H100B Pulse Oximeter User <strong>Manual</strong>Checking the Ni-MH Rechargeable BatteryPackageThe performance of a Ni-MH rechargeable batterypackage may deteriorate. To check the performance ofthe battery, follow the procedures below:1. Disconnect the pulse oximeter from the patientand stop all monitoring and measuringprocedures.2. Place the pulse oximeter in the charger standand connect the AC mains. Allow the battery tobe charged uninterruptedly for above 2.5 hours.For details about charging the Ni-MHrechargeable battery package, please refer tosection 3.4.- 26 -