<strong>Column</strong>s for Special ApplicationsLong Lifetime of RRHT <strong>Column</strong>s at Elevated Temperatures<strong>Column</strong>: ZORBAX SB-C18827700-9022.1 x 50 mm, 1.8 µmMobile Phase: A: 60% H 2 OB: 40% ACNFlow Rate: 1 mL/minTemperature: 80°CDetector: UV 254 nmSample:QC Test MixPlates1200010000800060004000200001499718853988115313051506164117761911204621812319250826432778291330853220346836033738388740544189432445374756489250335201Injection NumberLCRR005Comparison of Efficiencies – RRHT (1.8 µm)and Rapid Resolution (3.5 µm) <strong>Column</strong>s<strong>Column</strong> LengthResolving PowerN (3.5 µm)*Resolving PowerN (1.8 µm)High Resolution150 21,000 32,500100 14,000 24,00075 10,500 17,000**Ultra Fast50 7000 12,00030 4200 600020 — 350015 2100 2500Resolution α N 1/2*5 µm <strong>HPLC</strong> columns of the same length have 40% fewer plates (N-value); 4.6 mm ID**Available as a custom columnOrder online at www.agilent.com/chem/store81