7. Operation7.5 Calibration (option)Figure 7.8 CalibrationAttention! If the device is currently in measuring mode, triggering calibrationwill result in an interruption of measuring mode.After triggering the calibration process by pressing the key and enteringa valid password (if necessary), you will be prompted to enter the followingparameters.Concentration of the stock solutionDosing capacity of the dosing pumpPressing the key again displays the parameters required to calculate thespan (volume flow, gas temperature and system pressure).<strong>HM</strong> <strong>1400</strong> <strong>TRX</strong> 63