3.3 Calibration ProcedureTo perform a calibration:1) Make sure white cal standard and black trap are clean (see calibrationnotes Section 3).2) Press [menu] key #1 twice toenter MENU PAGE 2.L* 94.10 INCLUDEDC* 21.88h 193.97menu L*C*h D65 10 difKey#1 Key#2 Key#3 Key#4Press Twice3) Press [cal] key #2 to entercalibration function.exitMENU PAGE 2calcnfgKey#1 Key#2 Key#3 Key#4- The software datecode and copyrightare momentarily displayed.CALIBRATIONX-RITE <strong>SP88</strong>VERS. XXXX/XXXX-XCOPYRIGHT 1993- ‘‘READ WHITE CALIBRATIONPLAQUE’’ is displayed.☛ Pressing [exit] key #1 will abort the calibrationprocedure and calibration will not be updated.READ WHITECALIBRATION PLAQUEexitKey#1 Key#2 Key#3 Key#43-5➩