12.07.2015 Views

Proline Promass 83E/F/O Operation Manual - FMC Technologies

Proline Promass 83E/F/O Operation Manual - FMC Technologies

Proline Promass 83E/F/O Operation Manual - FMC Technologies

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

<strong>Proline</strong> <strong>Promass</strong> 83Technical dataZero point stability <strong>Promass</strong> F (high-temperature version)DNZero point stability <strong>Promass</strong> F (high-temperature version)[mm] [inch] [kg/h] or [l/h] [lb/min]25 1 1.80 0.066150 2 7.00 0.257280 3 18.0 0.6610Example for max. measured error[%]±1.0±0.5±0.200 2 4 6 8 10 12 14 16 18 t/hFig. 52: Max. measured error in % o.r. (example: <strong>Promass</strong> F, DN 25)a0004604Flow values (example)Turn down Flow Maximum measured error[kg/h] [lb/min] [% o.r.]500 : 1 36 1.323 1.5100 : 1 180 6.615 0.325 : 1 720 26.46 0.110 : 1 1800 66.15 0.12 : 1 9000 330.75 0.1Design fundamentals ä 114RepeatabilityDesign fundamentals ä 114.• Mass flow and volume flow (liquids):±0.025% o.r. (PremiumCal, for mass flow)±0.05% o.r.• Mass flow (gases): ±0.25% o.r.• Density (liquids): ±0.00025 g/cm 3• Temperature: ±0.25 °C ± 0.0025 · T °C; ±0.5 °F ± 0.0015 · (T - 32) °FIssue/Rev. 0.2 (12/12) MN0M022 • Page 113