CARDELL® <strong>MAX</strong>-<strong>12</strong><strong>HD</strong> User’s ManualSENSOR REMOVALCAUTION: For the comfort of the patient and to avoid damaging the sensor,do not pull on the cable when removing the sensor and clip from the sensor site,but rather, unclip the sensor and remove from placement site.When SpO2 monitoring is completed, remove the sensor from the patient.To remove the sensor and clip from the patient, press the clip open and remove.When the probe is removed from the patient, the message “SpO2 Probe OFF”is displayed and an audible alarm sounds, indicating a connection has beenlost. To acknowledge the alarm, press the SILENCE/RESET pushbutton. Themonitor silences the audible and visual alarms and the message “SpO2 ProbeOFF” remains on the display.To remove the sensor from the clip, grasp the end of each sensor pad and pull itthrough to the inside of the clip. The sensor should pop out of the clip easily.DO NOT pull on the cable.9.5 Alarm SetupThe SpO2 alarm is for parameter out of limit or abnormal status. When theparameter exceeds the limit, the monitor will give both visual and audioalarms.Alarm Range:ParameterRangeSpo2 upper limit lower limit to 100%SpO2 lower limit0 to upper limitPulse rate upper limitLower limit to 250 (bpm)Pulse rate lower limit30 to upper limitM1.0 5-15-0971