5. surGicaL ProceDure<br />
Drilling procedure for Standard and Standard Plus implants<br />
Preparati<strong>on</strong> of <strong>the</strong> alveolar ridge<br />
after opening <strong>the</strong> gingiva, a narrowly tapering ridge can be carefully reduced and<br />
smoo<strong>the</strong>d with a large round burr in order to obtain a flat b<strong>on</strong>e surface and provide<br />
a sufficiently wide area of b<strong>on</strong>e. When choosing <strong>the</strong> implant length (sLactive/sLa ®<br />
surface), <strong>the</strong> vertical reducti<strong>on</strong> of <strong>the</strong> b<strong>on</strong>e has to be c<strong>on</strong>sidered.<br />
Rule of thumb: <strong>the</strong> orofacial b<strong>on</strong>e wall around <strong>the</strong> implant must be at least<br />
1 mm thick.<br />
after <strong>the</strong> alveolar ridge has been prepared appropriately, preparati<strong>on</strong> of <strong>the</strong> implant<br />
bed can commence.<br />
Depending <strong>on</strong> <strong>the</strong> situati<strong>on</strong>, b<strong>on</strong>e scal-<br />
loping in <strong>the</strong> regi<strong>on</strong> of <strong>the</strong> subsequent<br />
implant bed (with <strong>the</strong> aid of a large<br />
round burr) is recommended prior to<br />
preparati<strong>on</strong> of <strong>the</strong> implant bed. <strong>the</strong> implantati<strong>on</strong><br />
site is <strong>the</strong>n marked with <strong>the</strong><br />
smallest round burr. it is <strong>the</strong>n widened<br />
with <strong>the</strong> remaining round burs in ascending<br />
order with c<strong>on</strong>stant cooling. this<br />
procedure makes it possible to correct<br />
<strong>the</strong> positi<strong>on</strong> slightly, if necessary.<br />
to insert an implant with an endosteal<br />
Ø 3.3 mm, <strong>the</strong> use of <strong>the</strong> round burr<br />
Ø 3.1 mm is not necessary and is not<br />
recommended.<br />
Ø 1.4 mm<br />
max. 800 rpm<br />
Ø 2.3 mm<br />
max. 800 rpm<br />
Ø 3.1 mm<br />
max. 800 rpm