CARDELL® <strong>MAX</strong>-<strong>12</strong><strong>HD</strong> User’s Manual10.2 NIBP SetupNIBP setup includes: select measurement (“work”) mode, select cuff size,select auto measurement time interval, change NIBP measurement scale,pressure compensation, dynamic blood pressure, adjust alarm limit and turnthe alarm sound on or off.Enter NIBP setup:Rotate the knob toand press knob to confirm. Rotate to NIBP, pressknob to confirm. Users can rotate the knob to select items shown in thefollowing figure and press the knob to confirm:Fig. 10-1 NIBP Setup MenuSelect cuff size: Large/SmallSelect work mode: Manual, Auto and StatSelect time interval for auto NIBP measurementChange NIBP measurement scale: mmHg or kPaPress Test: test static pressDynamic Blood Pressure: Start/Stop dynamic blood pressure monitoringAlarm limit: adjust alarm upper and lower limitsAlarm On/Off: Turn on or off NIBP alarm sound.M1.0 5-15-0979