13.07.2015 Views

HM 1400 TRX Total Mercury Analyser - Webshop, Gas Analysis ...

HM 1400 TRX Total Mercury Analyser - Webshop, Gas Analysis ...

HM 1400 TRX Total Mercury Analyser - Webshop, Gas Analysis ...

SHOW MORE
SHOW LESS
  • No tags were found...

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

7. Operation7.5 Calibration (option)Figure 7.8 CalibrationAttention! If the device is currently in measuring mode, triggering calibrationwill result in an interruption of measuring mode.After triggering the calibration process by pressing the key and enteringa valid password (if necessary), you will be prompted to enter the followingparameters.Concentration of the stock solutionDosing capacity of the dosing pumpPressing the key again displays the parameters required to calculate thespan (volume flow, gas temperature and system pressure).<strong>HM</strong> <strong>1400</strong> <strong>TRX</strong> 63

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!