<strong>Cardell</strong>® <strong>MAX</strong>-<strong>12</strong><strong>HD</strong> User’s ManualNOTE:During nebulization or suction (when not using closed suction system),in order to avoid moisture buildup and FilterLine occlusion, disconnect theAirway Adapter from the patient’s endotracheal tube.CAUTION: Loose or damaged connections may compromise ventilation orcause an inaccurate measurement of respiratory gases. Securely connect allcomponents and check connections for leaks.WARNING: When used with closed suction system, do not place the AirwayAdapter between the suction catheter and endotracheal tube.ConsumablesWith “T” Tube AdapterEndotrachealTube BoreDead SpaceFilterLine H Set, Regular (yellow) > 4.5 mm < 7.3 ccFilterLine H Set, Exotic (yellow) < 4.0 mm < 0.5 ccTable 3: Specifications for Intubated Consumables – Airway AdaptersNOTE: For more information on what to consider when choosing betweenairway adapters, refer to Appendix 3, “Dead Space – Cause, Effect &Control in Small Animal Anesthesia” by Dr. Robert Stein, founder of the<strong>Veterinary</strong> Anesthesia Support Group (www.VASG.org).92 M1.0 5-15-09