Testo 350 Calibration - Actoolsupply.com
Testo 350 Calibration - Actoolsupply.com
Testo 350 Calibration - Actoolsupply.com
- No tags were found...
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
www.actoolsupply.<strong>com</strong>5. Step by step - here, using the testo <strong>350</strong>Example: Control and adjustment of a CO module with test gas1000 ppm/1.4 % O 2.1. Switch the unit on and wait until the ON phase is over.2. Connect the test gas to the analyser via the flue gas probe.3. Open the test gas bottle and set the flow rate on theflowmeter to about 1.5 l/min.4. Press the Start key to start the pump in the analyser. Youmay need to regulate the test gas volume again (watch theflowmeter).www.actoolsupply.<strong>com</strong>25