CARDELL® <strong>MAX</strong>-<strong>12</strong><strong>HD</strong> User’s Manual8.3 Respiration Monitoring Preparation(1) Place electrodes in proper positions(2) Select proper respiration lead combination(3) Set respiration alarm limitsNOTE: Electrodes must be placed in proper positions.CAUTION: Patient motion may result in a respiration measurement error.8.4 Alarm SetupThe respiration-temperature alarm includes a parameter out-of-limit alarmand an abnormal status alarm. When the parameter is out of limit, the monitorwill give an alarm sound automatically, and the value displayed on the screenflashes at the same time.Set up alarm limits:Different parameters have different limitsWARNING: Alarm limits should be adjusted based on an individual patient’scondition.Parameter Range:ParameterRespiration upper limitRespiration lower limitApnea Alarm TimeAdjustment RangeLow limit to 150 bpm1 to upper limit5-<strong>12</strong>0 secondsAlarm for abnormal status: “electrode off”M1.0 5-15-0965