H100B Pulse Oximeter User <strong>Manual</strong>PR Alarm is from “1 + the lower limit of PRAlarm” to 300. Press the Up Arrow or Down Arrow button toincrease or decrease values. Choose PR Lo Alarm in menu, press theFunction button enter setup. The PR LoAlarm box changes from real line to brokenline. The adjustable range for the lower limitof PR Alarm is from 0 to “the upper limit ofPR Alarm – 1”. Press the Function button, confirm the alarmrange setup. Hi Alarm is always higher than Lo Alarm byat least 1 bpm. Press the On/Off button to exit the menu, andreturn to measurement state.3.3.12 Patient ID No. SetupThe oximeter can support 100 patient IDs, and 300-hourdata storage.When entering menu, press the Function button to setID (valid range is from 1 to 100). The ID display box onthe interface will change from real line to broken line.- 47 -