Technical data <strong>Proline</strong> <strong>Promass</strong> 83Installation instructions ä 1410.1.7 Operating conditions: InstallationInlet and outlet runsThere are no installation requirements regarding inlet and outlet runs.Length of connecting cableremote versionmax. 20 m (65 ft)System pressure ä 1510.1.8 Operating conditions: EnvironmentAmbient temperature rangeStorage temperature!Note!Sensor and transmitter:• Standard: –20 to +60 °C (-4 to +140°F)• Optional: –40 to +60 °C (-40 to +140°F)• Install the device at a shady location. Avoid direct sunlight, particularly in warm climatic regions.• At ambient temperatures below –20 °C (–4 °F) the readability of the display may be impaired.–40 to +80 °C (–40 to +175 °F), preferably +20 °C (+68 °F)Degree of protectionStandard: IP 67 (NEMA 4X) for transmitter and sensorShock resistance According to IEC 60068-2-31Vibration resistance Acceleration up to 1 g, 10 to 150 Hz, following IEC 60068-2-6CIP cleaningYesSIP cleaningYesElectromagnetic compatibility(EMC)To IEC/EN 61326 and NAMUR Recommendation NE 21Page 128 • MN0M022Issue/Rev. 0.2 (12/12)