13.07.2015 Views

HM 1400 TRX Total Mercury Analyser - Webshop, Gas Analysis ...

HM 1400 TRX Total Mercury Analyser - Webshop, Gas Analysis ...

HM 1400 TRX Total Mercury Analyser - Webshop, Gas Analysis ...

SHOW MORE
SHOW LESS
  • No tags were found...

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

9. Appendix: Technical data9 Appendix: Technical dataMeasuring range:0 - 45 to 0-400 µg/Nm³ selectableTested measuring range:Detectable limit:0 - 45 µg/Nm³< 0,5 µg/Nm³Reference method: Basic calibration according to VDI 3868 p. 2Maintenance interval according toEN15267-3:> 1 monthAvailability: > 95%Warm-up time:Setting time (90% time):Energy supply:Power consumption:Permissible ambient temperature:Measuring value output:<strong>Mercury</strong> detector:Partial gas volume flow:Converter120 minutes< 180 seconds230/400 V ( + 10%, - 15%), 50 Hz, 3L, N, PE, max. 10 kVA230 V ( + 10%, - 15%), 50 Hz, L, N, PE, max. 5 kVAMeasuring device: approx. 1200 WSampling probe: approx. 650 WSampling line: approx. 100 W / mSampling tube (heated): for tube lengths0.6m/1.0m/1.5: 600 W / 800 W / 1200 WMounted cooling unit (option): 850 W,rated current: 5.4 A, initial current: 22.0 A+ 5°C to +40°C4 - 20 mAUV absorption two-beam photometer (253.7nm)60 - 120 Nl/hthermocatalytic, running life: typ. 6 monthHousing Sheet steel cabinet, RAL 7035Class of protection:Dimensions:HxWxDWeight:IP551650 x 800 x 500 mm1750 x 800 x 500 mm with base/plinth 100 mm1850 x 800 x 500 mm with base/plinth 200 mm2350 x 800 x 600 with base/plinth andmounted cooling unitapprox. 200 – 240 kgSampling system: Sampling probe with heated 2 µm filter, IP 54sampling line, heated T max 200°C, PTFE, IP54Table 9.1: Technical data75 <strong>HM</strong> <strong>1400</strong> <strong>TRX</strong>

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!