www.actoolsupply.<strong>com</strong>Special advice for the adjustment process:! Make sure that the ambient temperature remains the samethroughout gas adjustment.! Wait until the devices have warmed up (e.g. about 30 minutesfor the testo <strong>350</strong> M/XL).! To ensure the greatest possible accuracy, in the ideal case thesame ambient conditions should prevail during gas adjustmentas during real gas measurement.! Do not choose a test gas concentration that is too low for theslope adjustment. See Point 4.4 for a re<strong>com</strong>mendation.! The zero point and slope value must be stable before adjustmentis started.! With mixed gases, allow for the influences due to cross-sensitivities.It may be that the corresponding cross-sensitivity alsohas to be adjusted.www.actoolsupply.<strong>com</strong>15