Operating Instructions - Snap-on Equipment

Operating Instructions - Snap-on Equipment Operating Instructions - Snap-on Equipment

snaponequipment.com
from snaponequipment.com More from this publisher
13.07.2015 Views

PC MaintenanceThe initial Gas Calibration window opens.Figure 3G-1: Initial Gas Calibration Screen✓ A warning displays and will automatically disappear followed by an autozero.2. The system then performs a zero calibration. In zero calibration, the gasanalyzer submits clean air, room air, into the sample system to verify thatthe analyzer reads properly when there is no gas present in the sample cell.✓ Make sure the vehicle exhaust sample probe is in fresh air, NOTsampling vehicle exhaust of any type.3. The Calibration Bottle Values screen then displays. Verify or change thevalues to match your calibration gas.— To change values, Tab to the value you want to change, press Deleteand input the correct value.Figure 3G-2: Calibration Bottle Values Screen4. Press OK. The Attach Calibration Bottle screen displays.Figure 3G-3: Attach Calibration Bottle Screen3G-2

PC Maintenance5. Attach the calibration bottle to the calibration port and open the calibrationbottle valve.6. Verify that the bottle pressure gauge reads at least 30 psi (207 kPa).7. Select Continue to perform the gas calibration.8. Turn the Gas Calibration bottle off and disconnect when the calibration iscomplete.Figure 3G-4: Gas Calibration Complete Screen9. When the gas calibration has successfully completed, select Exit tocontinue using FGA.If the gas calibration fails,1. Verify that the gauge(s) read at least 50 psi.2. Verify that the gas bottle is properly attached.3. Re-run the calibration.4. Verify that the proper gas values were entered.Leak CheckPerforming a leak check on the sample system helps to ensure that no vacuumleaks or hardware leaks exist to cause inaccurate readings. Leak check intervalsmay vary depending on the application and use, but once per week as part ofroutine maintenance is recommended, more frequently with heavy use.To perform a Leak Check,1. Press the Calibration button and choose Leak Check from the menu.— The Leak Check window opens.Figure 3G-5: Leak Check initial Screen3G-3

PC Maintenance5. Attach the calibrati<strong>on</strong> bottle to the calibrati<strong>on</strong> port and open the calibrati<strong>on</strong>bottle valve.6. Verify that the bottle pressure gauge reads at least 30 psi (207 kPa).7. Select C<strong>on</strong>tinue to perform the gas calibrati<strong>on</strong>.8. Turn the Gas Calibrati<strong>on</strong> bottle off and disc<strong>on</strong>nect when the calibrati<strong>on</strong> iscomplete.Figure 3G-4: Gas Calibrati<strong>on</strong> Complete Screen9. When the gas calibrati<strong>on</strong> has successfully completed, select Exit toc<strong>on</strong>tinue using FGA.If the gas calibrati<strong>on</strong> fails,1. Verify that the gauge(s) read at least 50 psi.2. Verify that the gas bottle is properly attached.3. Re-run the calibrati<strong>on</strong>.4. Verify that the proper gas values were entered.Leak CheckPerforming a leak check <strong>on</strong> the sample system helps to ensure that no vacuumleaks or hardware leaks exist to cause inaccurate readings. Leak check intervalsmay vary depending <strong>on</strong> the applicati<strong>on</strong> and use, but <strong>on</strong>ce per week as part ofroutine maintenance is recommended, more frequently with heavy use.To perform a Leak Check,1. Press the Calibrati<strong>on</strong> butt<strong>on</strong> and choose Leak Check from the menu.— The Leak Check window opens.Figure 3G-5: Leak Check initial Screen3G-3

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!