SINGLE-TOOTH SCREW-RETAINEDLab 7 - Finish copingConfirm the fit of the coping to the implant on the working model. Returnthe coping and the Abutment Screw to the clinician for patient try-in.Send to Clinician• Metal coping• Abutment Screw• Working modelClinical 8 - Remove healing abutmentSanitize the coping (per standard clinical procedure). Remove the HealingAbutment from the implant with the .050” (1.25mm) Hex Driver. Makesure the prosthetic platform is free of bone and soft tissue.Clinical 9 - Coping try-inPlace the custom coping onto the implant and hand-tighten theAbutment Screw with the .050” (1.25mm) Hex Driver (Regular or Long,as needed). Take a radiograph along the long axis of the implant toensure the abutment is seated completely on the implant. Confirm fitand contour, assuring there is adequate space for the veneer material.Adjust as necessary. Remove the coping from the implant and replacethe Healing Abutment with the .050” (1.25mm) Hex Driver. Return thecoping to the lab.Send to Lab• Metal coping• Abutment Screw• Working model• Shadeshop online at www.biohorizons.com50