US Clinical Trial Data SummaryFeasibility Pivotal * Continued AccessSubjects Enrolled 63 143 196Median Age (years) 11 (0.5-66) 7 (1-72) 5.4 (0.8 – 58)Successful Implant86.4% of attemptedimplants (51 / 59)88.1% of attemptedimplants (119 / 135)85.6% of attemptedimplants (137 / 160)Adverse Events 3.9% 5.9% 2.2%Clnically Successful94.6% 98.1% 99.1%OcclusionComposite Clinical Success 89.5% 91.9% 96.6%*Two arm study comparing surgical closure <strong>to</strong> device closureClinically Successful Occlusion: No residual leak or insignificant residual leakComposite Clinical Success: Clinically successful occlusion AND no major adverse eventThree US clinical studies were conducted <strong>to</strong> evaluate the GORE HELEX Septal Occluder. These studies were performedwith the original delivery system. The product described here is the same Occluder with a modified delivery system.Please note that the modified delivery system was not evaluated under the original US clinical study.