Operating Instructions - Snap-on Equipment
Operating Instructions - Snap-on Equipment Operating Instructions - Snap-on Equipment
PC MaintenanceThe initial Gas Calibration window opens.Figure 3G-1: Initial Gas Calibration Screen✓ A warning displays and will automatically disappear followed by an autozero.2. The system then performs a zero calibration. In zero calibration, the gasanalyzer submits clean air, room air, into the sample system to verify thatthe analyzer reads properly when there is no gas present in the sample cell.✓ Make sure the vehicle exhaust sample probe is in fresh air, NOTsampling vehicle exhaust of any type.3. The Calibration Bottle Values screen then displays. Verify or change thevalues to match your calibration gas.— To change values, Tab to the value you want to change, press Deleteand input the correct value.Figure 3G-2: Calibration Bottle Values Screen4. Press OK. The Attach Calibration Bottle screen displays.Figure 3G-3: Attach Calibration Bottle Screen3G-2
PC Maintenance5. Attach the calibration bottle to the calibration port and open the calibrationbottle valve.6. Verify that the bottle pressure gauge reads at least 30 psi (207 kPa).7. Select Continue to perform the gas calibration.8. Turn the Gas Calibration bottle off and disconnect when the calibration iscomplete.Figure 3G-4: Gas Calibration Complete Screen9. When the gas calibration has successfully completed, select Exit tocontinue using FGA.If the gas calibration fails,1. Verify that the gauge(s) read at least 50 psi.2. Verify that the gas bottle is properly attached.3. Re-run the calibration.4. Verify that the proper gas values were entered.Leak CheckPerforming a leak check on the sample system helps to ensure that no vacuumleaks or hardware leaks exist to cause inaccurate readings. Leak check intervalsmay vary depending on the application and use, but once per week as part ofroutine maintenance is recommended, more frequently with heavy use.To perform a Leak Check,1. Press the Calibration button and choose Leak Check from the menu.— The Leak Check window opens.Figure 3G-5: Leak Check initial Screen3G-3
- Page 87 and 88: Pocket PC MaintenanceTo perform the
- Page 89 and 90: Pocket PC Maintenance7. Press the S
- Page 91 and 92: Pocket PC MaintenanceFigure 2G-34:
- Page 93 and 94: Pocket PC MaintenanceLow Out FlowTh
- Page 95 and 96: Pocket PCOther Functions2HAboutThis
- Page 97 and 98: 3Using FGA with the PCThe Flexible
- Page 99 and 100: 3APC Software OverviewThe PC versio
- Page 101 and 102: PC Software OverviewCommands 1-10Co
- Page 103 and 104: PC Software OverviewCommandsFigure
- Page 105 and 106: 3BPC SoftwareInstallationThe Flexib
- Page 107 and 108: PC Software Installation5. If this
- Page 109 and 110: 3CPC Meters/Strip ChartThis chapter
- Page 111 and 112: PC Meters/GraphGas Bench MenuThe Ga
- Page 113 and 114: PC Meters/GraphStrip ChartTo view t
- Page 115 and 116: PC Meters/GraphStrip Chart Display
- Page 117 and 118: PC Meters/GraphWhen you select the
- Page 119 and 120: PC Meters/GraphTachometerUse the Ta
- Page 121 and 122: PC Automatic Tests3DTSI TestThis se
- Page 123 and 124: PC Automatic TestsTachometer Config
- Page 125 and 126: PC Automatic TestsIdle RPM TestFGA
- Page 127 and 128: PC Automatic TestsTSI Results16. Th
- Page 129 and 130: 3EPC Customer/VehicleFGA is designe
- Page 131 and 132: PC Customer/VehicleInformation bar
- Page 133 and 134: PC Customer/VehicleWhen the desired
- Page 135 and 136: 3FPC ResultsThe Test Results window
- Page 137: 3GPC MaintenanceGas CalibrationThis
- Page 141 and 142: PC MaintenanceNew NO x SensorNew NO
- Page 143 and 144: PC MaintenanceShop InformationThe S
- Page 145 and 146: PC MaintenanceThe Setup window cont
- Page 147 and 148: PC MaintenanceTSI Test SetupThe TSI
- Page 149 and 150: PC MaintenanceZero Purge Time(s)—
- Page 151 and 152: PC MaintenanceDiagnosticsService ut
- Page 153 and 154: PC Other Functions3HThis chapter co
- Page 155 and 156: 4General MaintenanceThis chapter ad
- Page 157 and 158: General MaintenanceOptionsPart Numb
- Page 159 and 160: General Maintenance3. Select the Ig
- Page 161 and 162: General MaintenanceWhen using the O
- Page 163 and 164: General Maintenance—Push the gray
- Page 165 and 166: General MaintenanceWhen FGA softwar
- Page 167 and 168: 5InterpretationThe section includes
- Page 169 and 170: Interpretation• Minimal NO is pro
- Page 171 and 172: Interpretation2. Repair the vehicle
- Page 173 and 174: Interpretation:The following is a c
- Page 175 and 176: AAppendix A:OBD II ConnectionsThe l
- Page 177 and 178: LIMITED WARRANTYTERMS AND CONDITION
PC Maintenance5. Attach the calibrati<strong>on</strong> bottle to the calibrati<strong>on</strong> port and open the calibrati<strong>on</strong>bottle valve.6. Verify that the bottle pressure gauge reads at least 30 psi (207 kPa).7. Select C<strong>on</strong>tinue to perform the gas calibrati<strong>on</strong>.8. Turn the Gas Calibrati<strong>on</strong> bottle off and disc<strong>on</strong>nect when the calibrati<strong>on</strong> iscomplete.Figure 3G-4: Gas Calibrati<strong>on</strong> Complete Screen9. When the gas calibrati<strong>on</strong> has successfully completed, select Exit toc<strong>on</strong>tinue using FGA.If the gas calibrati<strong>on</strong> fails,1. Verify that the gauge(s) read at least 50 psi.2. Verify that the gas bottle is properly attached.3. Re-run the calibrati<strong>on</strong>.4. Verify that the proper gas values were entered.Leak CheckPerforming a leak check <strong>on</strong> the sample system helps to ensure that no vacuumleaks or hardware leaks exist to cause inaccurate readings. Leak check intervalsmay vary depending <strong>on</strong> the applicati<strong>on</strong> and use, but <strong>on</strong>ce per week as part ofroutine maintenance is recommended, more frequently with heavy use.To perform a Leak Check,1. Press the Calibrati<strong>on</strong> butt<strong>on</strong> and choose Leak Check from the menu.— The Leak Check window opens.Figure 3G-5: Leak Check initial Screen3G-3