SP88/SP98 Spectrophotometer Operating Manual ... - X-Rite
SP88/SP98 Spectrophotometer Operating Manual ... - X-Rite SP88/SP98 Spectrophotometer Operating Manual ... - X-Rite
8.3 Replacing The Battery Pack1. Unplug the AC adaptor and turn the power switch Off.2. Carefully place the instrument on it’s top and lock the shoe next toinstrument.3. Remove battery access cover by sliding towards rear of unit.4. Grasp plastic tab that extends from battery pack and pull until pack isremoved.5. Slide new (or charged) battery pack into instrument until connector isproperly seated.6. Reinstall battery access cover bysliding into position.7. Discharged battery packs should becharged in the instrument for 16hours.Battery PackConnectorBattery AccessCover8-3
8.4 Replacing the Target Window☛ The target window alignment plug is only required when installing a target window on the SP88.Refer to Figure 11. Unplug the AC adaptor and turn the power switch Off.2. Carefully place instrument on endand remove old target window [1]by pushing outward from inside ofshoe [2]. Clean off any remainingadhesive from shoe [2].Fig 1[4][3]3. Insert target window alignmentplug [3] into sensor nose [4].[2]Refer to Figure 24. Compress shoe [2] all the waydown and lock in position.5. Remove paper backing [5] fromadhesive on new target window [1].[1]6. Place target window hole overalignment plug [3] and aligncrosshairs [6] squarely with shoe [2].7. Remove clear protective backing[7] from target window [1].Fig 2[2][3][5][6]8. Unlock shoe [2] and removealignment plug [3].Replacement Target Windows:SP88 InstrumentSP68-21-08 (Standard)SP68-21-14 (Large)SP98 InstrumentSP98-21[1][7]8-4
- Page 27 and 28: Positioning the Instrument on the W
- Page 29 and 30: 3.3 Calibration ProcedureTo perform
- Page 31 and 32: Calibration Procedure . . . continu
- Page 33 and 34: 4.2 Illuminant/Observer SelectionTh
- Page 35 and 36: Reference Entry . . . continued5) M
- Page 37 and 38: Sample Measurement . . . continued4
- Page 39 and 40: Using Measurement Averaging . . . c
- Page 41 and 42: Adjusting Tolerance Values . . . co
- Page 43 and 44: Adjusting Tolerance Values . . . co
- Page 45 and 46: 5.2 Pass/Fail MeasurementThe Pass/F
- Page 47 and 48: SECTION 6STORAGE OPERATIONSubjects
- Page 49 and 50: Storing Measurements . . . continue
- Page 51 and 52: Deleting Stored Measurements . . .
- Page 53 and 54: 6.3 Printing Stored MeasurementsThe
- Page 55 and 56: Printing Stored Measurements . . .
- Page 57 and 58: Tagging and Storing Operation . . .
- Page 59 and 60: Tagging and Storing Operation . . .
- Page 61 and 62: 6.5 QA-Master ‘‘Run Job’’ O
- Page 63 and 64: Setting RS-232 Communication Option
- Page 65 and 66: 7.2 Setting Operation OptionsThere
- Page 67 and 68: Setting Operation Options . . . con
- Page 69 and 70: Setting Operation Options . . . con
- Page 71 and 72: Setting Printout Options . . . cont
- Page 73 and 74: 7.4 Setting Date And TimeThe date a
- Page 75: 8.2 Cleaning the OpticsThe optics s
- Page 79 and 80: 8.6 Troubleshooting TipsInstrument
- Page 81 and 82: 9.2 Scanning a Bar CodeFollow these
- Page 83 and 84: Troubleshooting . . . continuedWas
- Page 85 and 86: Technical Specifications . . . cont
- Page 87 and 88: Display Messages . . . continuedCal
- Page 89 and 90: APPENDIX - DCOLOR CHECKThe color ch
- Page 91 and 92: COLOR CHECK RECORDDate: Illuminant:
8.4 Replacing the Target Window☛ The target window alignment plug is only required when installing a target window on the <strong>SP88</strong>.Refer to Figure 11. Unplug the AC adaptor and turn the power switch Off.2. Carefully place instrument on endand remove old target window [1]by pushing outward from inside ofshoe [2]. Clean off any remainingadhesive from shoe [2].Fig 1[4][3]3. Insert target window alignmentplug [3] into sensor nose [4].[2]Refer to Figure 24. Compress shoe [2] all the waydown and lock in position.5. Remove paper backing [5] fromadhesive on new target window [1].[1]6. Place target window hole overalignment plug [3] and aligncrosshairs [6] squarely with shoe [2].7. Remove clear protective backing[7] from target window [1].Fig 2[2][3][5][6]8. Unlock shoe [2] and removealignment plug [3].Replacement Target Windows:<strong>SP88</strong> InstrumentSP68-21-08 (Standard)SP68-21-14 (Large)<strong>SP98</strong> Instrument<strong>SP98</strong>-21[1][7]8-4