Change in sensitivity with temperature is found by dividing Equation (15) by R and differentiating with respectto temperature. This givesd(E/R) -P ab=(16)dT 8Since P and 8 increase with temperature at about the same rate, the change in sensitivity with temperature remainsapproximately constant at about 2.5 × 10 -3 µvolts bar -1 °C -1 or about 0.3 percent per degree from 10 to 50°C. Thus,over the normal range of temperatures encountered in water potential measurements, little, if any, temperaturecorrection for thermocouple sensitivity is required. However, A v does change with temperature. This temperaturedependence is discussed in Paragraph 3.5.24