7. Operation7.3 Zero offset (manual)Figure 7.6 The zero offset determination started manuallyThe zero offset determination can be started manually.Attention! If the device is currently in measuring mode, triggering a zero offsetmeasurement will result in an interruption of measuring mode.After triggering the zero offset measurement by pressing the key, acurrent measuring value ("zero" measuring value) will appear in the display inμg/Nm³. To ensure that the "zero" measuring value is not falsified by anyresidual quantities of mercury in the suction intake, you must wait a fewminutes for the value to stabilise before accepting it.After pressing the key, the zero offset measuring value in μg/Nm³ isstored as offset. The display changes back to the standard operating screenafter approx. 2 seconds.<strong>HM</strong> <strong>1400</strong> <strong>TRX</strong> 61