12. Appendix: Error messages and their elimination11. "Condensate container full"Description: Level sensor in condensate receptacle sends status message to PLCCauses: • Condensate receptacle is full Elimination: • Empty condensate receptacle• Level sensor defective • Check level sensor (floating contact should move freely), replace levelsensor if necessary12. "Zero offset drift too large"Description: Zero point offset values between 2 zero checks > 3%Causes: • Extremely soiled sampling system Elimination: • Clean (purge) sampling system• <strong>Gas</strong> sample cooling (gas drying) not operating correctly,condensation in gas duct and particularly in Hg photometer• See error elimination under 4 and 10(Clean quartz glass windows in Hg photometer,clean measuring and reference cuvettes if necessary)13. "Service request"Description: A maintenance interval has been reachedCauses: • The device has been operating without maintenance of 6 months Elimination: • Request service visit for maintenanceFurther information for operationDescription: Calibration value is not reachedCauses,counter-measures:• The measuring instrument must be checked for leaks from the sampling onwards by gradually closing the gas path. AThe criterion for an assessment as freeof leaks is a system pressure of 500 mbar, including at the fittings for the hoses carrying sample gas. In any case, the wearing parts of the sample gas pump(structure diaphragm, valve plates, sealing rings) should also be checked and replaced. In rare cases, leaks can also occur in the heated sampling system(sampling probe, sampling line), particularly at the connecting points.In the event of an error it is advisable to inform the device manufacturer / supplier.85 <strong>HM</strong> <strong>1400</strong> <strong>TRX</strong>

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!