Std 2 of 5Check Digit VerificationWhen enabled, the scanner will verify the check digit of the labels.START / ENDPROGRAMMING BARCODESDisable Check Digit VerificationDEFAULTEnable Check Digit Verification5-58 <strong>Magellan</strong> ® <strong>1000i</strong>


Standard 2 of 5Standard 2 of 5 — continuedCheck Digit TransmitWhen this option is enabled, the scanner will transmit the check digit.START / ENDPROGRAMMING BARCODESDisable Check Digit TransmissionEnable Check Digit TransmissionDEFAULTProduct Reference Guide 5-59


SymbologiesStandard 2 of 5 — continuedLength ControlFixed Length Decoding — When fixed length decoding is enabled, the scanner will decode abar code if the label length matches one of the configurable fixed lengths.Variable Length Decoding — When variable length decoding is enabled, the scanner willdecode a bar code if the label length falls in the range of the configurable minimum and maximumlength.Configuring Fixed Length Decoding:1. Scan the START/END bar code.2. Scan the Fixed Length Decoding bar code.3. Scan the START/END bar code.4. Set Length 1 to the first fixed length by following the Standard 2 of 5 Length 1, Length 2Programming Instructions below.5. Set Length 2 to the second fixed length (or to ‘000’ if there is only one fixed length) byfollowing the Standard 2 of 5 Length 1, Length 2 Programming Instructions below.Configuring Variable Length Decoding:1. Scan the START/END bar code.2. Scan the Variable Length Decoding bar code.3. Scan the START/END bar code.4. Set Length 1 to the first variable length by following the Standard 2 of 5 Length 1, Length 2Programming Instructions below.5. Set Length 2 to the second variable length by following the Standard 2 of 5 Length 1, Length2 Programming Instructions below.START / ENDPROGRAMMING BARCODESVariable Length DecodingDEFAULTFixed Length Decoding5-60 <strong>Magellan</strong> ® <strong>1000i</strong>


Standard 2 of 5Standard 2 of 5 — continuedStandard 2 of 5 Length 1, Length 2 Programming Instructions1. Scan the START bar code.2. Scan either the Set Length 1 or Set Length 2 bar code.3. Turn to Appendix C, Alpha-Numeric Pad and scan the three digits (zero padded) representingthe length in decimal notation.NOTEFor Standard 2 of 5 bar codes, all check anddata characters are included in the length calculations.4. Scan the END bar code.START / ENDPROGRAMMING BARCODESSet Length 1DEFAULT SETTING FOR THIS FEATURE: 008Set Length 2DEFAULT SETTING FOR THIS FEATURE: 050Product Reference Guide 5-61


SymbologiesStandard 2 of 5 — continuedStandard 2 of 5 StitchingEnables/disables stitching for Standard 2 of 5 bar codes. When parts of a Standard 2 of 5 labelare presented to the scanner with this feature enabled, the bar code parts will be assembled bythe scanner’s software, and the data will be decoded if all bar code proofing requirements aremet.NOTEOnly functions when Fixed Length Decoding is enabled.START / ENDPROGRAMMING BARCODESDisable Std 2 of 5 StitchingDEFAULTEnable Std 2 of 5 Stitching5-62 <strong>Magellan</strong> ® <strong>1000i</strong>


Standard 2 of 5Standard 2 of 5 — continuedMinimum ReadsThis feature specifies the minimum number of consecutive times a Standard 2 of 5 label must bedecoded before it is accepted as good read.START / ENDPROGRAMMING BARCODESMinimum = 1 ReadDEFAULTMinimum = 2 ReadsMinimum = 3 ReadsMinimum = 4 ReadsProduct Reference Guide 5-63


SymbologiesNOTES5-64 <strong>Magellan</strong> ® <strong>1000i</strong>


Appendix AProduct SpecificationsOptical and Read Performance ParametersParameterScan VolumeScan PatternScan RateMinimum ResolutionDepth of Field(100% UPC Labels)Specification70 cubic inches100 lines1,100 digital scan lines5 mil0 - 6”Minimum Print Contrast Ratio 25%Skew (Yaw)± 75Pitch± 65RollBetween 0 and 360Scanner Dimensions139.58mm5.5"97.84mm3.85"75.44mm2.97"83.25mm3.28"25.4mm1.0"Product Reference Guide A-1


Physical PropertiesParameterDimensions (Scanner only):Dimensions (Scanner w/BaseStation):Weight (Scanner)Weight (Base Station)3.3” x2.8” x3.7”5.5” x2.9” x3.8”7.0 oz.6.6 oz.SpecificationElectrical ParametersParameterOperating VoltageInput CurrentOperating (idle)Operating (label read)Specification(Two Models)5V unit = 5.0V ± 5%12V unit = 7-12V


Appendix BCable PinoutsStandard Cable Pinouts (Primary Interface Cables)RS-23212 CTS34 RTS5 RXD6 TXD78 VCC IN9 GND10Pin 110 Position RJ(Scanner End)IBM Port 5B/9B/171234 DATA -56 DATA +78 VCC_IN9 GND10Pin 110 Position RJ(Scanner End)Product Reference Guide B-1


Cable PinoutsIBM USB1234 D -56 D +78 VIN9 GND10Pin 110 Position RJ(Scanner End)USB & USB Keyboard1234 D -56 D +78 VBUS_VIN9 GND10Pin 110 Position RJ(Scanner End)Wand Emulation123 WAND~45678 VCC_IN9 GND10Pin 110 Position RJ(Scanner End)Keyboard Wedge12 KB_DATA3 AT_CLK4 KB_CLK56 AT_DATA78 VCC_IN9 GND10Pin 110 Position RJ(Scanner End)B-2 <strong>Magellan</strong> ® <strong>1000i</strong>


Appendix CAlpha-Numeric PadABCDEFProduct Reference Guide C-1


Alpha-Numeric Pad1234567890C-2 <strong>Magellan</strong> ® <strong>1000i</strong>


Appendix DDefault SettingsDefaults by SymbologyThe following is a partial list of key settings for each symbology type.Code TypeReadEnableChecksumVerificationEnableChecksumTransmissionEnableLabelIDUPC-A ✓ ✓ ✓ AUPC-E ✓ ✓ ✓ EEAN-13 ✓ ✓ ✓ FEAN-8 ✓ ✓ ✓ FFRSS-14R4RSS ExpandedRXCode 39 ✓ ✓ *PharmaCode 39ACode 128 ✓ #Interleaved 2 of 5 ✓ iCodabar ✓ %Code 93&MSI/Plessey ✓ @Standard 2 of 5 ✓ sInterface Default ExceptionsThe factory default settings indicated in the programming sections (in bold text) reflect factory configuration forthe RS-232 standard interface. The following tables list default exceptions by interface for the remaining availableinterfaces.Product Reference Guide D-1


IBM InterfacesIBM Interfaces include IBM-USB, IBM Port 9B, IBM Port5B and IBM Port17.ParameterIBM Interface TypeNumber of Host Transmit BuffersLabel I.D. TransmissionSuffix CharactersDefault SettingIBM Port 9BOne BufferDisableNo SuffixRS-232 Wincor/NixdorfParameterDefault SettingInterface TypeNumber of Host Transmit BuffersRS-232 ParityRS-232 Hardware ControlUPC-E Check Character ConversionUCC/EAN-128 Label IDCode 39 Label IDCode 93 Label IDCode 128 Label IDCodabar Label IDEAN-8 Label IDEAN-13 Label IDISBN Label IDInterleaved 2 of 5 Label IDStandard 2 of 5 Label IDMSI/Plessey Label IDUPC-E Label IDRSS-14 Label IDRSS Expanded Label IDRS-232-WNOne BufferOddCTS Flow ControlDisabled‘P’‘M’‘L’‘K’‘N’‘B’‘A’‘A’‘I’‘H’‘O’‘C’‘E’‘E’D-2 <strong>Magellan</strong> ® <strong>1000i</strong>


Interface Default ExceptionsWand EmulationParameterDefault SettingInterface TypeCode 39 Start/Stop Character TransmissionWand EmulationEnableKeyboardsKeyboard interfaces include USB Keyboard and Keyboard Wedge A-Y.ParameterKeyboard Wedge Interface TypeLabel ID TransmissionUSB KeyboardDisableDefault SettingProduct Reference Guide D-3


NOTESD-4 <strong>Magellan</strong> ® <strong>1000i</strong>


Appendix EKeyboard Function KeyMappingsKeyboard Model Cross ReferenceTable E-1 summarizes the keyboard models, their defined protocol, scancode set, and some unique features. Theremaining tables in this chapter provide the function key maps associated with each of the scancode sets.Table E-1. Keyboard Model Cross ReferenceModel TypeI/F IDTransmissionProtocolScancodeSetFunc.UseKeyCountryMapModeSupportPC/XT Foreign ALT Mode Wedge A PC/XT Scan Set 1 No NoAT;PS/2 25-286;PS/2 30-286;PS/2 50, 50Z;Wedge B AT/PS2 Scan Set 2 No NoPS/2 60,70,80,90,95Foreign ALT ModePS/2 25 and 30Wedge C AT/PS2 Scan Set 1 No NoForeign ALT ModePC/XT U.S. Mode Wedge D PC/XT Scan Set 1 Yes NoAT;PS/2 25-286;PS/2 30-286;PS/2 50, 50Z;PS/2 60,70,80,90,95U.S. Mode + specificcountry supportPS/2 25 and 30U.S. ModeIBM 3xxx Terminals(122-key keyboard)IBM 3xxx Terminals(102-key keyboard)PS55 5530T withJAPANESE DOS (TDOS)NEC 9801Wedge E AT/PS2 Scan Set 2 Yes YesWedge F AT/PS2 Scan Set 1 Yes NoWedge G AT/PS2 Scan Set 3 Yes NoWedge H AT/PS2 Scan Set 3 Yes NoWedge I AT/PS2 Japanese Yes NoDOSNECWedge JNEC 9801 Yes No9801Product Reference Guide E-1


Table E-2. Scanset 1 Function Key MapTable E-3. Scanset 2 Function Key MapASCII(hex)ASCII code Key Scancode00 NUL ALT right Make E0h 38h01 SOH ALT right Break E0h B8h02 STX ALT left Make 38h03 ETX ALT left Break B8h04 EOT CTRL left Make 1Dh05 ENQ CTRL left Break 9Dh06 ACK CTRL right Make E0h 1Dh07 BEL CTRL right Break E0h 9Dh08 BS BS 0Eh09 HT TAB right 0Fh0A LF RIGHT arrow (inner keypad) 4Dh + E00B VT TAB left 0Fh + S0C FF Enter (inner keypad) 1Ch + E00D CR CR 1Ch0E SO INSERT (inner keypad) 52h + E00F SI PAGE UP (inner keypad) 49h + E010 DLE PAGE DOWN (inner keypad) 51h + E011 DC1 HOME (inner keypad) 47h + E012 DC2 LEFT arrow (inner keypad) 4Bh + E013 DC3 DOWN arrow (inner keypad) 50h + E014 DC4 UP arrow (inner keypad) 48h + E0ASCII(hex)ASCII code Key Scancode00 NUL ALT right Make E0h 11h01 SOH ALT right Break E0h F0h 11h02 STX ALT left Make 11h03 ETX ALT left Break F0h 11h04 EOT CTRL left Make 14h05 ENQ CTRL left Break F0h 14h06 ACK CTRL right Make E0h 14h07 BEL CTRL right Break E0h F0h 14h08 BS BS 66h09 HT TAB right 0Dh0A LF RIGHT arrow (inner keypad) 74h + E00B VT TAB left 0Dh + S0C FF Enter (right keypad) 5Ah + E00D CR CR 5Ah0E SO INSERT (inner keypad) 70h + E00F SI PAGE UP (inner keypad) 7Dh + E010 DLE PAGE DOWN (inner keypad) 7Ah + E011 DC1 HOME (inner keypad) 6Ch + E012 DC2 LEFT arrow (inner keypad) 6Bh + E013 DC3 DOWN arrow (inner keypad) 72h + E014 DC4 UP arrow (inner keypad) 75h + E015 NAK F6 0Bh16 SYN F1 05h17 ETB F2 06h18 CAN F3 04h19 EM F4 0Ch1A SUB F5 03hE-2 <strong>Magellan</strong> ® <strong>1000i</strong>


Keyboard Model Cross Reference1B ESC ESC 76h1C FS F7 83h1D GS F8 0Ah1E RS F9 01h1F US F10 09hTable E-4. Scanset 3, 102-Key Function Key MapASCII (hex) ASCII code Key Scancode00 NUL ALT right Make 39h01 SOH ALT right Break F0h 39h02 STX ALT left Make 19h03 ETX ALT left Break F0h 19h04 EOT CTRL left Make 11h05 ENQ CTRL left Break F0h 11h06 ACK CTRL right Make 58h07 BEL CTRL right Break F0h 58h08 BS BS 66h09 HT TAB right 0Dh0A LF RIGHT arrow (inner keypad) 6Ah0B VT TAB left 0Dh + S0C FF Enter (inner keypad) 79h0D CR CR 5Ah0E SO INSERT (inner keypad) 67h0F SI PAGE UP (inner keypad) 6Fh10 DLE PAGE DOWN (inner keypad) 6Dh11 DC1 HOME (inner keypad) 6Eh12 DC2 LEFT arrow (inner keypad) 61h13 DC3 DOWN arrow (inner keypad) 60h14 DC4 UP arrow (inner keypad) 63h15 NAK F6 2Fh16 SYN F1 07h17 ETB F2 0Fh18 CAN F3 17h19 EM F4 1Fh1A SUB F5 27h1B ESC ESC 08h1C FS F7 37h1D GS F8 3Fh1E RS F9 47h1F US F10 4FhProduct Reference Guide E-3


Table E-5. Scanset 3 122-Key Function Key MapASCII (hex) ASCII code Key Scancode00 NUL ALT Right Make 39h01 SOH ALT Right Break F0h 39h02 STX ALT left Make 19h03 ETX ALT left Break F0h 19h04 EOT CTRL left (RESET) Make only 11h05 ENQ CTRL left (RESET) Make/Break 11h F0h 11h06 ACK ONLINE Enter Make only 58h07 BEL ONLINE Enter Make/Break 58h F0h 58h08 BS BS 66h09 HT TAB right 0Dh0A LF RIGHT arrow (inner keypad) 6Ah0B VT TAB left 0Dh + S0C FF CR (FIELD EXIT) Make only 5Ah F0h 5Ah0D CR CR (FIELD EXIT) Make/Break 5Ah0E SO INSERT (inner keypad) 65h0F SI FIELD + 79h10 DLE FIELD - 7Ch11 DC1 HOME (inner keypad) 62h12 DC2 LEFT arrow (inner keypad) 61h13 DC3 DOWN arrow (inner keypad) 60h14 DC4 UP arrow (inner keypad) 63h15 NAK F6 2Fh16 SYN F1 07h17 ETB F2 0Fh18 CAN F3 17h19 EM F4 1Fh1A SUB F5 27h1B ESC ESC 08h1C FS F7 37h1D GS F8 3Fh1E RS F9 47h1F US F10 4FhE-4 <strong>Magellan</strong> ® <strong>1000i</strong>


Keyboard Model Cross ReferenceTable E-6. Japanese DOS Function Key MapASCII value ASCII code Key Scancode00h NUL ALT right Make 31h01h SOH ALT right Break B1h02h STX ALT left Make 31h03h ETX ALT left Break B1h04h EOT CTRL left Make 41h05h ENQ CTRL left Break C1h06h ACK CTRL right Make 41h07h BEL CTRL right Break C1h08h BS BS 3Eh09h HT TAB right 3Ch0Ah LF RIGHT arrow (inner keypad) 4Dh0Bh VT TAB left 3Ch + S0Ch FF Enter (right keypad) 60h0Dh CR CR 3Bh0Eh SO INSERT (inner keypad) 52h0Fh SI PAGE UP (inner keypad) 49h10h DLE PAGE DOWN (inner keypad) 51h11h DC1 HOME (inner keypad) 4Ch12h DC2 LEFT arrow (inner keypad) 4Bh13h DC3 DOWN arrow (inner keypad) 4Ah14h DC4 UP arrow (inner keypad) 4Eh15h NAK F6 6Dh16h SYN F1 68h17h ETB F2 69h18h CAN F3 6Ah19h EM F4 6Bh1Ah SUB F5 6Ch1Bh ESC ESC 3Dh1Ch FS F7 6Eh1Dh GS F8 6Fh1Eh RS F9 70h1Fh US F10 71hProduct Reference Guide E-5


Table E-7. NEC 9801-Key Function Key MapASCII value ASCII code Key Scancode00h NUL unused n/a01h SOH CR 1Ch02h STX CAPS LOCK ON (make) 71h03h ETX CAPS LOCK OFF (break) F1h04h EOT CTRL left Make 74h05h ENQ CTRL left Break F4h06h ACK CTRL-C 60h07h BEL n/a n/a08h BS BS 0Eh09h HT TAB right 0Fh0Ah LF RIGHT arrow (inner keypad) 3Ch0Bh VT TAB left 0Fh + S0Ch FF DELETE 39h0Dh CR CR 1Ch0Eh SO INSERT (inner keypad) 38h0Fh SI KATAKANA LOCK ON (Make) 72h10h DLE KATAKANA LOCK OFF (Break) F2h11h DC1 HOME (inner keypad) 3Eh12h DC2 LEFT arrow (inner keypad) 3Bh13h DC3 DOWN arrow (inner keypad) 3Dh14h DC4 UP arrow (inner keypad) 3Ah15h NAK F6 67h16h SYN F1 62h17h ETB F2 63h18h CAN F3 64h19h EM F4 65h1Ah SUB F5 66h1Bh ESC ESC 00h1Ch FS F7 68h1Dh GS F8 69h1Eh RS F9 6Ah1Fh US F10 6BhE-6 <strong>Magellan</strong> ® <strong>1000i</strong>


Appendix FHost CommandsAccepting RS-232 CommandsThe scanner responds to the following RS-232 commands:COMMAND ASCII HEX COMMENTEnable Scanner E 0x45DisableScannerD0x44Reset Scanner R 0x52Not On FileIndicationBeep GoodRead ToneForce GoodRead ToneF 0x46 Long series of beepsB 0x42 Beeps if Good Read Beep is enabled! 0x01 Beeps regardless of beep settingBel ‘ 0x07 Force Good Read ToneIdentificationrequesti 0x69 Returns long response aHealth request h 0x68 Returns long response aStatus request s 0x73 Returns long response aa. Call Tech Support for information.If one of the above commands is received, the scanner will perform the steps indicated for the command. Hostcommands for other interfaces are also available. Contact Tech Support for more details.Product Reference Guide F-1


NOTESF-2 <strong>Magellan</strong> ® <strong>1000i</strong>


Appendix GSample SymbolsUPC-A0 50 1234567890Interleaved 2 of 5Code 1281234567890Code 128EAN-13Code 399 780330 290951BC321CodabarCode 93A13579B123456-9$Code 2 of 5123456Product Reference Guide G-1


Sample SymbolsRSS-14(01)00123456789012RSS Expanded0100123456789050G-2 <strong>Magellan</strong> ® <strong>1000i</strong>


ASCII ChartASCIIChar.Hex No.ASCIIChar.Hex No.ASCIIChar.Hex No.ASCIIChar.HexNo.NULSOHSTXETXEOTENQACKBELBSHTLFVTFFCRSOSIDLEDC1DC2DC3DC4NAKSYNETBCANEMSUBESCFSGSRSUS000102030405060708090A0B0C0D0E0F101112131415161718191A1B1C1D1E1FSP!“#$%&’()*+,-./0123456789:;?202122232425262728292A2B2C2D2E2F303132333435363738393A3B3C3D3E3F@ABCDEFGHIJKLMNOPQRSTUVWXYZ[\]^_404142434445464748494A4B4C4D4E4F505152535455565758595A5B5C5D5E5F‘abcdefghijklmnopqrstuvwxyz{|}~DEL606162636465666768696A6B6C6D6E6F707172737475767778797A7B7C7D7E7F


AustraliaDatalogic Scanning Pty LtdNorth Ryde, AustraliaTelephone: [61] (2) 9870 3200Fax: [61] (2) 9878 8688France and BeneluxDatalogic Scanning SarlLES ULIS Cedex, FranceTelephone: [33].01.64.86.71.00Fax: [33].01.64 46.72.44GermanyDatalogic Scanning GmbHDarmstadt, GermanyTelephone: 49 (0) 61 51/93 58-0Fax: 49 (0) 61 51/93 58 58ItalyDatalogic Scanning SpAVimercate (MI), ItalyTelephone: [39] (0) 39/62903.1Fax: [39] (0) 39/6859496JapanDatalogic Scanning KKShinagawa, Tokyo, JapanTelephone: 81 (0)3 3491 6761Fax: 81 (0)3 3491 6656Latin AmericaDatalogic Scanning, IncMiami, Florida, USATelephone: (305) 591-3222Fax: (305) 591-3007Spain and PortugalDatalogic Scanning Sarl Sucursal en EspañaMadrid, SpainTelephone: 34 91 746 28 60Fax: 34 91 742 35 33United KingdomDatalogic Scanning LTDWatford, EnglandTelephone: 44 (0) 1923 809500Fax: 44 (0) 1923 809 505www.scanning.datalogic.comDatalogic Scanning, Inc.959 Terry StreetEugene, OR 97402Telephone: (541) 683-5700Fax: (541) 345-7140©2005-2007 Datalogic Scanning, Inc. R44-2747 (Rev B) 9/07