US Clinical Trial Key LearningsGORE HELEX Septal Occluder is safe and effective for the closure of ASDsNo instances of perforation of cardiac structuresNo instances of thrombus on the deviceOptimal results when 2:1 device diameter <strong>to</strong> defect diameter ratio is used <strong>to</strong>close defects up <strong>to</strong> 18 mmClinical Trial DataDevice <strong>to</strong> Defect RatioAccess< 1.61.6 – 2.02.0 or greaterClinical Closure 92.6% 96.3% 100.0%Embolization Rate 3.0% 4.2% 0.0%Composite Success 80.0% 90.0% 95.0%*Information comes from Latson, et al., Analysis of fac<strong>to</strong>rs related <strong>to</strong> successful transcatheter closure of secundum atrial septal defects usingthe HELEX septal occluder. American Heart Journal 2006;151:1129.e7-1129.e11.