6. Operation6.3 Measuring modeFunctionally relevant parameters are cyclically monitored in measuring mode(see Appendix:Device parameters in delivery status and Appendix: Errormessages and their elimination).If one of these parameters moves outside its preset limit values, an errormessage is generated in the event memory with the date and time of the eventstart and is indicated by a flashing LED on the help key "?". The measuringdevice switches into standby mode after 120 seconds. If condensate is foundin the monitor this event has a higher priority, and the measuring device is putinto standby mode immediately. If the level sensor of the condensatereceptacle indicates "Condensate container full", switch-off (standby) onlyoccurs after 2 hours. Only after eliminating the error cause can the measuringdevice be switched into measuring mode. This is either done manually orautomatically, depending on the setting.48 <strong>HM</strong> <strong>1400</strong> <strong>TRX</strong>