<strong>Cardell</strong>® <strong>MAX</strong>-<strong>12</strong><strong>HD</strong> User’s ManualWARNING: If the SpO2 upper limit is set to 100%, then, it is equivalent tono alarm limit.Abnormal Status Alarm: “Probe Off” alarm9.6 PrecautionsWARNING: Prolonged use or the patient’s condition may require changingthe sensor site periodically. Change sensor site and check skin integrity,circulatory status, and correct alignment every 2~3 hours. If there is anychange, move the sensor to another site.CAUTION: Clean the sensor surface with 70% ethanol before and afteruse. But do not immerse it totally into the liquid.CAUTION:Do not autoclave, ethylene oxide sterilize or radiate the sensor.9.7 Cleaning and Maintenance9.7.1 Clean the sensor:(1) Clean the sensor with a soft cloth moistened in soapy water ormild detergent, saline (1%) or one of the following solutions:Microzid (pure), Mucocit (4%), Incidin (10%), Cidex (pure),Sporicidin (1:16), Mucaso (3%), Buraton (pure), alcohol (pure),Alconox (1:84), Cetylcide (1:63).(2) Clean the sensor surface with soft cloth and let it dry completely.(3) Wipe the receiving part and flashing part of the sensor with softcloth immersed in the detergent or alcohol.72 M1.0 5-15-09