CARDELL® <strong>MAX</strong>-<strong>12</strong><strong>HD</strong> User’s Manual5. CO 2 Waveform FillTo fill the CO 2 Waveform or not:MAIN MENU Rotate Knob CO2 Press Knob WaveformFilling Press Knob Rotate Knob Press Knob<strong>12</strong>.3 Alarm SetupCO2 monitoring alarm includes parameter limit alarm and abnormal statusalarm. Alarm is to give alert when the monitoring results are abnormal. It isaudible and visual with LED indicators and flashing readings.NOTE: Default alarm limits should be adjusted according to patientcondition.Parameter Limit Range:ParameterEtCO 2 Upper LimitEtCO 2 Lower LimitInCO 2 Upper LimitInCO 2 Lower LimitREST Upper LimitREST Lower LimitRangelower limit-100 (mmHg)0-upper limit (mmHg)lower limit-100 (mmHg)0-upper limit (mmHg)lower limit-100 (mmHg)0-upper limit (mmHg)<strong>12</strong>.4 PrecautionsCO 2 measurement can be affected by:- Changes in atmospheric pressure- Halogenated anesthetic vapors- Calibration drift- Fluid contaminationInaccurate CO 2 readings can be caused by:- Reuse of or failure to change sampling line.M1.0 5-15-0995