p <strong>the</strong> easy-to-read user guide ensures a reliable working sequence through colourcoded<br />

arrows and silic<strong>on</strong> sleeves.<br />

p clear illustrati<strong>on</strong>s and drill length stops allow <strong>the</strong> arranged instruments, screws and<br />

healing caps to be checked at a glance for correctness and completeness.<br />

p <strong>the</strong> instruments are positi<strong>on</strong>ed securely in <strong>the</strong> silic<strong>on</strong>e sleeves for sterilisati<strong>on</strong> and<br />

storage.<br />

p <strong>the</strong> cassette can be packed according to <strong>the</strong> working procedure (using <strong>the</strong><br />

handpiece or manually with <strong>the</strong> ratchet).<br />

p <strong>the</strong> surgical cassette also houses a separate screw c<strong>on</strong>tainer in which all <strong>the</strong><br />

required screws and healing caps are arranged, thus providing ease of access to<br />

<strong>the</strong>m.<br />

Guidelines for <strong>the</strong> sterilisati<strong>on</strong> of <strong>the</strong> Surgical cassette<br />

Method Temperature Exposure Time Dry Time<br />

steam sterilizati<strong>on</strong> Prevacuum cycle 134 °c/273 °f min. 4 –18 min 20 – 60 min*<br />

steam sterilizati<strong>on</strong> Gravity cycle<br />

no dry heat sterilizati<strong>on</strong>!<br />

134 °c/273 °f min. 40 min 20 – 60 min*<br />

*Instruments that not thoroughly dried may corrode.<br />

Before sterilisati<strong>on</strong>, <strong>the</strong> cassette is packed (e.g. sealed in foil or wrapped in towels)<br />

in order to keep it sterile.<br />

Important<br />

p chemical sterilisati<strong>on</strong> is not recommended!<br />

p Do not use dry heat sterilisati<strong>on</strong><br />

p ensure that <strong>the</strong> individual sterilisati<strong>on</strong> parameters comply with <strong>the</strong> current regulati<strong>on</strong>s<br />

of <strong>the</strong> respective country.<br />

In order to avoid damaging <strong>the</strong> surgical cassette during autoclaving, it must<br />

be placed correctly in <strong>the</strong> autoclave (see Fig.).<br />

see also DvD “straumann ® Dental implant system-surgery“,<br />

art. no. 150.541, “straumann surgical cassette“.