<strong>Cardell</strong>® <strong>MAX</strong>-<strong>12</strong><strong>HD</strong> User’s Manual4、Install the cuff hose to the NIBP connector of the patient monitor5、Make sure there is no block between the monitor and the hose6、Set blood pressure measurement correctly in the setup menu7、The cuff on the patient’s limb should be at the same level as the heart8、Press the blood pressure start key and start measuring blood pressureWARNINGS:1. Make sure there is no other pressure on the cuff.2. Wrong cuff size may result in inaccurate measurements.3. Make sure monitor is set to Large/Small corresponding to cuff used.4. To ensure the patient’s safety, never use cuff on the same limb where aninfusion is going on.5. Do not measure SpO2 or other parameters on the same limb where bloodpressure is measured.7. Do not apply cuff on an injured limb.8. Do not measure a patient’s blood pressure continuously or repetitively for along time.10. Use only accessories recommended by the manufacturer.11. Remove the cuff after turning the power off.10.4 Alarm Setup10.4.1 NIBP alarm ON/OFF and alarm limits setup82 M1.0 5-15-09