Technical data <strong>Proline</strong> <strong>Promass</strong> 83Performance characteristic<strong>Promass</strong> Eo.r. = of reading; 1 g/cm 3 = 1 kg/l; T = medium temperatureMaximum measured errorThe following values refer to the pulse/frequency output.The additional measured error at the current output is typically ±5 A.Design fundamentals ä 112.• Mass flow and volume flow (liquids): ±0.25% o.r.• Mass flow (gases): ±0.75% o.r.• Density (liquids)– Reference conditions: ±0.0005 g/cm³– Field density calibration: ±0.0005 g/cm³(valid after field density calibration under process conditions)– Standard density calibration: ±0.02 g/cm³(valid over the entire temperature range and density range ä 129)• Temperature: ±0.5 °C ± 0.005 · T °C; ±1 °F ± 0.003 · (T - 32) °FZero point stabilityDNZero point stability[mm] [inch] [kg/h] or [l/h] [lb/min]8 3/8 0.20 0.007415 ½ 0.65 0.023925 1 1.80 0.066240 1½ 4.50 0.165450 2 7.00 0.257380 3 18.00 0.6615Example for max. measured error[%]±1.0±0.5±0.200 1 2 4 6 8 10 12 14 16 18 t/hFig. 51: Max. measured error in % o.r. (example: <strong>Promass</strong> E, DN 25)a0004611Page 110 • MN0M022Issue/Rev. 0.2 (12/12)