46 3. Ir<strong>radiation</strong> Facility3.5 Measurement Setup, Methods and AnalysisIn the scope <strong>of</strong> this work, full depletion voltage (FDV) and bulk reverse current weremeasured on samples under study. The measurement setup as used <strong>in</strong> the laboratory isdescribed while the one used at the reactor was described <strong>in</strong> section 3.3.2. S<strong>in</strong>ce bothsetups are only slightly dierent, the same measurement method and analysis were used.3.5.1 Laboratory SetupHP 4284ALCR meterKEITHLEY 237HV meas. unitR=4.7kΩC=100nFcca. 3mBPPADGRFigure 3.10: Measurement setup as used <strong>in</strong> the laboratory. BP, TOP and GR are connectionsto sample back plane, pad and guard r<strong>in</strong>g, respectively.The ma<strong>in</strong> components <strong>of</strong> the measurement setup were a LCR meter, a precise ampermeterand a <strong>high</strong> voltage source (up to cca. 1000 V). All units were computer controlled.In the laboratory, a HP4284A LCR measure unit and a KEITHLEY 237 <strong>high</strong> voltagesource-measure unit were used. A Yokogawa UT15 temperature controller connected toa Pt100 sensor placed close to the sample was used to measure the sample temperature.All units were controlled and read out by a personal computer, where some on-l<strong>in</strong>e dataanalysis was also per<strong>for</strong>med.To measure the capacitance at dierent depletion depths, i.e. at dierent voltages,an external DC voltage has to be applied to the sample. In the laboratory, a KEITHLEY237 <strong>high</strong> voltage source-measure unit was used as an external power source. It can provideDC voltage up to 1100 V with a current limit <strong>of</strong> 1 mA. To shield the LCR meter from <strong>high</strong>DC bias<strong>in</strong>g voltages, decoupl<strong>in</strong>g capacitors were used as shown <strong>in</strong> gure 3.10. To reduce

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!