<strong>Proline</strong> <strong>Promass</strong> 83Technical dataMeasuring ranges for gases (<strong>Promass</strong> A)DNx[mm][inch]1 1/24 322 1/12 324 1/8 32Measuring ranges for gases (<strong>Promass</strong> I)DNx[mm][inch]8 3/8 6015 ½ 8015 FB ½ FB 9025 1 9025 FB 1 FB 9040 1 ½ 9040 FB 1 ½ FB 9050 2 9050 FB 2 FB 11080 3 110FB = Full bore versions of <strong>Promass</strong> IMeasuring ranges for gases (<strong>Promass</strong> X)DNx[mm][inch]350 14 200Calculation example for gas:• Sensor type: <strong>Promass</strong> F, DN 50• Gas: air with a density of 60.3 kg/m³ (at 20 °C and 50 bar)• Measuring range (liquid): 70000 kg/h• x = 90 (for <strong>Promass</strong> F DN 50)Max. possible full scale value:g max(G) = g max(F) · (G) ÷ x [kg/m³] = 70000 kg/h · 60.3 kg/m³ ÷ 90 kg/m³ = 46900 kg/hRecommended full scale valuesSee Page 127 ff. ("Limiting flow")Operable flow rangeGreater than 1000 : 1. Flows above the preset full scale value do not overload the amplifier, i.e.totalizer values are registered correctly.Issue/Rev. 0.2 (12/12) MN0M022 • Page 105