13.07.2015 Views

Dynamica General Catalog 2012_9_25.indd - Scientifix

Dynamica General Catalog 2012_9_25.indd - Scientifix

Dynamica General Catalog 2012_9_25.indd - Scientifix

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

AERO A700 SPECIFICATIONSColumn OvenSample InjectionDetectorGC Main UnitTemperatureAmbient +5°C to 420°C (increment of 1°C)Temperature Accuracy± 0.05°C (≤250°C)Temperature Ramp0.1 to 80°C / min (max. 22 ramp)Capacity16LTemperature Ambient +5°C to 420°CInjection PortsUp to 3 PortsHeating Zone 3Flow Control ModeAutomatic Flow / Pressure ControlPressure0 to 40 psiTemperature Ambient +5°C to 420°CNo. of DetectorsUp to 3 DetectorsDimensions (w x d x h)630 x 570 x 510 mmWeight55KgPower RequirementsAC 220 - 240V, 50HzFIDDetection LimitBaseline NoiseBaseline Drift≤ 5 x 10 -12 g/s (n-hexadecane)≤ 5 x 10 -14 A≤ 1 x 10 -13 A/30minNPDDetection LimitBaseline NoiseBaseline DriftN ≤ 5 x 10 -12 g/s (Azobenzene)P ≤ 5 x 10 -13 g/s (Malathion)≤ 2 x 10 -13 A≤ 4 x 10 -13 A/30minTCDSensitivityBaseline NoiseBaseline Drift10,000mV • ml/mg (n-hexadecane)≤ 30μV (99.999% H 2 gas)≤ 100μV/30minFPDDetection LimitBaseline NoiseBaseline DriftS ≤ 8 x 10 -11 g/s (Parathion-methyl)P ≤ 8 x 10 -13 g/s (Parathion-methyl)≤ 8 x 10 -11 A≤ 5 x 10 -12 A/30minECDSensitivityBaseline NoiseBaseline DriftRadiation Source≤ 3 x 10 -14 g/ml (γ-BHC)15μV≤ 100μV/30minNi 63 , ≤ 10mCiPIDDetection LimitBaseline NoiseBaseline DriftS ≤ 5 x 10 -12 g/s (benzene)≤ 5 x 10 -14 A≤ 1.2 x 10 -12 A/30min48